Search Results
Found 1 results
510(k) Data Aggregation
(43 days)
The LITHOCLAST® MASTER Handpiece and the 3.2 mm probe have the same intended use as their predicate accessories, that is, for the fragmentation of urinary tract calculi, including renal, ureteral and bladder stones, through rigid or The 3.2 mm probe is specifically indicated for use in semirigid endoscopes. fragmenting large stones in the bladder and kidney.
MASTER Handpiece: The MASTER Handpiece is 18 mm in diameter, 228 mm long, and is constructed of aluminum alloy. The proximal end of the handpiece is connected to the LITHOCLAST pressure regulator. The chosen treatment probe is attached to the distal end of the handpiece using a screw cap. The metal projectile inside the handpiece is driven forward by the pressure from the pressure regulator. The energy from the projectile is transmitted to the probe by a shock transmitter at the distal end of the handpiece. The handpiece is sealed at both the distal and proximal ends, which: (1) prolongs the life time of the handpiece by preventing dirt, moisture and other particles from entering the handpiece; and (2) allows for the changing of probes during a treatment without the risk of breaking sterility.
3.2 mm Probe: The 3.2 mm probe is a rigid Type 304 surgical grade stainless steel rod that acts to couple the shockwave from the handpiece to the target stone. The probe is reusable and may be sterilized by steam sterilization according to standard hospital procedures.
The provided text describes a 510(k) summary for the ELECTRO MEDICAL SYSTEMS SA LITHOCLAST® MASTER Handpiece and 3.2 mm Probe, seeking clearance based on substantial equivalence to predicate devices. The submission details various tests performed to support this claim for both the handpiece and the probe.
Here's an analysis of the acceptance criteria and study information, based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative acceptance criteria or detailed reported device performance in a table format. Instead, it describes various tests performed to demonstrate substantial equivalence to existing predicate devices. The implication is that the performance of the new device (MASTER Handpiece and 3.2 mm Probe) was considered acceptable if it performed comparably to, or improved upon, the predicate devices in the specific tests mentioned.
Here's an interpretation of the performance and implied criteria based on the text:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| MASTER Handpiece: | |
| Equivalent/Improved Lifetime | "Life time... tests were performed to support the substantial equivalence of the MASTER Handpiece." (The new sealed design is stated to "prolong the life time of the handpiece"). |
| Equivalent/Improved Probe Displacement | "Probe displacement tests were performed to support the substantial equivalence of the MASTER Handpiece." |
| Maintain Sterility during Probe Changes | The sealed design "allows for the changing of probes during a treatment without the risk of breaking sterility." (This is a design feature offering an advantage over the predicate, which required a silicone sleeve). |
| 3.2 mm Probe: | |
| Equivalent/Appropriate Probe Velocity | "Probe velocity... tests... were performed to support the substantial equivalence of the 3.2 mm probe." |
| Equivalent/Appropriate Probe Displacement | "Probe... displacement tests... were performed to support the substantial equivalence of the 3.2 mm probe." |
| Equivalent/Appropriate Impact Pressure | "Impact pressure... studies were performed to support the substantial equivalence of the 3.2 mm probe." |
| Equivalent/Appropriate Disintegration Efficiency | "Disintegration efficiency... studies were performed to support the substantial equivalence of the 3.2 mm probe." |
| Safe In-vitro Tissue Impact (No excessive collateral damage) | "in-vitro tissue impact studies were performed to support the substantial equivalence of the 3.2 mm probe." |
| Reusable and Sterilizable by Steam | "The probe is reusable and may be sterilized by steam sterilization according to standard hospital procedures." (Implied criterion: ability to withstand standard sterilization methods without degradation). |
| Efficacy for fragmenting large stones in bladder and kidney (New indication) | The 3.2 mm probe is "specifically indicated for use in fragmenting large stones in the bladder and kidney." The tests listed above collectively support the safety and effectiveness for this indication, comparing its performance to smaller diameter predicate probes. While not a direct "criterion" in the same way as a numerical threshold, demonstrating its suitability for this specific use is a key aspect of its clearance. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify sample sizes for any of the tests performed (lifetime, displacement, velocity, impact pressure, disintegration efficiency, in-vitro tissue impact studies).
The data provenance (country of origin, retrospective/prospective) is also not mentioned in this summary. It's likely these were in-house laboratory tests conducted by ELECTRO MEDICAL SYSTEMS SA, which is based in Nyon, Switzerland.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable to this submission. The tests described are bench or in-vitro performance studies, not clinical studies involving expert interpretation of patient data to establish ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable. No human interpretation or adjudication of results is mentioned for these performance tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no MRMC study mentioned, as this is a device for fragmenting kidney stones, not an AI-assisted diagnostic tool where human readers would be involved.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable in the context of an AI algorithm. The device itself (handpiece and probe) is the standalone entity, and its performance was evaluated in various bench and in-vitro tests without human assistance during the actual fragmentation process, although a surgeon would operate it.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the performance tests mentioned, the "ground truth" would be established by:
- Measurement of physical parameters: For "lifetime," "probe displacement," "probe velocity," and "impact pressure," the ground truth would be directly measured physical quantities using appropriate sensors and equipment.
- Quantitative results of efficacy: For "disintegration efficiency," the ground truth would be objectively measurable outcomes like the extent of stone fragmentation or reduction in stone mass.
- Observable effects: For "in-vitro tissue impact studies," the ground truth would involve observable or measurable effects on the tissue, often compared against a control or predicate device's effects.
8. The sample size for the training set
This is not applicable. This is a hardware medical device clearance, not an AI/machine learning application requiring a training set.
9. How the ground truth for the training set was established
This is not applicable for the same reason as above.
Ask a specific question about this device
Page 1 of 1