510(k) Data Aggregation

    K Number
    K251758
    Device Name
    Serafin®
    Manufacturer
    Date Cleared
    2025-09-11

    (94 days)

    Product Code
    Regulation Number
    872.5470
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    Serafin**®**

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K232074
    Device Name
    Serafin®
    Manufacturer
    Date Cleared
    2024-02-22

    (225 days)

    Product Code
    Regulation Number
    872.5470
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Serafin**®**

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serafin is intended for the orthodontic treatment of malocclusion.

    Device Description

    Serafin® is composed of series of transparent removable orthodontic appliances manufactured using thermoforming technology based on a 3D model of the patient's teeth, customized according to a Dental Practitioner's specific prescription, indicated for treatment of dental malocclusion by applying continuous gentle force to realign teeth.

    The time for each stage of aligner usage is set in the personal treatment plan, and this varies for each patient and the Dental Practitioner's choice. Each set of aligners needs to be worn for at least 2 weeks. Patients the total treatment time lasts about 6 to 24 months.

    AI/ML Overview

    The provided text is a 510(k) summary for the Serafin® orthodontic clear aligner system. It focuses on demonstrating substantial equivalence to a predicate device (Invisalign) through non-clinical testing. Crucially, this document does not contain information about clinical studies with human subjects, algorithm performance metrics (such as accuracy, sensitivity, or specificity), or the use of AI/machine learning.

    Therefore, I cannot provide details on the following aspects of your request:

    • A table of acceptance criteria and reported device performance related to AI/algorithm output (as no such data is present).
    • Sample size used for a test set for AI/algorithm performance.
    • Number of experts used to establish ground truth for a test set.
    • Adjudication method for a test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone (algorithm only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for AI/algorithm performance.
    • Sample size for the training set for an AI/algorithm.
    • How the ground truth for the training set was established for an AI/algorithm.

    The document focuses on the physical, chemical, and biological properties of the device itself (the clear aligner), not on computational performance or diagnostic capabilities associated with AI.

    Here's what the document does provide regarding acceptance criteria and testing, albeit for the physical device, not an AI algorithm:

    1. Acceptance Criteria and Device Performance (for the physical device):

    Acceptance Criteria (Set Test Criteria)Reported Device Performance
    Physical and Chemical Tests:All test results met the preset test criteria, except as noted for Flexural Modulus.
    AppearanceConformed to standard
    DimensionConformed to standard
    Water sorption & solubilityLow water uptake, Low water solubility (conformed to standard)
    3-point flexural strength (Flexural Modulus, Flexural Strength, Tear Strength)Flexural Modulus did not meet standard requirement of 1500 MPa. The document states this requirement is not applicable as the device is designed for subtle force, not rigidity. Flexural Strength and Tear Strength conformed to standard.
    HardnessConformed to standard
    pHConformed to standard
    Potassium permanganate reducing substances residueConformed to standard
    Residue on evaporationConformed to standard
    Heavy metalsConformed to standard
    Ultraviolet-visible spectrumConformed to standard
    Biological Evaluation:Testing demonstrated biocompatibility.
    CytotoxicityConformed to standard
    Skin sensitizationConformed to standard
    Acute oral mucosa irritationConformed to standard
    Intracutaneous reactivityConformed to standard
    Acute systemic toxicityConformed to standard
    IrritationConformed to standard
    Biological risk assessment (ISO 10993-1, ISO 7405)Conducted
    Toxicological risk assessment (for leachables/degradation products)Conducted to monitor and mitigate biological risk.

    2. Sample size used for the test set and the data provenance:

    • No information on a "test set" for an AI algorithm.
    • The non-clinical testing was performed on "test articles" (implied to be samples of the Serafin® aligner material/device). The exact number of samples is not stated.
    • Data Provenance: The testing was done by the manufacturer (TNS Co., Ltd., based in South Korea) or their designated testing facilities. It's non-clinical lab testing of material properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as the testing is for physical/chemical/biological properties of the device material, not for an AI algorithm requiring expert-established ground truth. Standards (e.g., ISO, MFDS, ADA ANSI) define the test procedures and acceptance criteria.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This concept applies to human reader studies or AI validation with human adjudicated ground truth, which is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done, as this submission is for a physical medical device (orthodontic aligner), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable, as there is no AI algorithm being evaluated in this submission. The "3D software" mentioned is for generating a 3D model of teeth and treatment plan, which is then reviewed by a dental practitioner, not for AI-driven diagnosis or analysis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the physical device testing, the "ground truth" or reference was established by international and national standards (ISO 20795-2, ISO 10993 series, MFDS guidelines, ADA ANSI 41-2020) for material properties (e.g., flexural strength, water sorption) and biocompatibility (e.g., cytotoxicity).

    8. The sample size for the training set:

    • Not applicable, as there is no AI algorithm with a distinct "training set" described in this document.

    9. How the ground truth for the training set was established:

    • Not applicable.