Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153226
    Date Cleared
    2016-03-24

    (139 days)

    Product Code
    Regulation Number
    870.5900
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solex 7™ Intravascular Heat Exchange Catheter connected to the CoolGard 3000®/Thermal Regulation System is indicated for use:

    · In cardiac surgery patients to achieve and or maintain normothermia during surgery and recovery/intensive care. (Maximum use period = 4 days)

    · To induce, maintain and reverse mild hypothermia in neurosurgery and recovery/intensive care. (Maximum use period = 4 days)

    · In fever reduction, as an adjunct to other antipyretic therapy, in adult patients with cerebral infarcerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. (Maximum use period = 7 days)

    Device Description

    The Solex 7 Intravascular Heat Exchange Catheter (Solex 7 Catheter) is a sterile, single use 9.3F flexible catheter designed for placement in the Superior Vena Cava from an insertion site in the Jugular and Subclavian Veins. The Solex 7 Catheter is connected to a single use, disposable CoolGard 3000 or Thermogard XP® Start-Up Kit (SUK) and the CoolGard 30000 or Thermogard XP® Console, all of which comprise the ZOLL Intravascular Heat Exchange The Start-Up Kit (SUK) and the CoolGard 3000 or Thermogard XP Console are System. supplied separately. The ZOLL Heat Exchange System is also designed for use with an off-theshelf temperature probe. The Solex 7 Catheter is comprised of a polyurethane shaft and a serpentine shaped PET balloon at the distal end. The blood contact surfaces of the catheter incorporate a hydrophilic heparin coating.

    The catheter has five lumens, two of which when connected to the Start-Up Kit, are used to circulate sterile saline in a closed loop circuit for heat exchange with the blood in the central venous system. Warmed or chilled saline is pumped through the heat exchange lumens, inflating the diameter of the serpentine balloon that interfaces with the patient's blood to warm or cool circulating blood. The inflow/outflow lumens form a closed-loop system through which the warmed or chilled saline circulates. The warmed or chilled saline is not infused into the patient.

    Additional lumens of the Solex 7 Catheter consist of a 0.032" guidewire compatible lumen that can also be used as a primary infusion lumen, and two additional infusion lumens within the catheter shaft.

    AI/ML Overview

    The document provided describes the Solex 7™ Intravascular Heat Exchange Catheter and Start-Up Kit and its substantial equivalence to predicate devices, based on nonclinical testing. This is a 510(k) premarket notification, which demonstrates a device is at least as safe and effective as a legally marketed predicate device, rather than proving effectiveness through clinical trials required for a PMA.

    Therefore, the study described here is primarily a set of nonclinical tests to show that the modified device performs similarly to its predicate devices and meets its own design specifications.

    Here's an analysis of the acceptance criteria and the "study" (nonclinical testing) that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides multiple tables (Table 3, Table 4, Table 5, Table 6, Table 7) detailing various tests, their methods, and results. These tables effectively serve as the acceptance criteria and reported device performance. I will consolidate key examples below.

