Search Results
Found 9 results
510(k) Data Aggregation
(120 days)
Solution Administration Sets
For the administration of fluids from a container into the patient's vascular system through a vascular access device.
The proposed devices consist of Solution Administration Sets. These devices include Basic, Secondary, CONTINU-FLO solution sets, Stand-Alone devices and Chemotherapy devices (see Table 2 for a list of subject device set names per product family). They are single use disposable, non-pyrogenic, sterile devices intended for the administration of fluids from a container into the patient's vascular system.
This is a 510(k) premarket notification for "Solution Administration Sets" by Baxter Healthcare Corporation. The document states that the devices are substantially equivalent to a predicate device (K203609 cleared on September 30, 2021).
Here's the breakdown of the acceptance criteria and the study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't provide a specific table of acceptance criteria with corresponding performance values in the format usually seen for AI/ML devices. Instead, it describes general conformance to recognized standards and the positive outcomes of various tests.
Acceptance Criteria (Standard / Test) | Reported Device Performance |
---|---|
ISO 80369-7: 2021 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications) | All proposed devices were found to be in conformance with this standard. Performance testing included mechanical (tensile strength), pressure (burst, leakage, backflow, internal), stress cracking, resistance, torque, spike insertion/removal force, drop form accuracy, vacuum, pump/set integrity, DEHP content. |
ISO 8536-4 (Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed) | Complete sets meet the performance requirements of this standard (mentioned in relation to priming volume and tubing types). |
ISO 10993-1 (Biological Evaluation of Medical Devices) / FDA-2013-D-0350 Guidance | The proposed devices are biocompatible and appropriate for their intended use. Biocompatibility tests conducted: Cytotoxicity, Sensitization, Intracutaneous (Irritation) Reactivity, Acute Systemic Toxicity, 30 Day Systemic Repeat Dose Toxicity Study, Material Mediated Pyrogen, Hemolysis. All met acceptance criteria. |
USP Particulate Matter in Injections | Filter performance testing included particulate retention, integrity, air filter flow. Particulate matter testing met the USP Acceptance criteria. |
Microbial Ingress Testing (Baxter's testing strategy, per K223175) | All test results met their acceptance criteria, demonstrating the absence of microbial ingress into the sterile fluid path during simulated clinical use, supporting appropriate design for intended use. |
ISO 11137-1: 2006 (Sterilization of health care products - Radiation - Part 1) | Sterilization process established per this standard. Devices sterilized via radiation with a minimum Sterility Assurance Level (SAL) of 10-6. |
ISO 11137-2: 2013 (Sterilization of health care products - Radiation - Part 2) | Minimum Sterilizing Dose (MSD) established and validated as per Method 1. Continued validity confirmed via periodic dose audit studies. |
ISO 11607-1: 2019 (Packaging for terminally sterilized medical devices - Part 1) | Package verification testing performed per this standard (Simulated Distribution per ASTM D4169-22) and included visual (ASTM F1886), seal strength (ASTM F88), and bubble test (ASTM F2096-11). All met requirements. |
ASTM F1980-21 (Accelerated Aging of Sterile Barrier Systems and Medical Devices) | 2-year shelf-life confirmed via accelerated aging. |
Note: This submission is for a traditional medical device (solution administration sets), not an AI/ML device. Therefore, the questions related to AI/ML specific studies (sample size for test set, data provenance, number of experts for ground truth, adjudication, MRMC study, standalone performance, training set sample size, training set ground truth) are not applicable to this document. The "tests" described are standard engineering, biocompatibility, and sterilization validations for physical medical devices.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable, as this is a traditional medical device, not an AI/ML device. The testing described is bench testing and biocompatibility assessments, not a study involving patient data or a specific test set in the AI/ML context.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as this is a traditional medical device, not an AI/ML device. Ground truth as typically defined for AI/ML models is not relevant here.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as this is a traditional medical device, not an AI/ML device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable, as this is a traditional medical device, not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as this is a traditional medical device, not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as this is a traditional medical device, not an AI/ML device. Instead of "ground truth," the device relies on conformance to established international and national standards (ISO, ASTM, USP) and predefined acceptance criteria for various physical, chemical, and biological tests.
8. The sample size for the training set
Not applicable, as this is a traditional medical device, not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable, as this is a traditional medical device, not an AI/ML device.
