510(k) Data Aggregation

    K Number
    K170500
    Manufacturer
    Date Cleared
    2017-05-16

    (84 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    SIROLaser Advance+

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SIROLaser Advance+ is intended for intra- and extra-oral surgery including incision, hemostasis, coagulation and vaporization of soft tissue including marginal and epithelial lining of free gingiva and is indicated for: frenectomy; frenotomy; implant recovery; gingivectomy; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of diseased epithelial lining; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty ; excision of lesions; exposure of unerupted/partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket, sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; fibroma removal; gingival incision and excision; treatment of canker sores; herpetic ulcers of the oral mucosa; laser soft tissue curettage; reduction of gingival hypertrophy.

    Whitening: The SIROLaser Advance+ is indicated for light activation for bleaching materials for teeth whitening and for laser-assisted whitening/bleaching of teeth.

    Low Level Laser Therapy: The SIROLaser Advance+ is intended to emit energy in the red and infrared spectrum to provide topical heating for the purpose of elevating tissue temperary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, and for the temporary increase in local blood circulation and or temporary relaxation of muscles.

    Device Description

    The SIROLaser Advance+ is a solid state laser in which laser energy is generated by internal diodes. The laser output energy of the device is in the infrared and red spectra at wavelengths of 970 nm and 660nm, respectively. The output settings of the SIROLaser Advance+ can be adjusted by the user and also features preset operating modes which are designed to correspond with its proposed surgical, periodontic, and endodontic applications.

    SIROLaser Advance+ primarily consists of a laser control unit with graphical touch screen user interface, a laser handpiece, and single-use, disposable, "EasyTip" laser fiber guides (for surgical applications), as well as, reusable "MultiTip" light guides (for non-contact, low-power, "soft laser" applications). The laser control unit contains the 970 nm and 660 nm laser diodes, electronic hardware, touchscreen graphical user interface, and software to control the laser functionality.

    The SIROLaser Advance+ handpiece assembly consists of an internal handpiece body with laser activation finger switch. The handpiece includes a removable stainless steel outer sleeve which can be cleaned and sterilized between patient procedures. The SIROLaser Advance+ is also offered with an optional wireless footswitch.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the SIROLaser Advance+ device, which is an laser surgical instrument. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than establishing new safety and efficacy through extensive clinical trials with specific acceptance criteria related to disease detection or treatment outcomes.

    Therefore, the information you're looking for, such as a table of acceptance criteria with reported device performance related to a clinical outcome, sample sizes for test sets, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone performance, and detailed training set information, is not typically included or required in this type of regulatory submission because the device itself does not make a diagnostic or prognostic claim that would necessitate such studies.

    Instead, the "acceptance criteria" for a 510(k) submission generally revolve around demonstrating that the new device is as safe and effective as a legally marketed predicate device. This is achieved through non-clinical performance data, primarily focusing on engineering, safety, and operational specifications.

    Here's a breakdown based on the provided text, explaining why certain requested information is absent:


    Acceptance Criteria and Study Supporting Device Equivalence (Not Clinical Performance)

    The "acceptance criteria" in this context refer to the successful demonstration of equivalence to predicate devices and adherence to relevant safety and performance standards. The "study" refers to the non-clinical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission for substantial equivalence based on safety and operational performance, there isn't a direct equivalent to a clinical "acceptance criteria" table and reported "device performance" in terms of clinical accuracy (e.g., sensitivity, specificity for a diagnostic device).

    Instead, the performance verification is against engineering standards and specific operational specifications, which are compared to predicate devices for substantial equivalence.

