510(k) Data Aggregation
(251 days)
The Sleep Apnea Feature is an over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Watch and Phone.
This feature is intended to detect signs of moderate to severe obstructive sleep apnea in the form of significant breathing disruptions in adult users 22 years and older, over a twonight monitoring period. It is intended for on demand use.
This feature is not intended for users who have previously been diagnosed with sleep apnea. Users should not use this feature to replace traditional methods of diagnosis and treatment by a qualified clinician. The data provided by this device is also not intended to assist clinicians in diagnosing sleep disorders.
The Samsung Sleep Apnea Feature leverages wrist-worn PPG and actigraphy technology to create an over-the-counter (OTC) assessment of moderate-to-severe obstructive sleep annea for adults. When enabled, the device utilizes the wearable platform's PPG-derived SpO2 to monitor the user's sleep for repetitive, relative decreases in their blood oxygenation indicative of significant breathing disruptions associated with sleep apnea. Each on-demand assessment period requires two successful nights of observation within 10 days. After two qualifying assessment nights. the device will display the result on the wearable, after which, the user is guided to the phone for additional information. This provides the user with health information so that they may seek out medical attention. No raw signal data, including the SpO2 signal, is provided to the user nor is it able to be shared with clinicians.
The Samsung Sleep Apnea Feature consists of two mobile medical applications, one on the wearable (e.g., Samsung Galaxy Watch) and the other on the connected mobile phone (e.g., Samsung Galaxy Phone), both commercial off-the-shelf general computing platforms. Communications between the two devices are accomplished by encrypted Bluetooth/BLE connection via standard protocols for data transfer. The wearable component of the Sleep Apnea Feature runs in the wearable's operating system allowing it to verify the identification/qualification of the hardware, request SpO2/accelerometer signals via private APIs, display information on the screen display, and send data and receive commands to the phone Sleep Apnea Feature on the associated phone. The phone component of the Sleep Apnea feature provides a UI for onboarding, labeling, and status as well as the ability for device updates.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Sensitivity for detecting moderate-to-severe obstructive sleep apnea (AHI ≥ 15) | 82.7% (95% CI: [76.7%, 87.6%]) - Passed |
| Specificity for not detecting signs of moderate-to-severe obstructive sleep apnea (AHI < 15) | 87.7% (95% CI: [83.1%, 91.4%]) - Did Not Pass initial criterion. Modified Calculation (Post-hoc): 91.1% (95% LCB: 86.9%) - Passed after re-evaluation considering 10 previously undiagnosed subjects who were "false positives" but benefited from the device. |
| Software Validation (Moderate Level of Concern) | Appropriate documentation provided in accordance with FDA 2005 guidance. |
| Cybersecurity | Aligns with FDA 2014 guidance and conforms to Section 524B of the FD&C Act. |
| Non-clinical performance testing (hardware compatibility, input signal quality, handling noisy/missing data, poor signal quality) | Bench testing (SpO2 Data Integrity, Accelerometer Sensor Performance) and on-human testing (Sleep Time, On-human Sleep SpO2 Accuracy, On-human Sleep SpO2 Coverage, On-human Stationary SpO2 Accuracy, Low Perfusion) were submitted and confirmed platform capabilities and data quality. |
| Electromagnetic compatibility (EMC) and electrical, mechanical, and thermal safety of hardware components | Performance data provided (implied to have passed, as no issues were raised). |
| Biocompatibility of skin-contacting hardware components | Biocompatibility evaluation performed (implied to have passed, as no issues were raised). |
| Software Verification, Validation, and Hazard Analysis (including technical specifications, hardware characteristics, mitigation of failures) | Documentation provided and aligned with appropriate FDA guidance. |
| Human Factors and Usability Testing (correct use, interpretation of output, understanding when to seek medical care) | Self-Selection: 16/20 intended users correctly identified themselves; 4/4 non-intended users correctly identified themselves. Performance Testing: Completed 5 simulated use scenarios successfully. Knowledge Questions: Participants understood limitations and correct actions, despite one initial use error regarding contact with a doctor. Conclusion: Can be used safely and effectively. |
| Labeling (description, hardware/OS requirements, sensor data, warnings, interpretation, summary of clinical performance) | Labeling is sufficient and satisfies 21 CFR 801.109, including IUF, description, precautions, clinical data summary, AE list, and safe use instructions. Limitations section presented important contraindications, warnings, and precautions. |
Study Details
-
Sample size used for the test set and the data provenance:
- Total enrolled subjects: 620
- Subjects who completed the investigation: 573 (620 - 47 did not complete)
- Test set size for sensitivity/specificity calculation: 202 (True Positives + False Negatives) for sensitivity; 268 (True Negatives + False Positives) for specificity. The total for these calculations is 470 subjects.
- Data Provenance:
- Country of Origin: Not explicitly stated, but the sponsor information lists "Samsung Research America, Mountain View, CA 94043 USA," suggesting the study was conducted within the United States.
- Retrospective or Prospective: Prospective, as it states "A non-randomized, open-label, multi-center, single-blind study was conducted in an enriched adult population with an accredited sleep lab recruiting and enrolling subjects..."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The ground truth was established by "physician's assessment on corresponding PSG from an FDA-cleared PSG device as the clinical gold standard."
- The text does not specify the number of physicians or their individual qualifications (e.g., years of experience, specific board certifications), only that it was a "physician's assessment." The study was conducted in an "accredited sleep lab," implying appropriately qualified professionals.
-
Adjudication method for the test set:
- The text does not explicitly state a formal adjudication method like "2+1" or "3+1" for reading the PSG results to establish ground truth. It refers to "physician's assessment," which typically implies a clinical standard interpretation, likely by qualified sleep physicians within the accredited labs.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not conducted or reported. This study assessed the standalone performance of the AI device against a physician-interpreted PSG gold standard.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance study was done. The study "compared the result from the Samsung Sleep Apnea Feature, the Device Under Test (DUT), with physician's assessment on corresponding PSG from an FDA-cleared PSG device as the clinical gold standard." This indicates the device's algorithm generated results independently, which were then compared to the ground truth.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth used was physician's assessment on corresponding Polysomnography (PSG) results from an FDA-cleared PSG device, referred to as the "clinical gold standard." This falls under the category of "expert assessment" based on comprehensive physiological data.
-
The sample size for the training set:
- The development of the machine-learned algorithms utilized datasets from "over 1000 subjects, split into separate training, tuning, and testing datasets." The exact sample size specifically for the training set is not provided, only the total number of subjects for development phases.
-
How the ground truth for the training set was established:
- The text states that "datasets from representative populations were utilized from over 1000 subjects" for development, and these datasets were "maintained independently from the final verification and validation activities." It is implied that the ground truth for these development datasets would have also been established using PSG or similar gold standard methods, consistent with clinical practice for sleep apnea diagnosis, but the precise methodology for establishing ground truth specifically for the training set is not detailed.