510(k) Data Aggregation

    K Number
    K202188
    Date Cleared
    2021-02-19

    (199 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Safety Blood Collection Sets for Single Use

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Safety Blood Collection Sets for Single Use and blood collection tube/syringe are used together for the collection of venous blood. The winged needle is designed with a safety shield to cover the needle immediately following blood collection to aid in the protection against accidental needlestick injury.

    Device Description

    The Safety Blood Collection Sets for Single Use are used in routine venipuncture procedures. The winged needle is designed with a safety shield, which can be activated to cover the needle immediately the following venipuncture to aid in the protection against accidental needlestick injury. The Safety Blood Collect Sets is compatible for use with a tube and syringe. The product is to be used by appropriately trained healthcare professionals, only following the product instructions. The Blood Collection needle is manufactured from tubular stainless steel sharpened at both ends that is attached to the hub. The hub is threaded on one side to connect to the needle holder, which is used to guide the needle into an evacuated blood collection tube. This end of the needle is the shorter end and is fitted with a protective rubber cap and a needle holder. The opposite end of the needle is 3/4" long for withdrawing blood and is fitted with a needle sheath. The needle holder and needle sheath protect the needle. The safety feature is operated through the release of a latch mechanism whereby the user slides a winged cover over the needle, as it is removed from the patient. Once the needle is covered, the safety cover locks in place. This mechanism is substantially equivalent to the predicate device. The proposed devices are packaged as sterile, single-use, and single patient use only. The proposed device consists of ten components: 1. needle holder 2. rubber sleeve 3. puncture needle 4. needle hub (male luer lock connector) 5. connect base (female luer lock connector) 6. flexing tube 7. safety shield 8. wings 9. patient-end tube needle 10. needle sheath

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Safety Blood Collection Sets for Single Use). It details the device's characteristics, comparison to a predicate device, and various performance, biocompatibility, and sterility tests conducted. However, it does NOT describe a study that proves the device meets specific acceptance criteria in terms of AI or algorithm performance, nor does it involve human readers, expert consensus for ground truth, or an MRMC study.

    The document discusses "acceptance criteria" only in the context of the physical and biological testing of the blood collection set (e.g., sterilization, biocompatibility, mechanical performance). It explicitly states: "No clinical study is included in this submission." This type of 510(k) submission generally relies on demonstrable substantial equivalence to a legally marketed predicate device through non-clinical testing, rather than novel clinical performance studies involving human subjects or AI algorithms.

    Therefore, many of the requested bullet points, particularly those related to AI/algorithm performance, human readers, sample sizes for test/training sets for AI, ground truth establishment for AI, and MRMC studies, are not applicable to the content of this document.

    I will address the applicable parts based on the provided text, and explicitly state when information is not present or not relevant to the document's content.


    Description of Acceptance Criteria and Proving Device Meets Them

    The document describes the acceptance criteria in the context of various performance, biocompatibility, and sterility tests for a physical medical device (blood collection set). It does not pertain to the performance of an AI/algorithm-based device.

    1. A table of acceptance criteria and the reported device performance:

    The document states that "All of the pre-determined acceptance criteria were met" for various tests, but it does not provide a specific table with the acceptance criteria values alongside numerical performance results for the device. It generally states compliance with relevant ISO and ASTM standards.

    Here's a summary of the stated compliance:

    Test CategoryStandards Complied WithReported Device Performance
    Performance TestingISO 9626 (Stainless steel needle tubing), ISO 7864 (Sterile hypodermic needles), ISO 80369-7 (Small-bore connectors for intravascular/hypodermic applications), ISO 80369-20 (Common test methods for small-bore connectors), ISO 23908 (Sharps injury protection), ISO 6009 (Color coding for identification)Complies with all noted standards. (e.g., Hub/Needle bond strength for 21G > 44N, complying with ISO7864)
    BiocompatibilityISO10993-1 (Biological evaluation of medical devices, specifically for externally communicating devices, contact circulating blood for