Search Results
Found 4 results
510(k) Data Aggregation
(333 days)
- The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
- The Solitaire™ Revascularization Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
- The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ische to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (< 70 cc by CTA or MRA, < 25 cc by MR-DWI). Endovascular therapy with the device should start within 6-16 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
The Solitaire™ Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ Revascularization Device is intended for use in the neurovasculature such as the internal carotid artery, M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Revascularization Device facilitates clot retrieval and has Platinum/Iridium radiopaque markers on the proximal and distal ends and also features radiopaque markers along the circumference of the working length of the devices are supplied sterile and are intended for single-use only.
The prompt asks to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text.
Based on the provided information, the submission is for a labeling modification to reduce the recommended minimum vessel diameter for the Solitaire™ Revascularization Device, 4 mm device models. The core device design and Indications for Use remain unchanged from the predicate devices. Therefore, the performance criteria and studies focus on demonstrating that this labeling modification does not introduce new safety or effectiveness concerns.
Here's the breakdown of the information as requested:
1. Table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Bench Testing: | |
| Durability: Ability to withstand simulated use (delivery, resheathing, retrieval) in a representative tortuous model. | Acceptance criteria met. (The test demonstrates the device can endure the mechanical stresses under simulated use with the reduced vessel diameter.) |
| Radial Force: Maintain sufficient radial force at the minimum vessel diameter. | Acceptance criteria met. (The test confirms appropriate interaction with the vessel wall even in smaller diameters.) |
| Clinical Performance & Safety (for reduced vessel diameter): Equivalent clinical performance and safety profile in vessels down to 1.5-2.0mm compared to the predicate device in its approved vessel sizes. | Retrospective subgroup analysis of the STRATIS registry data demonstrated that the subject Solitaire™ Revascularization 4 mm device, with vessel diameters of 1.5 to 2.0 mm, has similar clinical performance and safety profile compared to the predicate Solitaire™ Revascularization 4 mm device. This indicates the device continues to meet expected safety and effectiveness when used in smaller vessels within the new recommended range. |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: The exact sample size for the retrospective analysis of the STRATIS registry data is not explicitly stated in the provided text. It mentions "a retrospective subgroup analysis."
- Data Provenance: The data is from the STRATIS registry, described as a retrospective analysis. The country of origin is not specified, but STRATIS (Stroke TreAtment with a Solitaire stent-retriever and Intravenous t-PA) is an international registry, so the data likely encompasses multiple countries.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the text. The STRATIS registry would have had various clinical endpoints and adjudication processes, but the details of experts establishing ground truth for the specific retrospective subgroup analysis are not described.
4. Adjudication method for the test set:
This information is not provided in the text.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study was performed. The device is a mechanical thrombectomy device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
No standalone algorithm performance was done as this is a medical device (thrombectomy device), not an imaging or diagnostic algorithm.
7. The type of ground truth used:
For the clinical performance, the ground truth would be based on patient outcomes data collected as part of the STRATIS registry (e.g., successful reperfusion rates, clinical disability scores, adverse event rates), as adjudicated in the original registry study design. For the bench tests, the ground truth is based on engineering measurements against predefined specifications.
8. The sample size for the training set:
- Training Set Sample Size: The concept of a "training set" in the context of device approval (especially for a physical medical device and a labeling modification) is not applicable in the same way it would be for an AI/ML algorithm. The device design and previous iterations would have been "trained" through extensive R&D and prior clinical trials for predicate devices, but there isn't a "training set" in the computational sense.
- The "retrospective analysis of subject vessel size was performed using the STRATIS registry data" to support the labeling modification, implying that this existing clinical data was used for validation rather than for "training" a new device or algorithm.
9. How the ground truth for the training set was established:
As above, the concept of a "training set" with ground truth establishment in the AI/ML sense is not applicable for this device submission. The existing clinical data from the STRATIS registry, which includes patient outcomes, would serve as the "ground truth" for evaluating the clinical performance of the device in smaller vessels. The establishment of this ground truth would have been defined by the STRATIS registry protocol, including clinical assessments and imaging.
Ask a specific question about this device
(366 days)
The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The subject 6-40-10 Solitaire™ Platinum Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ Platinum Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Platinum Revascularization Device facilitates clot retrieval. The Solitaire™ Platinum Revascularization Device has Platinum Iridium radiopaque markers on the working cell length, proximal and distal ends. The subject Solitaire™ Platinum Revascularization Device is a portfolio expansion to the predicate Solitaire™ Platinum Revascularization Device (K153071).
