510(k) Data Aggregation

    K Number
    K072289
    Date Cleared
    2007-11-01

    (77 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SPINEFRONTIER DORADO VBR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORADO™ Intervertebral Body Cage is a spinal intervertebral body cage intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolithesis at the involved level(s). The SpineFrontier DORADO™ Intervertebral Body Cage is intended to be used with supplemental spinal fixation system(s) (Example: Posterior Pedicle Screw).

    Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

    Device Description

    The DORADO™ Intervertebral Body Cage is a spinal intervertebral body cage intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two spinal levels from L2-S1. The system is comprised of devices of various fixed heights and footprints to fit the anatomical needs of a wide variety of patients. The device has raised contours on the superior and inferior surfaces that will resist implant pullout.

    AI/ML Overview

    The provided text describes the 510(k) summary for the DORADO™ Intervertebral Body Cage, a medical device. It does not contain information related to software/AI performance or studies involving human readers or expert consensus for ground truth. Therefore, I cannot fulfill all parts of your request as the information is not present in the provided document.

    However, I can extract information related to the device's performance testing and acceptance criteria as described for this physical medical device.

    Acceptance Criteria and Study for DORADO™ Intervertebral Body Cage

    The provided document describes the mechanical performance testing of the DORADO™ Intervertebral Body Cage to demonstrate substantial equivalence to predicate devices, rather than a clinical study evaluating an AI/software device. The "acceptance criteria" and "device performance" in this context refer to the mechanical properties and successful adherence to ASTM standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    Static CompressionASTM F2077-03, "Test Methods for Intervertebral Body Fusion Devices"Demonstrated comparable functional and performance characteristics to the predicate device.
    SubsidenceASTM F2267-04, "Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression"Demonstrated comparable functional and performance characteristics to the predicate device.
    Push-out ResistanceASTM Draft Standard F-04.25.02.02, "Static Push-out Test Methods for Intervertebral Body Fusion Devices." Draft #2 - August 29, 2000Demonstrated comparable functional and performance characteristics to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: "Representative samples of the device were tested." Specifically, the "17mmx25mm device was selected for testing to support the addition of the 17mm device." This implies a limited number of devices were tested per dimension to represent the entire product line.
    • Data Provenance: Not applicable in the context of clinical data. This refers to mechanical testing conducted in a laboratory setting.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This is not applicable as the study involved mechanical testing of a physical device, not an AI/software device requiring expert interpretation for ground truth.

    4. Adjudication Method

    Not applicable for mechanical testing.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    Not applicable as this is a physical medical device, not an AI/software diagnostic tool.

    6. Standalone Performance

    The testing performed was for the standalone mechanical integrity of the device against established ASTM standards. The results "demonstrate comparable functional and performance characteristics to the predicate device."

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is established by ASTM (American Society for Testing and Materials) standards for intervertebral body fusion devices. These standards define the acceptable mechanical properties and testing methodologies.

    8. Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of mechanical performance testing of a finished medical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.