    Acceptance Criteria (General for 510(k))Reported Device Performance (from Non-Clinical Data)
    Safety and Electrical Standards- Conformity to IEC 60601-1 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance).
    • Conformity to IEC 60601-1-2 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance. Collateral Standard: Electromagnetic compatibility). |
      | Laser Safety Standards | - Conformity to IEC 60825-1 (Safety of laser products – Part 1: Equipment classification and requirements).
    • Conformity to IEC 60601-2-22 (Medical electrical equipment Part 2: Particular Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment). |
      | Usability | - Usability study conducted in conformity with IEC 62366 (Medical devices – Application of usability engineering to medical devices). |
      | Biocompatibility | - Biocompatibility assessment of patient contacting components according to ISO 10993-5 (Biological evaluation of medical devices Part 5: Tests for cytotoxicity). |
      | Software Validation | - Validation of the device's software in conformity with IEC 62304 (Medical device software – Software lifecycle processes). |
      | Functional Equivalence | - Wavelengths: 970 nm (-10 nm/+15 nm) & 660 nm (+/- 5 nm), comparable to predicates (Epic 10: 940 nm; SIROLaser Advance: 970 nm; THOR VR: 660 nm).
    • Optical Power: 970 nm: 0.2W-7.0W (CW), 14W (peak); 660 nm: 25mW, 50mW, 100mW. Comparable to predicates (Epic 10: 10W; SIROLaser Advance: 7.0W, 14W peak; THOR VR: 30mW, 75mW, 200mW).
    • Emission Modalities & Pulse Duration: Continuous Wave, Chopped (1Hz-10kHz), Peak Pulse (1.5kHz-20kHz), comparable to predicates.
    • Aiming Beam: 660 nm (+/- 5 nm), 1mW (max), comparable to predicates (625-670 nm, 1mW max).
    • Fiber Diameter: 200 µm, 320 µm, comparable to predicates (200 µm, 300 µm, 400 µm).
    • User Interface: Color touch screen graphical user interface, same as primary predicate. |

    2. Sample Size Used for the Test Set and Data Provenance

    The document states, "No human clinical data was included to support substantial equivalence." Therefore, there is no "test set" in the sense of patient data for clinical performance evaluation. The "test set" here refers to the actual device units tested under various engineering standards and for biocompatibility. The provenance would be the Dentsply Sirona testing labs or certified third-party testing facilities.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not Applicable. Since no human clinical data was included, there was no need for experts to establish ground truth on patient cases. The qualification for the engineers and technicians performing the non-clinical tests would be their certifications in electrical, laser, mechanical, and software engineering, as well as biocompatibility.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not Applicable. Adjudication methods are used in clinical studies to resolve discrepancies in expert opinions on patient data. This was not a clinical study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not Applicable. This device is a laser surgical instrument, not an AI-assisted diagnostic or therapeutic tool that would involve human "readers" interpreting output.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not Applicable. This is a hardware device (laser) with accompanying software for control. "Standalone performance" in this context refers to the device's adherence to its specified performance parameters (e.g., power output, wavelength stability, safety features) without human intervention in operating the device itself during these tests, which were performed per the listed IEC standards.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    For this 510(k) submission, the "ground truth" for the non-clinical tests is:

    • Engineering Specifications and International Standards: The device's performance (e.g., laser power, wavelength, electrical safety, EMC, software functionality) is compared against pre-defined, industry-accepted standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, IEC 62366, IEC 62304) and the manufacturer's own design specifications.
    • Biocompatibility Standards: The patient-contacting components are tested against ISO 10993-5 (cytotoxicity) for biological safety.

    8. The Sample Size for the Training Set

    Not Applicable. There is no "training set" in the context of machine learning or AI models, as this device does not incorporate such algorithms for diagnostic or prognostic purposes.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As there is no training set as defined for AI or machine learning, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103753
    Date Cleared
    2011-02-15

    (54 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SIROLASER ADVANCE, SIROLASER XTEND, SIROLASER XTEND UPGRADED, FONALASER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue; marginal and interdental gingival and epithelial lining of free gingiva; frenectomy; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions; exposure of unerupted/partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; Fibroma removal; Gingival incision and excision; Treatment of canker sores; herpetic ulcers of the oral mucosa; Laser soft tissue curettage; Reduction of gingival hypertrophy.

