510(k) Data Aggregation

    K Number
    K171956
    Date Cleared
    2017-12-12

    (166 days)

    Product Code
    Regulation Number
    876.1075
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile endoscope biopsy sampling need to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope.

    Device Description

    Sterile endoscope biopsy sampling needle, Model: DT-EN-W7 is mainly composed of needle head, PEEK outer tube assembly, and handle. Sizes are available in 18 - 22 gauge needles. The length of the needles range is between 0 and 80 mm. The needle is used by advancing the entire device to the site of soft tissue sampling through endoscopic channel aperture. The needle is advanced with gentle but firm pressure. Once the needle is in position, stab hard into the tissue and connect the external end of needle tube with aspirator in negative pressure. Retreat the needle head into tube, and get the needle out of the endoscope through the channel. The sample can then be expelled from the stylet notch.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Sterile Endoscope Biopsy Sampling Needle". This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets explicit acceptance criteria through a standalone study in the way an AI/ML device would be assessed.

    Therefore, many of the specific questions about acceptance criteria for an AI/ML device are not directly applicable or answerable from this document. However, I can extract the relevant performance data summaries and the non-clinical testing conducted to establish substantial equivalence, which serves a similar purpose in the context of this device type.

    Here’s a breakdown based on the information available:

    1. A table of acceptance criteria and the reported device performance

    The document defines "Technological characteristics" which act as acceptance criteria for the device, and then reports that performance bench testing was successfully completed for these. However, exact numerical "reported device performance" values against each criterion are not provided in this summary, only that they met the requirements.

    Acceptance Criteria (Technological Characteristics)Reported Device Performance
    Smoothness and leakproofnessSuccessfully completed
    Details: Needle head moves freely and smoothly in outer tube. Tip stays in outer tube after retraction. Pull rod fixed apace when operating aspirator and turning it. No air leakage except air passageway.Meets all requirements
    Joint strengthSuccessfully completed
    Details: Connection of outer tube and handle ≥15N; connection of needle head and handle ≥15N.Meets all requirements
    Puncture forceSuccessfully completed
    Details: Needle tip sharp with maximum puncture force ≤ 8 N.Meets all requirements
    Stiffness and resistance to breakageSuccessfully completed
    Details: Good stiffness with span 15mm ±0.1mm, load 10N ±0.1N, deflection value ≤ 0.50mm. Not broken in the same plane at 20 degrees by two-way repeated bending for 20 times with span 17.5mm ±0.1 mm.Meets all requirements
    Reducing substanceSuccessfully completed
    Details: KMnO4 solution consumption volume of test liquid