Search Results
Found 2 results
510(k) Data Aggregation
(133 days)
These powder-free sterile light brown colored surgeon's gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time inMinutes, 0.01 µg/cm²/minute | |
|---|---|---|
| 1. | Carmustine (BCNU) (3.3 mg/ml) | 0.20 |
| 2. | Cisplastin, (1.0 mg/ml) | >240 |
| 3. | Cyclophosphamide (20 mg/ml) | >240 |
| 4. | Doxorubicin HCl (2.0 mg/ml) | >240 |
| 5. | Etoposide (20 mg/ml) | >240 |
| 6. | Fluorouracil (50.0 mg/ml) | >240 |
| 7. | Methotrexate (25 mg/ml) | >240 |
| 8. | Paclitaxel (6.0 mg/ml) | >240 |
| 9. | Thiotepa (10.0 mg/ml) | 82.2 |
| 10. | Vincristine Sulfate (1.0 mg/ml) | >240 |
Please note that the following drug has extremely low permeation time of less than 30 minutes: Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 0.20 minute.
The proposed device is a disposable device. It is not made with natural rubber latex. Instead, the gloves are formulated using neoprene synthetic polymer and are coated with nitrile coating. The gloves are manufactured using exact same material used in the currently cleared device, Duraprene SMT gloves, that have been legally marketed by Cardinal Health under K102500. The gloves are manufactured using molds that feature anti-slip finish, independent thumb, and tapered mechanically locking cuffs to help reduce cuff roll down. They are light brown in color and are offered powder-free and sterile.
The acceptance criteria and device performance are as follows:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard/Test/FDA Guidance) | Reported Device Performance (Results Summary) |
|---|---|---|
| Biocompatibility: Primary Skin Irritation | ISO 10993-10 | Gloves are non-irritating. (Pass) |
| Biocompatibility: Guinea Pig Maximization | ISO 10993-10 | Gloves do not display any potential for sensitization. (Pass) |
| Physical Characteristics: Dimensions | ASTM D3577 | Meet requirements |
| Physical Characteristics: Physical Properties | ASTM D3577 | Meet requirements for rubber surgical gloves |
| Freedom from Holes | 21 CFR 800.20 & ASTM D3577 | Tested in accordance with ASTM D5151 with acceptable results (Meets AQL requirements) |
| Powder Residual | ASTM D3577 tested using ASTM D6124 standard test method | Gloves meet powder level requirements for "Powder-Free" designation per ASTM D3577. Results generated values < 2mg of residual powder per glove. |
| Chemotherapy Drug Permeation | ASTM D6978 | Exceeded maximum testing time of 240 minutes for most drugs. Carmustine (BCNU) (3.3 mg/ml) showed permeation time of 0.20 minutes. Thiotepa (10 mg/ml) showed permeation time of 82.2 minutes. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific test mentioned (e.g., number of gloves tested for permeation, number of animals for biocompatibility). The testing was conducted according to established international and national standards (ISO 10993-10, ASTM D3577, ASTM D6124, ASTM D6978, 21 CFR 800.20). These standards typically specify the required sample sizes for such tests to ensure statistical validity. The data provenance is not specified in terms of country of origin, but it is implied to be from testing labs adhering to these standards, likely in association with Cardinal Health for regulatory submission in the USA. The data would be prospective, as it was generated specifically for this 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This device is a medical glove, a physical product, and the "ground truth" for its performance is established through objective laboratory testing against material and safety standards, not through expert consensus on interpretation of data like in an imaging study. Therefore, the concept of "experts used to establish ground truth" in the way it applies to diagnostic AI is not relevant here. The ground truth is defined by the technical specifications and performance limits set by the referenced ASTM and ISO standards and FDA regulations. The "experts" in this context would be the qualified laboratory personnel conducting the tests and the regulatory bodies defining these standards.
4. Adjudication Method for the Test Set
Not applicable. As noted above, performance is measured against objective standards, not through subjective human interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human readers interpret images or other data, and the AI's impact on their performance is being evaluated. This document pertains to a medical glove, where performance is assessed through physical and chemical testing.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, a standalone performance assessment was done. The device performance (e.g., physical properties, biocompatibility, chemotherapy drug permeation) was evaluated independently through laboratory testing, without a "human-in-the-loop" component in the sense of a diagnostic interpretation task. The "algorithm" here is the glove itself, and its performance is measured directly against predefined criteria.
