510(k) Data Aggregation

    K Number
    K162784
    Device Name
    Silk'n HST
    Date Cleared
    2017-03-25

    (173 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Silk'n HST

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Silk'n HST is an over-the-counter home use device intended for non-invasive treatment of mild to moderate periorbital wrinkles for adult women who have Fitzpatrick skin types I-IV.

    Device Description

    The Silk'n HST device is designed as a hand held device, which utilizes both RF and low power light energies. The device includes RF generator, LEDs at wavelengths of 630±20nm and 850±20nm (red and IR wavelengths, respectively) and a temperature stabilizer. The Silk'n HST device consists of an applicator and an adaptor. The applicator is a handheld unit used for treatment. The treatment surface is located at the applicator tip and comes in direct contact with the skin. The applicator is equipped with an ON/OFF switch that also selects the energy level as manifested using indicators panel. The adaptor connects the applicator to the electrical outlet using its plug and the applicator socket.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device Name: Silk'n HST
    Intended Use: Over-the-counter home use device intended for non-invasive treatment of mild to moderate periorbital wrinkles for adult women who have Fitzpatrick skin types I-IV.

    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for effectiveness appears to be a statistically significant reduction of at least 1.0 score in the average score of the Fitzpatrick Wrinkle Severity Scale, according to at least 2 out of 3 blinded dermatologists, measured 3 months post-treatment end compared to baseline.

    Acceptance CriterionReported Device Performance
    Effectiveness:
    Statistically significant reduction of at least 1.0 score in average Fitzpatrick Wrinkle Severity Scale score.Achieved an average statistically significant reduction of 1.49 ± 0.51 Fitzpatrick scores (p