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device: Solitaire™ Platinum Revascularization Device (6x40 mm)
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes both biocompatibility testing and non-clinical bench testing. There is also clinical data leveraged from a registry.
Biocompatibility Acceptance Criteria & Performance:
| Test Category | Test Description | Method | Acceptance Criteria | Reported Device Performance (Conclusion) |
|---|---|---|---|---|
| Cytotoxicity | L929 MTT Cytotoxicity | ISO 10993-5 | Viability is $\geq$ 70%. | Acceptance criteria met |
| Sensitization | Guinea Pig Maximization Sensitization | ISO 10993-10 | Test article does not elicit a sensitization response. | Acceptance criteria met |
| Irritation | Intracutaneous Irritation Test | ISO 10993-10 | Differences in the mean test and control scores of the extract dermal observations are $<$ 1.0. | Acceptance criteria met |
| Acute Systemic Toxicity | Acute Systemic Injection Test | ISO 10993-11 | No abnormal clinical signs and weight loss in excess of 10%. Temperature rise $\geq$ 0.5°C | Acceptance criteria met |
| Materials Mediated Rabbit Pyrogen | (Not specified) | (Not specified) | Acceptance criteria met | |
| Hemo-compatibility | Hemolysis | ISO 10993-4 | No significant differences between the test article extract and negative control article results. The test article is considered non-hemolytic | Acceptance criteria met |
| Partial Thromboplastin Time | (Not specified) | Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway. | Acceptance criteria met | |
| Platelet and Leukocyte Count | (Not specified) | Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material. | Acceptance criteria met | |
| Complement Activation C3a and SC5b-9 Assay | (Not specified) | Levels of C3a and SC5b-9 are comparable and less than the positive control. | Acceptance criteria met | |
| Thrombosis | (Not specified) | Thrombo-resistance properties are acceptable in clinical use. | Acceptance criteria met | |
| Genotoxicity | Bacterial Mutagenicity Test | ISO 10993-3 | Test article is considered non-mutagenic | Acceptance criteria met |
| In-vitro Mouse Lymphoma Assay | (Not specified) | Test article is considered non-mutagenic | Acceptance criteria met | |
| In-vivo Mouse Micronucleus Assay | (Not specified) | Test article is considered non-mutagenic | Acceptance criteria met |
Non-Clinical Bench Testing Acceptance Criteria & Performance:
| Test Description | Method | Acceptance Criteria | Reported Device Performance (Conclusion) |
|---|---|---|---|
| Total System Length | Total system length was measured from the distal tip of the distal marker to the proximal tip of the delivery system. | The stent length should be adequate for its intended use for delivery via standard microcatheters. | Acceptance criteria met |
| Marker-to-Marker Length | Marker-to-marker length is representative of the total stent length. Total stent length was measured 100% in process. | The stent length should be adequate for its intended use for delivery via standard microcatheters. | Acceptance criteria met |
| Delivery Force | Delivery force was measured through a representative tortuous anatomical model. | The stent must be below delivery force specification. | Acceptance criteria met |
| Resheathing Force | Resheathing force was measured through a representative tortuous anatomical model. | The stent must be below re-sheathing force specification. | Acceptance criteria met |
| Multiple Resheathing Durability | Multiple resheathing durability was evaluated on the ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use. | The stent must be able to be deployed and resheathed up to four times. | Acceptance criteria met |
| Body Finger Marker Coil Tensile | Body finger marker coil tensile was performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol body finger. | The body finger marker coil tensile should be greater than or equal to existing tensile strength specification. | Acceptance criteria met |
| Radial Force | Radial force was measured 100% in process. | The stent must be within existing radial force specification. | Acceptance criteria met |
| Radiopacity | Radiopacity was verified by both a material and dimensional analyses. | The radiopaque body markers must be visible using standard catheter laboratory equipment. | Acceptance criteria met |
| Ease of Use (Fluorosafe Marker Location) | Ease of use (fluorosafe marker location) was measured from the distal tip of the distal marker to the distal end of the fluorosafe marker. | Less required fluoroscopy is better than more required fluoroscopy. | Acceptance criteria met |
| Particulate | Particulate was evaluated for generation under simulated use in a representative tortuous anatomical model. | The stent must not generate clinically unacceptable particulate. | Acceptance criteria met |
Clinical Performance (Leveraged from STRATIS Registry 4x40 Subgroup):
| Confirmatory Clinical Data | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Safety Data: | ||
| Symptomatic ICH | As low as reasonably achievable, comparable to predicate | 2.8% (7/252) |
| Mortality at 90 days | As low as reasonably achievable, comparable to predicate | 16.2% (48/296) |
| Study Device-Related SAEs | As low as reasonably achievable, comparable to predicate | 0.3% (1/296) |
| Performance Data: | ||
| Successful revascularization (TICI 2b-3) | High success rate, comparable to predicate | 90.4% (227/251) |
| Functional Independence (mRS 0-2) at 90 days | High rate, comparable to predicate | 60.3% (167/277) |
2. Sample Size Used for the Test Set and Data Provenance
- Non-Clinical Animal Testing: The study used an "in-vivo Yorkshire cross swine model." The number of animals is not explicitly stated, but it evaluated 4 out of 8 proposed target sites (Right Vertebral, Left Vertebral, Right Internal Thoracic, Left Internal Thoracic, Right Costocervical, Left Costocervical, Renal, Ascending Pharyngeal).