Ask a specific question about this device
(56 days)
Solution Administration Sets with 0.2 Micron Filter
For the retention of microorganisms and removal of air and particulate matter from infusion fluids.
The Solution Administration Sets with a 0.2 micron filter product line consists of sterile, non-pyrogenic, single use disposable devices used for the administration of fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. They are indicated for the retention of microorganisms and removal of air and particulate matter from infusion fluids. The filter consists of a 0.2 micron polyethersulfone (PES) solution membrane and 0.1 micron polyvinylidene fluoride air vent membrane enclosed in a copolyester housing.
The provided text is a 510(k) summary for a medical device (Solution Administration Sets with 0.2 Micron Filter) and, as such, focuses on demonstrating substantial equivalence to a predicate device for regulatory clearance rather than a comprehensive study evaluating device performance against established acceptance criteria in a research context.
This document does not contain the kind of detailed information about a study that would rigorously prove a device meets acceptance criteria in the typical scientific sense (e.g., sample sizes, ground truth establishment, expert qualifications, MRMC studies). It is a regulatory submission, so the "studies" are verification tests to ensure the modified device is equivalent to the predicate.
However, I can extract the acceptance criteria mentioned and the reported "performance" based on the provided text, while noting the limitations in the depth of information available.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Reported Device Performance (as stated in the document) |
---|---|
Performance Data | |
Air diffusion | All tests met the acceptance criteria. |
Bubble point | All tests met the acceptance criteria. |
Gravity flow rate | All tests met the acceptance criteria. |
Flow rate post sterile water conditioning | All tests met the acceptance criteria. |
Flow rate post parenteral nutrition conditioning | All tests met the acceptance criteria. |
Bacterial retention | All tests met the acceptance criteria. |
Biocompatibility | |
Cytotoxicity | All tests met the acceptance criteria. |
Systemic Toxicity | All tests met the acceptance criteria. |
Intracutaneous | All tests met the acceptance criteria. |
Hemolysis | All tests met the acceptance criteria. |
Pyrogen | All tests met the acceptance criteria. |
Sensitization | All tests met the acceptance criteria. |
USP Physicochemical | All tests met the acceptance criteria. |
Note: The document only states that "All tests met the acceptance criteria" without providing the specific numerical or qualitative thresholds for those criteria. It implies that these criteria were pre-established internally by Baxter Healthcare Corporation for their risk analysis and design verification.
Here's why the other requested information is largely not present in this type of regulatory document:
2. Sample size used for the test set and the data provenance:
- Not explicitly stated. The document is a summary and does not include the detailed protocols, sample sizes, or statistical analysis reports from the bench tests or biocompatibility assessments.
- Data Provenance: The tests were conducted by Baxter Healthcare Corporation. The document doesn't specify the country of origin for the data beyond that. These are typically internal corporate studies (retrospective in the sense that they are conducted on manufactured samples, but prospective in terms of the test design).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not mentioned. This information is typically relevant for studies involving human interpretation (e.g., image analysis, clinical evaluations). For bench tests and biocompatibility tests of a physical device, the "ground truth" is typically defined by scientific principles, international standards (e.g., ISO-10993), and validated test methodologies. There's no "expert ground truth" in the sense of consensus from adjudicators.
4. Adjudication method for the test set:
- Not applicable/Not mentioned. As above, adjudication is not a standard part of these types of engineering and biological safety tests. The results are typically quantitative measurements against defined specifications.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a completely different type of study, relevant for AI/radiology devices. This document is a 510(k) for an administration set with a filter, not an AI diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This question pertains to AI algorithms. The device discussed is a physical medical device.
7. The type of ground truth used:
- For Performance Data (e.g., flow rate, bacterial retention): The "ground truth" would be established by the validated test methods themselves, based on physical and microbiological principles, often referenced to international standards or established industry practices for filter performance.
- For Biocompatibility: The "ground truth" is defined by the requirements of international standards like ISO-10993-1 and FDA guidance, which specify the types of biological responses that are considered acceptable or unacceptable.
8. The sample size for the training set:
- Not applicable. This question refers to machine learning models. This device does not involve a "training set" in that context.
9. How the ground truth for the training set was established:
- Not applicable. As above, no training set for an AI model is involved.
Summary of the Study that Proves the Device Meets Acceptance Criteria:
The document describes a series of nonclinical bench tests and biocompatibility assessments conducted by Baxter Healthcare Corporation.