7. The Type of Ground Truth Used
The ground truth used is based on objective measurements and predefined standards (ASTM D3577, ASTM D6978, ISO 10993-10, 21 CFR 800.20). For example:
- Physical properties: Measured values (e.g., dimensions, tensile strength) compared against specified limits in ASTM D3577.
- Freedom from holes: Determined by a water leak test (ASTM D5151) and assessed against AQL (Acceptable Quality Level) requirements as per 21 CFR 800.20 and ASTM D3577.
- Powder residual: Quantitative measurement of powder per glove (ASTM D6124) compared against the < 2.0 mg/glove requirement for "powder-free."
- Chemotherapy drug permeation: Measured breakthrough detection time of specific chemotherapy drugs compared against a standard detection limit (0.01 µg/cm²/minute) over a specified testing period (up to 240 minutes) detailed in ASTM D6978 for each drug.
- Biocompatibility: Results from established biological evaluation tests (e.g., primary skin irritation, guinea pig maximization) against criteria in ISO 10993-10.
8. The Sample Size for the Training Set
Not applicable. This device is a physical product, not a software algorithm that requires a "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, this is a physical medical device, not an AI/ML algorithm requiring a training set.
Ask a specific question about this device
(64 days)
These powder-free sterile surgeon's gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities.
Gloves are formulated using neoprene. They are coated with nitrile coating and are offered powder-free and sterile.
Here's a breakdown of the acceptance criteria and the study information for the Cardinal Health Sterile Neoprene Powder-Free Surgical Gloves with Nitrile Coating, based on the provided text:
Based on the provided 510(k) summary, this document describes a device for general medical use (surgical gloves), not an AI/ML-based device. Therefore, many of the requested fields related to AI/ML studies, such as sample size for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance, and how ground truth was established, are not applicable to this submission. The "study" here refers to standard performance testing for medical gloves.
1. Table of Acceptance Criteria and Reported Device Performance
Device: Sterile Neoprene Powder-Free Surgical Gloves with Nitrile Coating
| Test / Acceptance Criteria | Reported Device Performance |
|---|---|
| Primary Skin Irritation | Gloves are non-irritating. |
| Guinea Pig Maximization (Sensitization Potential) | Gloves do not display any potential for sensitization. |
| Dimensions (meets requirements of ASTM D3577) | Gloves meet requirements of ASTM D3577. |
| Physical Characteristics (Tensile Strength, Elongation, etc.) (meets requirements for rubber surgical gloves per ASTM D3577) | Gloves meet requirements for rubber surgical gloves per ASTM D3577. |
| Freedom from Holes (meets requirements of 21 CFR 800.20 and ASTM D3577) | Gloves meet requirements of 21 CFR 800.20 and ASTM D3577. |
| Powder Residual (meets powder level requirements for "Powder-Free" designation per ASTM D3577, tested using ASTM standard D6124 - below 2mg/glove) | Results generated values below 2mg of residual powder per glove, meeting powder level requirements for "Powder-Free" designation per ASTM D3577 tested using ASTM standard D6124, Standard test method for residual powder on medical Doves. |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated for each test. Testing for medical gloves typically involves sampling a certain number of gloves from production lots according to relevant ASTM standards (e.g., ASTM D3577 for physical properties and freedom from holes, ASTM D6124 for powder residual).
- Data Provenance: The manufacturing location is listed as "7/111 Moo 4, Highway 331 Mabyangporn Pluakdaeng, Rayong, 21140, Thailand." The data is likely prospective testing performed on manufactured gloves.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This is not an AI/ML device where expert ground truth is established in the same manner. The "ground truth" for glove performance is determined by adherence to established governmental regulations (21 CFR 800.20) and industry standards (ASTM D3577, ASTM D6124), often through instrumental measurements and standardized methods rather than expert consensus on complex interpretations.
4. Adjudication method for the test set
- Not Applicable. No adjudication method, such as 2+1 or 3+1, is mentioned or would be relevant for these types of physical and chemical tests on surgical gloves. Performance is assessed against quantitative thresholds defined by standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an AI/ML device.
7. The type of ground truth used
- The ground truth for this device is based on established regulatory requirements (21 CFR 800.20) and recognized industry performance standards (ASTM D3577, ASTM D6124). These standards define the acceptable range or threshold for specific physical and chemical properties of the gloves.
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device, so there is no "training set."
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/ML device.
Ask a specific question about this device
Page 1 of 1