Search Results
Found 2 results
510(k) Data Aggregation
(84 days)
STERRAD(R) CYCLESURE(R) 24 BIOLOGICAL INDICATOR
The STERRAD® CYCLESURE® 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD® Sterilization Systems and Cycles:
MODEL | CYCLE |
---|---|
STERRAD® 100S | Standard |
STERRAD® 50 | Standard |
STERRAD® 200 | Standard |
STERRAD® NX® | Standard |
Advanced | |
STERRAD® 100NX® | Flex |
EXPRESS |
The STERRAD® CYCLESURE® 24 Biological Indicator is a self-contained standalone biological monitor intended for the routine monitoring of the STERRAD® Sterilization Process. It consists of a glass fiber disc containing a minimum of 1x10° Geobacillus stearothermophilus spores, a glass ampoule containing nutrient growth medium, a cap and liner closing the vial and a chemical indicator on top of the cap. The cap contains two small circular openings that allow for diffusion of hydrogen peroxide vapor into the vial. The relatively small size of the circular openings serves as a restriction to this diffusion.
Here's a breakdown of the acceptance criteria and study information for the STERRAD® CYCLESURE® 24 Biological Indicator, based on the provided 510(k) summary:
Device: STERRAD® CYCLESURE® 24 Biological Indicator
1. Acceptance Criteria and Reported Device Performance:
The document describes several studies performed to confirm the biological indicator performs as intended. The acceptance criteria for these tests are implicitly "Passed," indicating that the device met the required performance standards.
Study Performed | Acceptance Criteria | Reported Device Performance |
---|---|---|
Evaporation | Passed | Passed |
Verification of Positive BI Color | Passed | Passed |
Bacteriostasis | Passed | Passed |
BI Validation in STERRAD® Systems (Dose Response) | Passed | Passed |
Verification of Minimum Incubation Time | Passed | Passed |
Growth Promotion | Passed | Passed |
Spore Resistance Testing | Passed | Passed |
2. Sample Size and Data Provenance:
The document does not explicitly state the sample sizes used for each test. It mentions that "Testing was conducted" and "Performance testing demonstrated," but no specific numbers of biological indicators tested are provided.
The data provenance is implied to be prospective testing conducted by the applicant (Advanced Sterilization Products) as part of their submission for clearance of the modified device. The country of origin for the data is not specified, but the applicant's address is in Irvine, CA, USA.
3. Number and Qualifications of Experts for Ground Truth:
This information is not applicable to this type of device (biological indicator) and study. Biological indicators are designed to provide a direct readout (growth or no growth of spores) to indicate sterility. There is no human interpretation of images or complex data involved that would require expert consensus.
4. Adjudication Method:
Not applicable. See point 3. The results of the biological indicator (growth or no growth) are objective and do not require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret output with and without AI assistance to assess changes in accuracy or efficiency. A biological indicator functions as a direct sterility monitor and does not involve human interpretation in the same way.
6. Standalone Performance Study (Algorithm Only):
Yes, a standalone performance study was done in the sense that the biological indicator itself (the "algorithm" in this context) was tested independently to verify its function. The listed studies (Evaporation, Verification of Positive BI Color, Bacteriostasis, BI Validation, Minimum Incubation Time, Growth Promotion, Spore Resistance Testing) all assess the intrinsic performance of the biological indicator without human intervention or interpretation beyond observing the final growth result.
7. Type of Ground Truth Used:
The ground truth used for this device is based on microbiological principles and refers to the known outcome of spore killing under specific sterilization conditions.
- Positive Controls: Unsterilized or partially sterilized biological indicators are expected to show spore growth.
- Negative Controls: Properly sterilized biological indicators are expected to show no spore growth.
- D-value and Z-value parameters: For spore resistance testing, the ground truth relates to the known resistance of Geobacillus stearothermophilus spores to hydrogen peroxide under defined conditions.
- Growth Promotion: The ground truth is the ability of the medium to support the growth of the target microorganism.
8. Sample Size for the Training Set:
Not applicable. This device is not an AI/ML algorithm that learns from a training set. It is a physical biological indicator.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. See point 8.
Ask a specific question about this device
(487 days)
STERRAD (R) CYCLESURE (R) 24 BIOLOGICAL INDICATOR
The STERRAD® CYCLESURE® 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD® Sterilization Systems and Cycles:
- STERRAD® 100S
- STERRAD® 50
- STERRAD® 200
- STERRAD® NX®
- STERRAD® 100NX® (Standard, Flex and Express Cycles)
- For STERRAD® 100NX® DUO Cycle, the STERRAD® CYCLESURE® 24 Biological Indicator should only be used in a test pack configuration (REF 20243).