- Objective: To evaluate the effect of a material modification (change in solution membrane material from one hydrophilic polyethersulfone to another hydrophilic polyethersulfone) in the 0.2 micron filter within their Solution Administration Sets. The goal was to establish substantial equivalence to the previously cleared predicate device (K964850).
- Tests Performed:
- Performance Data: Air diffusion, bubble point, gravity flow rate, flow rate post sterile water conditioning, flow rate post parenteral nutrition conditioning, and bacterial retention.
- Biocompatibility: Cytotoxicity, Systemic Toxicity, Intracutaneous, Hemolysis, Pyrogen, Sensitization, and USP Physicochemical. These were conducted in accordance with ISO-10993 and FDA guidance.
- Results: The document states that "All tests met the acceptance criteria."
- Conclusion: Based on these tests, Baxter concluded that "The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate devices that are currently marketed for the same intended use."
This is a regulatory study designed to show equivalence, not an independent research study to establish novel performance claims.
Ask a specific question about this device
(177 days)
HEALTH LINE INTERNATIONAL CORPORATION'S PRIMARY IV SOLUTION ADMINISTRATION SET WITH CHECK VALVE
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow requlator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow requlator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
The provided documents are a 510(k) clearance letter and an "Indications for Use" statement for the Health Line International Corporation's IV Solution Administration Set with Check Valve. They do not contain information about acceptance criteria or a study proving that a device meets such criteria.
The 510(k) process is for demonstrating substantial equivalence to a predicate device, not for proving performance against specific acceptance criteria through a study as would be detailed for an AI/ML medical device.
Therefore, I cannot extract the requested information from the provided text.
Ask a specific question about this device
(315 days)
PENTADEFLUE IV SOLUTION ADMINISTRATION SETS
To Administer IV Fluids/medications to the patient's vascular system.
To administer IV Fluids into a patient's vascular system
The PENTADEFLU IV Administration Sets are Single Use, Sterile, Non-Pyrogenic devices used to administer IV fluids/medications to a patient's vascular system via gravity control.
The provided text describes the 510(k) summary for the PENTADEFLU IV Administration Sets, a medical device. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than undergoing extensive clinical trials with acceptance criteria and comparative effectiveness studies as might be seen for novel therapeutic devices or AI-powered diagnostics.
Therefore, the information requested, such as acceptance criteria based on performance metrics, sample sizes for test and training sets, expert adjudication methods, MRMC studies, and ground truth establishment, is generally not present or applicable in the context of this 510(k) summary. The summary focuses on engineering and biocompatibility testing, along with demonstrating similarity to a predicate device.
Here's an analysis based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Reported Device Performance | Comments |
---|---|---|
Mechanical Properties | Verified | "subjected to performance and safety testing to verify mechanical properties and functioning" - Specific numerical criteria and results are not provided in this summary. |
Functioning | Verified | "subjected to performance and safety testing to verify mechanical properties and functioning" - Specific functional criteria and results are not provided in this summary. |
Biocompatibility | Verified | "biocompatibility... using FDA recognized Standards, where applicable" - Specific tests and results are not provided. |
Sterility | Verified | "sterility... using FDA recognized Standards, where applicable" - Specific tests and results are not provided. |
Design | No modifications affect safety/effectiveness | "There are no modifications to the device design that affect safety and effectiveness of the PENTADEFLU IV Administration Sets" compared to the predicate. |
Technological Characteristics | Same as predicate device | "The PENTADEFLU IV Administration Sets have the same technological characteristics as the legally marketed predicative IV Administration Sets." |
2. Sample size used for the test set and the data provenance
- Not Applicable / Not Provided. This 510(k) summary does not describe a "test set" in the context of an AI/ML algorithm or diagnostic accuracy study. The testing mentioned refers to engineering, performance, biocompatibility, and sterility testing of the physical device. The provenance of samples for these tests (e.g., manufacturing batches) is not detailed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable / Not Provided. Ground truth establishment by experts during a diagnostic accuracy study is not relevant to this device's 510(k) submission.
4. Adjudication method for the test set
- Not Applicable / Not Provided. Adjudication methods are relevant for diagnostic studies, not for the type of device described here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable / Not Provided. This device is an IV administration set, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable / Not Provided. This device is a physical medical administration set, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not Applicable / Not Provided directly in this format. For a physical device like an IV set, "ground truth" would refer to its adherence to engineering specifications, biocompatibility standards, and sterility requirements as verified by standardized testing methods and regulatory guidelines (e.g., ISO standards, FDA regulations for IV sets). This isn't "ground truth" in the diagnostic sense.