    Test CategoryAcceptance Criteria (Test Method Summary)Reported Device Performance (Results)
    Solex 7 Catheter Bench Performance
    Visual InspectionEnsure catheter is smooth without major pits.Catheter samples met the acceptance criterion for smooth appearance with no pits.
    Catheter Indwell Life (7 days)Verify no saline leakage during simulated use for seven days.Catheter samples met the acceptance criterion for normal function for seven days without leakage.
    Particulate TestingDetermine catheter coating integrity following simulated indwell and flow rate conditions.Solex 7 samples and Cool Line samples tested demonstrated levels of particulate below that required in USP <788>.
    Heat Exchange TestingDetermine heat exchange capability of catheter.Catheter samples met the acceptance criterion for a heat exchange power of 125 Watts minimum.
    In Vitro Thrombogenicity TestingMeasurement of thrombogenic behavior of catheter following 7 day simulated use by radioactive platelet quantification and visual inspection.Coated catheter samples showed statistically significant decreases (p < 0.05) in thrombogenicity when compared to normalized thrombogenic behavior of uncoated catheters.
    Start-Up Kit (SUK) Packaging Testing
    SUK Package Seal Peel TestEnsure seal peel data is not less than specified acceptance criteria.SUK package seals met the acceptance criterion for peel strength.
    Start-Up Kit Performance Testing
    SUK Flow TestVerify that air flows through the SUK system.SUK systems met the acceptance criterion for air flow.
    Modified SUK Indwell Life TestVerify that Custom Luers and SUK Tubing are able to function normally at specified temperatures for the labeled indwell period.Custom Luers and SUK tubing met the acceptance criterion for normal function for 7 days at the specified temperature.
    Custom Luer Catheter Bench Performance
    Guidewire PassageVerify ability to frontload, backload, and remove J-tip 0.032" guidewire from catheter.Catheters met the acceptance criterion for the J-tip 0.032" guidewire to be frontloaded, backloaded, and removed from the catheters.
    Balloon Deflation Using Deflation Syringe (20 mL Slip-Fit)Verify that balloons collapse upon aspiration at specified temperature using supplied slip-fit syringe.Catheter balloons met the acceptance criterion of collapse upon aspiration using the supplied slip fit syringe.
    Luer Standards Performance Testing
    Water/Liquid LeakageTested using test method similar to test specified in ISO594-1&2 (§4.2.1).All samples met the requirements of the test method used.
    Stress CrackingTested using test method similar to that specified in ISO594-2 (§4.7).All samples met the requirements of the test method used.
    Biocompatibility Testing
    CytotoxicityMEM Elution Test.The reactivity grade of all test article extract samples was “0”.
    Hemocompatibility (ASTM Hemolysis)ASTM Hemolysis (Extract Method) Test.The difference between the hemolytic indexes of the test article and the negative control equals 0.00 percent. This places the test article in the non-hemolytic range.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document generally refers to "samples" or "catheter samples" without specifying precise numerical sample sizes for most bench tests. However, for "Particulate Testing," it mentions "Solex 7 samples and Cool Line samples."

    • Sample Size: Not explicitly stated numerically for most tests. It refers to "catheter samples," "SUK systems," "all samples," etc.
    • Data Provenance: This is a nonclinical study conducted by ZOLL Circulation, Inc. (submitter). The tests are laboratory-based and simulated, rather than involving human subjects or real-world clinical data. Therefore, concepts like country of origin for data or retrospective/prospective don't directly apply in the same way they would for a clinical study. The tests were performed to verify performance against international standards (ISO, EN ISO, ASTM, USP).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable (N/A). The "ground truth" for these tests is defined by established engineering and medical device standards (e.g., ISO, ASTM, USP criteria for tensile strength, leakage, biocompatibility, etc.) and the device's own design specifications. Expert consensus in the traditional sense (like for medical image interpretation) is not directly involved in setting the pass/fail criteria for these bench tests. The human factors testing involved "users" but not experts establishing ground truth for objective performance metrics in the same way.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This section is N/A. Adjudication methods like 2+1 or 3+1 are used in clinical studies where subjective expert review is required (e.g., for ambiguous medical findings). For these nonclinical bench tests, results are typically objective measurements or observations against predefined pass/fail criteria outlined in the test methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is N/A. The document describes a medical device (intravascular heat exchange catheter and start-up kit), not an AI diagnostic or assistance system. Therefore, an MRMC comparative effectiveness study or assessment of human reader improvement with AI assistance is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is N/A. The device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on a combination of:

    • International Standards: ISO 10555-1, ISO 594-1, ISO 594-2, EN ISO 10555-1, USP <788>, ASTM Hemolysis standards.
    • Device Design Specifications: Specific dimensional tolerances, flow rates, heat exchange power, tensile strength, burst/leak resistance, indwell life, and biocompatibility criteria defined by the manufacturer for the device.
    • Human Factors Test Scenarios: User tasks designed to assess the ease and correctness of device usage as per the Instructions for Use (IFU).

    8. The sample size for the training set

    This section is N/A. This is not a machine learning or AI device, so there is no "training set" in the context of algorithm development.

    9. How the ground truth for the training set was established

    This section is N/A for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1