- Provenance: Prospective, animal model.
- Confirmatory Clinical Data (Leveraged from STRATIS Registry):
- Sample Size: A total of 984 patients were enrolled in the STRATIS Registry. From this, the analysis for the 4x40 Solitaire devices (predicate and reference) included 296 patients who were treated only with these devices.
- Provenance: Prospective, multi-center, observational registry conducted in the United States (55 sites).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Animal Testing: The document does not specify the number of experts or their qualifications for evaluating the animal testing results (e.g., Luminal Thrombus, Endothelial Cell Coverage, etc.).
- Clinical Data (STRATIS Registry):
- Clinical Events Committee: An independent Clinical Events Committee was used for adverse event adjudication. The number of experts or their specific qualifications (e.g., type of physician, years of experience) are not detailed.
- Core Lab: An independent Core Lab was used for procedural image assessment. The number of experts or their specific qualifications are not detailed.
4. Adjudication Method for the Test Set
- Clinical Data (STRATIS Registry): The document states that an "independent Clinical Events Committee for adverse event adjudication and an independent Core Lab for procedural image assessment" were used. This indicates an independent adjudication process, but the specific method (e.g., 2+1, 3+1 consensus) is not detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was not done as described for AI assistance. The clinical data leveraged was from an observational registry evaluating clinical outcomes of the device in humans. This study was not comparing human readers with and without AI assistance; rather, it was assessing the device's performance in a real-world clinical setting.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Not applicable. This document describes a medical device (revascularization device), not an algorithm or AI software. Therefore, there is no "standalone" algorithm-only performance to report. The performance discussed for the clinical data is the device's performance in situ within a human patient, with clinicians performing the procedure.
7. The Type of Ground Truth Used
- Biocompatibility: In vitro and in vivo laboratory test results (e.g., cell viability, immune responses, clotting times, mutagenicity).
- Bench Testing: Engineering measurements and evaluations against design specifications.
- Animal Testing: Histopathological and physiological observations in an animal model (e.g., luminal thrombus, endothelial cell coverage, inflammation).
- Clinical Data (STRATIS Registry):
- Adverse Events: Adjudicated by an independent Clinical Events Committee, likely based on medical records, clinical observations, and definitions of adverse events.
- Procedural Image Assessment: Adjudicated by an independent Core Lab, likely measuring revascularization success (TICI scores) from imaging data.
- Functional Independence (mRS 0-2): Patient outcomes measured by modified Rankin Scale (mRS) at 90 days.
8. The Sample Size for the Training Set
- Not applicable. For this type of medical device (physical revascularization device), there is no "training set" in the context of an AI/ML algorithm. The "training set" concept is relevant for data-driven models that learn from data.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the provided document.
Ask a specific question about this device
(75 days)
The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The subject 4-20-05 and 6-24-06 Solitaire™ Platinum Revascularization Devices are designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The device is designed for use in the neurovasculature such as the internal carotid artery. M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries.
The provided text describes the Solitaire™ Platinum Revascularization Device and its performance data to establish substantial equivalence with a predicate device. The information primarily focuses on bench testing and biocompatibility rather than clinical studies involving human or animal subjects for the new devices (4-20-05 and 6-24-06 Solitaire™ Platinum Revascularization Devices).
