510(k) Data Aggregation

    K Number
    K191915
    Date Cleared
    2020-03-06

    (233 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This surgeon's glove is a device made of synthetic rubber latex intended to be worn by surgeons and or operating room personnel to protect a surgical wound from contamination, and are tested for use with chemotherapy drugs.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification for a surgical glove, not an AI-powered diagnostic device. Therefore, the requested information about acceptance criteria, study design for AI evaluation (sample size, data provenance, expert ground truth, MRMC study, standalone performance, training set details), and adjudication methods is not applicable to this document.

    The document describes the performance of the SensiCare PI Sterile Powder-Free Polymer Coated Polyisoprene Surgical Glove, Coated with Aloe Vera, specifically its resistance to chemotherapy drug permeation.

    Here's how the provided information relates to the "device" (the surgical glove) under review:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for these gloves, as relevant to the provided data, would implicitly be that the breakthrough detection time (permeation time) for chemotherapy drugs should be sufficiently long to provide protection, ideally exceeding 240 minutes for most drugs. The device performance is reported directly in the tables on pages 3 and 5.

    Acceptance Criteria & Reported Performance for SensiCare PI Sterile Surgical Gloves (Natural and Green Colors)

    Chemotherapy DrugConcentrationAcceptance Criteria (Implied)Reported Performance (Breakthrough Detection Time in Minutes) - Natural Color (Page 3)Reported Performance (Breakthrough Detection Time in Minutes) - Green Color (Page 5)
    Bleomycin15 mg/ml>240 minutes>240 minutes>240 minutes
    Busulfan6 mg/ml>240 minutes>240 minutes>240 minutes
    Carboplatin10.0 mg/ml>240 minutes>240 minutes>240 minutes
    Carmustine (BCNU)3.3 mg/mlLonger time preferred22.5 (23.4, 23.8, 22.5) minutes13.7 (14.4, 14.4, 13.7) minutes
    Cisplatin1.0 mg/ml>240 minutes>240 minutes>240 minutes
    Cyclophosphamide (Cytoxan)20 mg/ml>240 minutes>240 minutes>240 minutes
    Dacarbazine (DTIC)10.0 mg/ml>240 minutes>240 minutes>240 minutes
    Doxorubicin Hydrochloride2.0 mg/ml>240 minutes>240 minutes>240 minutes
    Epirubicin (Ellence)2 mg/ml>240 minutes>240 minutes>240 minutes
    Etoposide (Toposar)20.0 mg/ml>240 minutes>240 minutes>240 minutes
    Fludarabine25.0 mg/ml>240 minutes>240 minutes>240 minutes
    Fluorouracil50.0 mg/ml>240 minutes>240 minutes>240 minutes
    Idarubicin1.0 mg/ml>240 minutes>240 minutes>240 minutes
    Ifosfamide50.0 mg/ml>240 minutes>240 minutes>240 minutes
    Mechlorethamine HCI1.0 mg/ml>240 minutes>240 minutes>240 minutes
    Melphalan5 mg/ml>240 minutes>240 minutes>240 minutes
    Methotrexate25 mg/ml>240 minutes>240 minutes>240 minutes
    Mitomycin C0.5 mg/ml>240 minutes>240 minutes>240 minutes
    Mitoxantrone2.0 mg/ml>240 minutes>240 minutes>240 minutes
    Paclitaxel (Taxol)6.0 mg/ml>240 minutes>240 minutes>240 minutes
    Paraplatin10 mg/ml>240 minutes>240 minutes>240 minutes
    Rituximab10 mg/ml>240 minutes>240 minutes>240 minutes
    Thiotepa10.0 mg/mlLonger time preferred24.4 (25.3, 24.4, 25.6) minutes13.8 (13.8, 16.4, 14.8) minutes
    Vincristine Sulfate1.0 mg/ml>240 minutes>240 minutes>240 minutes

    Warning: The document explicitly states "Warning: Do not use with Carmustine and Thiotepa" for both glove types due to their limited breakthrough times.

    The remaining points (2-9) are not applicable as this is not an AI/software device:

    1. Sample sized used for the test set and the data provenance: Not an AI study. This relates to physical testing of gloves for chemical permeation. The standard for such testing would typically involve multiple glove samples to ensure reproducibility and statistical validity, but the specific sample size for each drug test is not provided in this regulatory letter.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for chemical permeation is established through laboratory testing, not expert consensus (e.g., using analytical chemistry methods to detect drug breakthrough).
    3. Adjudication method: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This is for evaluating changes in human reader performance with AI.
    5. If a standalone was done: Not applicable. This refers to algorithm-only performance.
    6. The type of ground truth used: For these gloves, the "ground truth" is the measured breakthrough time of the chemotherapy drugs through the glove material, determined by standardized laboratory methods (e.g., ASTM F739 or similar).
    7. The sample size for the training set: Not applicable. This is not an AI device.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document is a regulatory letter for a medical device (surgical gloves) and details its performance against specific chemical permeation, which are the relevant "acceptance criteria" and "study results" for this type of product. The questions posed are designed for AI-driven medical devices and thus, are largely not relevant to this document.