The STERRAD® CYCLESURE® 24 Biological Indicator is a self-contained standalone biological monitor intended for the routine monitoring of the STERRAD® Sterilization Process. It consists of a glass fiber disc containing a minimum of 1x10^6 Geobacillus stearothermophilus spores, a glass ampoule containing nutrient growth medium, a cap and liner closing the vial and a chemical indicator on top of the cap. The cap contains two small circular openings that allow for diffusion of hydrogen peroxide vapor into the vial. The relatively small size of the circular openings serves as a restriction to this diffusion. There are no changes to the STERRAD® CYCLESURE® 24 Biological Indicator from the device cleared in K102884. The change is to the labeling to add the STERRRAD® 100NX® DUO Cycle to the CYCLESURE® 24 Biological Indicator Indications for Use Statement.
Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study for the STERRAD® CYCLESURE® 24 Biological Indicator:
1. Table of Acceptance Criteria and Reported Device Performance
The summary states that the device "meets the same performance criteria as the predicate devices." While specific numerical acceptance criteria for each test (e.g., specific thresholds for reduction, growth rates, etc.) are not explicitly detailed in the provided text, the categories of performance criteria tested are:
Acceptance Criteria Category | Reported Device Performance |
---|---|
Evaporation | Passed |
Verification of Positive BI Color | Passed |
Bacteriostasis | Passed |
BI Validation in the STERRAD® Sterilization Systems (Dose Response) | Passed |
2. Sample Size Used for the Test Set and Data Provenance
The summary does not provide specific sample sizes for the test set for each of the studies listed. It states: "Therefore, only following performance data related to the DUO Cycle are included in this submission." This implies that the tests were conducted specifically for the DUO Cycle.
- Sample Size: Not specified.
- Data Provenance: The document implies the data was generated specifically for this submission to support the new DUO Cycle, so it would be prospective data, likely collected in a laboratory setting at Advanced Sterilization Products. The country of origin of the data is not explicitly stated, but given the applicant's address, it's highly probable to be United States.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of device (a biological indicator for sterilization) does not typically involve human expert interpretation in the same way as medical imaging devices. The "ground truth" for a biological indicator is based on microbiological culturing results (i.e., presence or absence of Geobacillus stearothermophilus growth), which is an objective outcome rather than a subjective interpretation. Therefore, the concept of "experts establishing ground truth" as it applies to subjective assessments is not relevant here.
4. Adjudication Method for the Test Set
As the "ground truth" for a biological indicator is objective microbiological growth, there is no adjudication method (like 2+1 or 3+1 consensus) required for the test set. The outcome is determined by scientific observation of bacterial growth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This is a biological indicator for monitoring sterilization processes, not a diagnostic imaging device that involves human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable or performed for this device.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, in a sense. The "device" itself (the biological indicator) is a standalone entity that provides an objective output (color change in the chemical indicator and potential microbiological growth). The performance studies evaluate the indicator's ability to accurately reflect sterilization efficacy without human intervention during the "reading" of the growth. The "algorithm" here is the biological process itself, not a computer algorithm.
7. Type of Ground Truth Used
The ground truth used for evaluating the performance of a biological indicator is microbiological growth/non-growth.
- For the "BI Validation in the STERRAD® Sterilization Systems (Dose Response)" test, the ground truth would be the known efficacy of the sterilization parameters (e.g., specific hydrogen peroxide exposure levels, time, temperature) to kill bacterial spores, and the subsequent observation of whether the Geobacillus stearothermophilus spores in the BI were killed or survived. This is essentially outcomes data in a controlled experimental setting.
- For "Verification of Positive BI Color" and "Bacteriostasis," the ground truth would be the expected visual change or lack of inhibitory effect on growth based on standard microbiological techniques.
8. Sample Size for the Training Set
The concept of a "training set" is not relevant for this device. Biological indicators do not use machine learning or AI algorithms that require training data. Their function is based on established biological principles of spore resistance and growth in response to sterilization, which are inherently fixed properties, not learned ones.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" for this biological indicator, this question is not applicable. The device's performance relies on its inherent biological properties and manufacturing consistency, not on an algorithm trained on data.
Ask a specific question about this device
Page 1 of 1