8. The sample size for the training set
- Not Applicable / Not Provided. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
- Not Applicable / Not Provided. There is no "training set."
Summary of the Study:
The "study" described in the 510(k) summary for the PENTADEFLU IV Administration Sets is a series of performance and safety tests conducted to demonstrate the device's conformance to established standards and its substantial equivalence to a legally marketed predicate device (Victus IV Administration Sets, K023469).
The key aspects of the study are:
- Type of Study: Bench testing for mechanical properties, functioning, biocompatibility, and sterility. There are no clinical trials or human subject studies described for this 510(k).
- Purpose: To verify that the device meets safety and performance requirements for its intended use and is substantially equivalent to the predicate device.
- Methodology: Testing was conducted "using FDA recognized Standards, where applicable." The specific standards are not listed in this summary, but they would typically include ISO standards for medical devices, specifically those pertaining to IV administration sets (e.g., ISO 8536 series), as well as biocompatibility standards (e.g., ISO 10993 series) and sterility validation.
- No specific acceptance criteria values (e.g., tensile strength, flow rate tolerances) or numerical performance results are included in this high-level summary. The statement merely indicates that such testing was performed and verified.
- Comparison to Predicate: The submission asserts that the device has "the same technological characteristics" as the predicate and that "no modifications to the device design that affect safety and effectiveness" have been made. This forms the primary basis for the substantial equivalence determination.
Ask a specific question about this device
(56 days)
BLOOD/SOLUTION ADMINISTRATION SET, MODEL 33.015
Administration of Intravenous fluids and drugs.
The Tuta Healthcare Pty. Limited Blood Administration Set is a device used to administer fluids, blood and blood products from a container to a patient's vascular system Through a catheter or venous access system inserted into a vein.
Use of Needle-Free Access site may aide in the prevention of needlestick injury.
The Tuta Healthcare Blood/Solution Administration Set is designed administer fluids from a container to a patient's vascular system through a needle catheter inserted into a vein. The pump helps to control the rate of flow of fluids from the container to the patient.
The provided text describes a 510(k) summary for the Tuta Healthcare Blood/Solution Administration Set, comparing it to the legally marketed Baxter Healthcare's Solution Administration Set (K924721). The submission aims to demonstrate substantial equivalence, focusing on design, materials, intended use, and performance.
However, the document does not contain the kind of detailed information typically found in studies for AI/ML-enabled medical devices or diagnostic devices, especially regarding acceptance criteria, specific performance metrics (like sensitivity, specificity, AUC), sample sizes for test/training sets, expert qualifications, or adjudication methods for ground truth, as these are not relevant to this type of device (an administration set) or the type of substantial equivalence submission presented.
The study referenced is a laboratory bench testing to assess the new device against the predicate device.
Here's a breakdown of the information that is available in the provided text in relation to your questions:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Same indication statement as predicate device | Yes, "administration of fluids from a container to a patient's vascular system through a needle catheter inserted into a vein." |
Same technological characteristics (design, materials, principle of operation) as predicate device | Yes, similar components (PVC tubing, roller regulators, perforators) and same principle of operation. |
New characteristics do not affect safety or effectiveness | No new characteristics identified that would affect safety or effectiveness. |
Descriptive characteristics precise enough to ensure equivalence | Yes. |
Biocompatibility of fluid path materials | Materials suitable for limited contact (tested according to General Program Memorandum #G95). |
Performance (e.g., Flow rate) compares favorably to predicate device | Flow rate testing of the Blood/Solution Administration Set "compares favourably" to the Baxter Healthcare's Solution Administration Set. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified. The document states "Laboratory beach testing has been performed," but does not provide details on the number of units or test repetitions.
- Data Provenance: The testing was "Laboratory beach testing," implying it was conducted in a controlled environment. The manufacturer is based in Australia, but the testing location is not explicitly stated. It is a prospective study as tests were performed specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This device is an administration set, not a diagnostic device requiring expert interpretation of results or establishing ground truth based on clinical expert consensus. The "ground truth" here is the performance of the predicate device and established safety standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This device is an administration set. Adjudication methods are typically relevant for diagnostic studies where there's variability in interpretation or a need for consensus on clinical findings.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is not an AI-enabled diagnostic tool, and no human reader studies were conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device does not have an algorithm or AI component.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" in this context is the performance of the legally marketed predicate device (Baxter Healthcare's Solution Administration Set K924721) and adherence to recognized safety and performance standards (e.g., biocompatibility testing per General Program Memorandum #G95).
