Found 1 results

510(k) Data Aggregation

    K Number
    K110118
    Date Cleared
    2011-02-17

    (30 days)

    Product Code
    Regulation Number
    868.2775
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    STIMTRODE SINGLE USE NERVE STIMULATOR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stimtrode™ is intended to be used by an anesthetist during a peripheral nerve block procedure for nerve localization using invasive electrodes/needles (supplied separately).

    Device Description

    The Stimtrode device is a battery powered peripheral nerve stimulators that can be used for

    • · nerve locating using invasive electrodes/needs (not supplied)
      The stimulus is generated by a constant current source. The waveform is a square wave with 3 options for pulse width. These are: 0.1, 0.3 and 0.5 milliseconds.
      The units will continuously check for a closed circuit at 2Hz. Once a closed circuit is detected, the twitches will repeat at 2Hz, until an open circuit is detected.
      The unit is permanently attached to the anode. The anode comprises a modified ECG type pad. The cathode comprises of a nerve locating needle that is attached at time of use.
    AI/ML Overview

    The provided text is a 510(k) summary for the Stimtrode Nerve Stimulator. It describes the device, its intended use, and its technological characteristics. However, it does not contain any information about acceptance criteria, a study proving device performance, sample sizes used, data provenance, ground truth establishment, or any comparative effectiveness studies.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Xavant Technology, XPOD/XMAP Nerve Stimulator K072092) and confirming its regulatory classification and intended use.