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Note: The table below combines information from multiple sections of the document to present the acceptance criteria and the conclusion for each test. The "Reported Device Performance" column reflects the conclusion stated in the document, which consistently indicates that the device met the acceptance criteria and is substantially equivalent to the predicate.
| Test Category / Description | Method | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Biocompatibility | |||
| Cytotoxicity | ISO 10993-5 | Viability is ≥70%. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-5 and are substantially equivalent to the predicate device. |
| Sensitization | ISO 10993-10 | Test article does not elicit a sensitization response. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-10 and are substantially equivalent to the predicate device. |
| Irritation | ISO 10993-10 | Differences in the mean test and control scores of the extract dermal observations are < 1.0. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-10 and are substantially equivalent to the predicate device. |
| Acute Systemic Toxicity | ISO 10993-11 | No abnormal clinical signs and weight loss in excess of 10%. Temperature rise ≥0.5°C | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-11 and are substantially equivalent to the predicate device. |
| Hemolysis | ISO 10993-4 | No significant differences between the test article extract and negative control article results. The test article is considered non-hemolytic. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-4 and are substantially equivalent to the predicate device. |
| Partial Thromboplastin Time | ISO 10993-4 | Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-4 and are substantially equivalent to the predicate device. |
| Platelet and Leukocyte Count | ISO 10993-4 | Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-4 and are substantially equivalent to the predicate device. |
| Complement Activation C3a and SC5b-9 Assay | ISO 10993-4 | Levels of C3a and SC5b-9 are comparable and less than the positive control. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-4 and are substantially equivalent to the predicate device. |
| Thrombosis | ISO 10993-4 | Thrombo-resistance properties are acceptable in clinical use. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-4 and are substantially equivalent to the predicate device. |
| Bacterial Mutagenicity Test | ISO 10993-3 | Test article is considered non-mutagenic. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-3 and are substantially equivalent to the predicate device. |
| In-vitro Mouse Lymphoma Assay | ISO 10993-3 | Test article is considered non-mutagenic. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-3 and are substantially equivalent to the predicate device. |
| In-vivo Mouse Micronucleus Assay | ISO 10993-3 | Test article is considered non-mutagenic. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for ISO 10993-3 and are substantially equivalent to the predicate device. |
| Bench Testing | |||
| Delivery Force | Measurement through representative tortuous anatomical model. | Stent must be below delivery force specification. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for delivery force and are substantially equivalent to the predicate device. |
| Re-sheathing Force | Measurement through representative tortuous anatomical model. | Stent must be below re-sheathing force specification. | The subject 4-20-05 and 6-24-06 devices met the acceptance criteria for re-sheathing force and are substantially equivalent to the predicate device. |
| Multiple Re-sheathing Durability | Evaluation in a representative tortuous model beyond recommended passes/re-sheathings. | The subject 4-20-05 and 6-24-06 devices must reliably deploy and resheath up to four (4) times. | The subject 4-20-05 and 6-24-06 devices showed no irregularities, breaks, kinks, body marker migration, glue separations, or other observed defects after all deployment and re-sheathing attempts. The devices met the acceptance criteria and are substantially equivalent to the predicate device. |
| Body Marker Tensile | Testing of laser weld strength of Platinum/Iridium markercoil to Nitinol distal finger. | Body marker should be greater than or equal to existing tensile strength specification. | The subject 4-20-05 and 6-24-06 devices met acceptance criteria for body marker tensile and are substantially equivalent to the predicate device. |
| Body Marker Radiopacity | Verification analysis of body markers. | The radiopaque body markers must be visible using standard catheter laboratory equipment. | The subject 4-20-05 and 6-24-06 devices met acceptance criteria for body marker radiopacity and are substantially equivalent to the predicate device. |
| Proximal Marker to Distal Marker | 100% in-process measurement of laser cut and electro-polished stent length. | Length of stent must meet all inspection criteria. | The subject 4-20-05 and 6-24-06 devices met acceptance criteria for proximal marker to distal marker and are substantially equivalent to the predicate device. |
| Af Temperature | 100% in-process tracking of heat set parameters. | Temperature should be less than or equal to existing Af temperature specification. | The subject 4-20-05 and 6-24-06 Solitaire™ Platinum Revascularization Devices met acceptance criteria for Af and are substantially equivalent to the predicate device. |
| Radial Force | 100% in-process measurement. | Stent must be within existing radial force specification. | Radial force was measured 100% in-process. The subject 4-20-05 and 6-24-06 devices met acceptance criteria for radial force and are substantially equivalent to the predicate device. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document does not explicitly state numerical sample sizes for most of the bench tests (e.g., how many devices were tested for delivery force). For "Multiple Re-sheathing Durability," it implies that devices were tested for "up to four (4) times" deployment and re-sheathing, but not the number of physical devices sampled. For "Proximal Marker to Distal Marker" and "Radial Force," it states "100% in-process" measurement, which implies all manufactured units were tested. For biocompatibility tests, details on sample sizes are not provided within this summary sheet.