8. The sample size for the training set
Not applicable. There is no "training set" as this is a physical medical device and not an AI/ML model.
9. How the ground truth for the training set was established
Not applicable, as there is no training set. The "ground truth" for comparison is the predicate device's established performance and regulatory compliance.
Ask a specific question about this device
(3 days)
SOLUTION ADMINISTRATION SETS WITH CAPLESS LUER ACTIVATED VALVE
Ask a specific question about this device
(88 days)
SOLUTION ADMINISTRATION SET
The solution adminstration set will be used to administer fluids to a patient's vascular system from a container through a needle or catheter inserted into a vein. The proposed solution set will be primarily used to administer solutions containing the chemotherapeutic drug paclitaxel, but can also be used for general solution administration with Baxter Flo-Gard® (6200 and 6300 series) and Colleague" volumetric infusion pumps.
The proposed solution set will be used for the administration of intravenous solutions containing the chemotherapeutic drug paclitaxel. The package insert for paclitaxel recommends that the drug be administered parenterally through a polyethylene-lined set because of the concern with diethyl-2-hexylphthalate (DEHP) leaching from DEHP plasticized polyvinyl chloride (PVC). The proposed set is designed to meet the package insert recommendations in that it contains polyethylene-lined tubing, an in-line 0.22 um filter, and no DEHP plasticized PVC in contact with the solution.
The provided 510(k) Premarket Notification document (K981792) is for a "Solution Administration Set". This type of medical device is a physical product, not an AI/ML algorithm. Therefore, many of the requested criteria related to AI/ML device studies (e.g., sample size for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, ground truth establishment for training set) are not applicable to this submission.
However, based on the non-clinical tests described, here's an attempt to extract relevant information and note the inapplicable sections:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|
Functional performance meets requirements for administering IV solutions. | "proposed paclitaxel device meets or exceeds all functional requirements" |
Compatibility with paclitaxel (does not leach DEHP into solution). | Compatibility with paclitaxel "has been generated." |
No DEHP plasticized PVC in contact with the solution (specific design feature). | "contains polyethylene-lined tubing... and no DEHP plasticized PVC in contact with the solution." |
In-line 0.22 um filter is present and functional. | "contains... an in-line 0.22 um filter." |
New device is substantially equivalent to predicate device (Baxter nitroglycerin set K83284A). | Stated as "similar" with "principal differences" accounted for (spike design, 0.22µm filter, TEHTM plasticizer). |
All solution contact materials are identical to materials used in legally marketed devices. | "All solution contact materials to be used in the proposed set are identical to materials used in legally marketed devices under comparable conditions of use." |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified in the document. The document refers to "Data regarding the functional performance" being "generated," but does not provide details on the number of units tested.
- Data Provenance: Not explicitly stated, but assumed to be generated from non-clinical laboratory testing performed by Baxter Healthcare Corporation. It is prospective testing of the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This criterion is not applicable. The device is a physical administration set; its "ground truth" is its physical and chemical properties and functional performance, which are evaluated through direct testing, not expert interpretation of data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This criterion is not applicable. Adjudication methods are typically used for expert consensus on image interpretation or complex clinical outcomes. The functional testing of a physical device does not involve such methods.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This criterion is not applicable. The device is a physical medical device, not an AI/ML algorithm. No human reader studies with AI assistance were performed or are relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This criterion is not applicable. The device is a physical medical device, not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for this device's performance would be established through direct physical and chemical testing standards and laboratory analyses. For example:
- Measurement of flow rates.
- Chemical analysis for DEHP leaching.
- Filtration efficacy.
- Material compatibility tests.
- Sterility validation (implied for an IV set, though not explicitly detailed here).
8. The sample size for the training set
- This criterion is not applicable. The device is a physical medical device, not an AI/ML algorithm. "Training sets" are for AI/ML models.
9. How the ground truth for the training set was established
- This criterion is not applicable. The device is a physical medical device, not an AI/ML algorithm.
Ask a specific question about this device
(164 days)
SOLUTION ADMINISTRATION SET WITH CAPPED LUER ACTIVATED VALVE
Baxter solution administration sets with capped luer activated valves are intended for use with a vascular access device for the administration of drugs and solutions. The capped luer activated valve will replace the Y-injection site on the set. The valve can be connected to standard male luer adapters (e.g., syringes or sets), for continuous or intermittent fluid administration or the withdrawal of fluids.
The subject of this submission is a capped luer activated valve which Baxter intends to incorporate into currently marketed solution sets. The capped luer activated valve will replace the Y-injection site on the set for continuous or intermittent fluid administration or the withdrawal of fluids. The normally closed valve is opened by removing the cap and inserting a standard male luer adapter such as on syringes or sets to the female end of the valve automatically closes when the male luer is disconnected. The valve is intended to be capped when not in use.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Solution Administration Sets with Capped Luer Activated Valve:
Acceptance Criteria and Study Analysis
The provided 510(k) summary for the "Solution Administration Sets with Capped Luer Activated Valve" describes a device intended to replace a Y-injection site on existing solution administration sets. The submission focuses on demonstrating substantial equivalence to a predicate device, the Abbott I.V. Sets with Capped Reflux Valve Port.
It's important to note: The provided document is a 510(k) summary, not a full study report. As such, detailed information regarding specific acceptance criteria, test methodologies, and data breakdowns is very limited. The document states that "A description of the functional testing along with test results has been provided," implying these details were part of the complete 510(k) submission to the FDA, but they are not present in this public summary.
Therefore, many of the requested fields below cannot be fully populated from the provided text.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Acceptance Criteria (from document) | Reported Device Performance (from document) |
---|---|---|
Functional Performance | (Not explicitly detailed in this summary) | "The data indicate that the proposed valve meets or exceed all functional requirements and support its suitability for use in Baxter solution administration sets." |
Explanation: The document broadly states that the device was evaluated for "functional performance" and met "all functional requirements." However, it does not specify what those functional requirements were (e.g., specific flow rates, pressure resistance, leakage limits, number of activations without failure, material compatibility, sterilization effectiveness, etc.).
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): Not specified in the provided text.
- Data Provenance: Not specified in the provided text. The testing would have been conducted by Baxter Healthcare Corporation, likely at their facilities, but no geographic origin or retrospective/prospective nature is mentioned.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. This device is a mechanical medical device (solution administration set). The "ground truth" for its performance is established through engineering and functional testing, not typically through human expert adjudication as would be the case for diagnostic AI.
- Qualifications of Experts: N/A
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. For a mechanical device, testing results are objectively measured against predefined engineering specifications. There is no subjective adjudication process mentioned or typically used for this type of device.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done? No. This type of study is relevant for diagnostic devices that humans interpret (e.g., radiology AI). This submission is for a mechanical solution administration set, which does not involve human interpretation in its primary function, nor does it involve "AI."
- Effect Size of AI Assistance: Not applicable.
6. Standalone Performance (Algorithm Only)
- Standalone Performance Done? N/A. This device does not have an "algorithm" in the sense of AI or software interpreting data. Its performance is entirely "standalone" in that it functions as a mechanical component.
7. Type of Ground Truth Used
- Type of Ground Truth: Engineering specifications and performance metrics. For example, leakage tests, flow rate measurements, pressure resistance, ability to connect and disconnect luer adapters, durability/number of cycles, material biocompatibility, and sterilization validation would all contribute to the "ground truth" of the device's functional performance. The document only broadly refers to "functional requirements."
8. Sample Size for the Training Set
- Sample Size (Training Set): Not applicable. This submission is for a mechanical medical device, not an AI/ML model that requires a training set. The development of such devices involves design, prototyping, and iterative functional testing, but not a "training set" in the computational sense.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth Establishment (Training Set): Not applicable.
Summary of Limitations from the Document:
The provided 510(k) summary is a high-level overview. It confirms that functional testing was performed and that the device met its requirements for substantial equivalence. However, it lacks the detailed technical specifications and study breakdowns that would be present in the full original 510(k) submission to the FDA. Specifically, it does not provide:
- Exact numerical acceptance criteria for any functional test.
- Quantified test results (e.g., "passed 1000 cycles" vs. "meets requirements").
- The number of units tested (sample size).
- Specific test methodologies.
- Any information related to clinical trials or human factors testing, beyond "intended for use with a vascular access device."
The document's primary purpose is to declare substantial equivalence to a predicate device (Abbott I.V. Sets with Capped Reflux Valve Port) based on similar technological characteristics and demonstrated functional performance, allowing the device to be marketed.