- Data Provenance: The data is from "bench testing" and "biocompatibility" studies, primarily conducted in a laboratory setting. There is no information regarding the country of origin of the data. The data is prospective as it was generated specifically for the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable. The tests described are primarily physical/mechanical bench tests and in-vitro/in-vivo (animal for biocompatibility) biocompatibility tests. "Ground truth" in the context of expert consensus (like for medical image interpretation) is not relevant for these types of engineering and biological safety evaluations. The acceptance criteria are based on established standards (ISO) and internal specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. The tests are directly measured against predefined numerical or qualitative acceptance criteria. There is no expert adjudication process for the results of these bench or biocompatibility tests mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not done. This document pertains to the regulatory submission for a physical medical device (revascularization device) and focuses on engineering performance and biocompatibility, not on AI-assisted diagnostic or interpretative tasks typically assessed by MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
A standalone performance study of an algorithm was not done. This device is a physical medical device, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the bench tests are the established engineering specifications and physical properties of the device, as measured by standard laboratory methods. For biocompatibility, the ground truth is based on biological safety standards (e.g., ISO 10993 series) which define acceptable biological responses. There is no "expert consensus," "pathology," or "outcomes data" specifically mentioned as ground truth for these specific validation tests of the new devices, as the focus is on physical and material properties, and the new devices are considered substantially equivalent to a predicate.
8. The sample size for the training set
This is not applicable. The document describes the performance testing of a physical medical device (Solitaire™ Platinum Revascularization Device), not a machine learning or AI algorithm that requires a "training set."
9. How the ground truth for the training set was established
This is not applicable, as there is no training set for this type of device submission.
Ask a specific question about this device
(62 days)
The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The Solitaire™ Platinum device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The device is designed for use in the neurovasculature such as the internal carotid artery, M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the device facilitates clot retrieval and has Pt-Ir radiopaque markers on the working cell length, proximal and distal ends. The Solitaire™ Platinum device is a modification to the currently cleared Solitaire™ 2 Revascularization Device to increase the working length radiopacity.
The provided text is a 510(k) summary for the Solitaire™ Platinum Revascularization Device. It describes the device, its indications for use, and how it was deemed substantially equivalent to predicate devices. However, this document does not contain information about an AI/ML-based device, nor does it detail acceptance criteria and a study proving an AI/ML device meets those criteria.
The information requested in your prompt (acceptance criteria table, sample sizes for test/training sets, expert involvement, adjudication methods, MRMC studies, standalone performance, ground truth types, etc.) is specific to the development and validation of AI/ML medical devices. The Solitaire™ Platinum Revascularization Device is a physical medical device (a thrombectomy device), and its evaluation primarily involved bench testing, animal usability testing, and comparison to predicate physical devices.
Therefore, I cannot fulfill your request to create:
- A table of acceptance criteria and reported device performance for an AI/ML device, as this document does not describe AI/ML device performance. The listed "performance data" refers to mechanical properties and visibility of a physical device.
- Sample sizes used for the test set and data provenance related to an AI/ML model, because no AI/ML model is discussed.
- Number of experts used to establish ground truth or their qualifications.
- Adjudication method for a test set.
- MRMC comparative effectiveness study results or effect size for human readers with/without AI assistance.
- Standalone (algorithm only) performance data.
- Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for an AI/ML model.
- Sample size for the training set for an AI/ML model.
- How ground truth for the training set was established for an AI/ML model.
The document explicitly states: "No animal testing for safety and efficacy or clinical studies were performed as there is no change to the indications for use or the fundamental scientific technology of the device." and "No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the new model number." This further confirms the absence of typical AI/ML validation studies such as those involving human readers or large datasets for model training and testing.
In summary, the provided text does not contain the information necessary to describe the acceptance criteria and the study that proves an AI/ML device meets those criteria, as it pertains to a physical medical device.
Ask a specific question about this device
Page 1 of 1