Ask a specific question about this device
(84 days)
SOLUTION ADMINISTRATION SET WITH 0.22 MICRON HIGH PRESSURE EXTENDED LIFE FILTER
Not Found
The subject of this submission is the Millipore 0.22 micron filter which Baxter intends to incorporate into various configurations of solution sets for the administration of intravenous solutions. The filter is manufactured by Millipore and has been recently cleared for marketing in Abbott solution administration sets covered by K960466. Baxter will purchase the 0.22 micron filter from Millipore and incorporate it into currently marketed solution administration sets. Baxter is making no changes to the design, components or materials of the Millipore filter.
This 510(k) summary describes a new solution administration set incorporating a 0.22 micron filter manufactured by Millipore, intended for use with intravenous solutions. The submission focuses on the filter itself, as Baxter will be integrating it into their existing solution administration set designs. The predicate devices are Abbott and Baxter solution sets, both incorporating 0.22 micron filters.
Here's an analysis of the provided information based on your requested criteria:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Reported Device Performance |
---|---|
Functional Performance | Met or exceeded all functional requirements. |
Reverse Pressure Filter Integrity | Met (implied by "meets or exceeds all functional requirements"). |
Unit Venting | Met (implied by "meets or exceeds all functional requirements"). |
Gravity Flow | Met (implied by "meets or exceeds all functional requirements"). |
Inlet Axial Stress | Met (implied by "meets or exceeds all functional requirements"). |
Housing Integrity/Female Luer Fitment | Met (implied by "meets or exceeds all functional requirements"). |
Forward Pressure Filter Integrity | Met (implied by "meets or exceeds all functional requirements"). |
Downstream Particle Count | Met (implied by "meets or exceeds all functional requirements"). |
Accelerated Endurance | Met (implied by "meets or exceeds all functional requirements"). |
Outlet Deflection Stress | Met (implied by "meets or exceeds all functional requirements"). |
In-Line Filter Wettability | Met (implied by "meets or exceeds all functional requirements"). |
Biocompatibility | Met or exceeded all functional requirements (this usually encompasses biocompatibility testing). |
Bacterial Endotoxins | Met (implied by "functional, microbiological, and drug compatibility data indicate that the proposed filter meets or exceeds all functional requirements"). |
Bioburden Evaluation | Met (implied by "functional, microbiological, and drug compatibility data indicate that the proposed filter meets or exceeds all functional requirements"). |
Structural Integrity | Met or exceeded all functional requirements. |
Burst Strength | Met (implied by "meets or exceeds all functional requirements"). |
MVI Resistance | Met (implied by "meets or exceeds all functional requirements"). |
Drug Compatibility | Met (explicitly stated: "drug compatibility data indicate that the proposed filter meets or exceeds all functional requirements"). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated. The summary mentions "Data regarding the functional performance [...] were generated by Millipore," but does not specify the number of units tested for each criterion.
- Data Provenance: The data was generated by Millipore, the manufacturer of the 0.22 micron filter. It is not specified whether the data is retrospective or prospective, nor the country of origin, though given the context of a US 510(k) submission, it's highly likely to be considered within a US regulatory framework.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this type of medical device submission. This refers to clinical studies where expert consensus or diagnostic interpretation is required for ground truth. For a filter, the "ground truth" is typically defined by standardized physical, chemical, and microbiological tests.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical trials or diagnostic studies to resolve discrepancies between multiple expert readers. For a filter, objective measurements are taken, and results are compared against predefined acceptance limits.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. This submission is for a physical medical device (a filter), not an AI-powered diagnostic tool. Therefore, MRMC studies and the concept of human reader improvement with AI assistance are irrelevant here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is not applicable. As mentioned above, this is a physical device, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device is based on objective, quantitative measurements obtained from established test methods for filter performance, biocompatibility, and structural integrity. This includes:
- Physical measurements (e.g., burst strength, flow rates).
- Chemical purity assessments (e.g., bacterial endotoxins, particle counts).
- Microbiological testing (e.g., bioburden evaluation) demonstrating sterile filtration capabilities.
- Drug compatibility studies.
These "ground truths" are derived from validated test procedures and industry standards relevant to medical filters.
8. The Sample Size for the Training Set
This is not applicable. This device is a physical product, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable, as there is no training set for a physical filter device.
Ask a specific question about this device
Page 1 of 1