    Device Description

    The SIROLaser Advance is a Diode Laser System for dental soft tissue surgery. The system is comprised of the laser unit with handpiece, laser fibers of different diameters, bendable tips, and a wireless foot control. The SIROLaser Advance has the following basic functions: Ability to emit laser radiation either in continuous wave mode (cw) or in chopped mode; Setup and display of treatment parameters: power, treatment time, chop frequency, duty cycle; Selection of predetermined settings for different indications; Upgradeable firmware.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    Acceptance Criteria and Device Performance for SIROLaser Advance

    Note: The provided document is a 510(k) summary, which generally focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive clinical study data with detailed acceptance criteria and performance metrics typically found in a clinical trial report. Therefore, direct "reported device performance" against explicit, quantifiable acceptance criteria is not provided in the same way it would be for a novel device undergoing a full clinical efficacy trial. The acceptance is typically based on meeting safety and performance standards equivalent to the predicate.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category/TypeAcceptance Criteria (Implicit from Predicate & Standards)Reported Device Performance (as per 510(k) Summary)
    SafetyEquivalent to predicate devices (Sirona SIROLaser Advance (K0092660) and Biolase iLase) and relevant safety standards."Performance testing to validate the safety... of the SIROLaser Advance includes electrical safety, electromagnetic compatibility..."
    Effectiveness/FunctionalityEquivalent to predicate devices for intended use (intra- and extra-oral soft tissue surgery including incision, excision, hemostasis, coagulation, and vaporization)."Performance testing to validate the... effectiveness of the SIROLaser Advance includes... validation testing of both hardware and software functions."
    The device possesses the reported basic functions:
    • Emit laser radiation in continuous wave or chopped mode
    • Setup and display of treatment parameters (power, time, frequency, duty cycle)
    • Selection of predetermined settings for different indications
    • Upgradable firmware |
      | Substantial Equivalence | Demonstrably similar in intended use, technology, design, and performance characteristics to legally marketed predicate devices, or differences do not raise new questions of safety or effectiveness. | The device is deemed "substantially equivalent" to Sirona SIROLaser Advance (K0092660) and Biolase iLase. |

    2. Sample Size for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a clinical performance study with a defined sample size for patients or data collected for demonstrating direct clinical efficacy.

    Instead, the submission relies on:

    • Performance Testing: This likely refers to engineering, electrical safety, electromagnetic compatibility (EMC), and software validation testing performed on the device itself or in controlled environments. The sample size for these tests would typically be a limited number of devices or components, not a patient population.
    • Comparison to Predicate Devices: The primary basis for clearance is demonstrating substantial equivalence to previously cleared devices. This implies that the clinical safety and effectiveness of the type of device (dental soft tissue laser) have already been established by the predicate devices.

    Therefore:

    • Sample size used for the test set: Not applicable in the context of a clinical test set from patient data. The relevant "sample" would be the device itself and its components for engineering and performance validation.
    • Data provenance: Not applicable for a clinical test set. The validation data would originate from the manufacturer's internal testing facilities (e.g., for electrical safety, EMC, software).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. Since there is no clinical "test set" with patient data requiring expert adjudication, there were no experts used to establish ground truth in this context. The 510(k) process for this type of device typically relies on engineering data and comparison to predicate devices, not de novo clinical trials with independent expert assessment of patient outcomes.


    4. Adjudication Method for the Test Set

    Not applicable. As no clinical test set requiring expert ground truth establishment is described, no adjudication method was employed.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.

    • The document does not mention any studies involving human readers or comparing AI-assisted vs. non-AI-assisted performance.
    • The SIROLaser Advance is a physical medical device (a laser system), not an AI/software as a medical device (SaMD) that typically relies on diagnostic imaging or data for interpretation where MRMC studies are relevant.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    Not applicable. The SIROLaser Advance is a hardware device for surgical procedures and does not rely on an "algorithm only" or AI component for standalone performance assessment in the way SaMDs do. The laser itself is the tool, operated by a human.


    7. Type of Ground Truth Used

    For the purpose of the 510(k) submission, the "ground truth" implicitly refers to:

    • Engineering Standards and Specifications: For electrical safety, EMC, and hardware/software functionality, the ground truth is defined by adherence to recognized international and national standards (e.g., IEC standards for medical electrical equipment) and the device's own design specifications.
    • Predicate Device Performance: The "ground truth" for clinical safety and effectiveness is largely established by the long-standing, safe, and effective use of the predicate devices (Sirona SIROLaser Advance (K0092660) and Biolase iLase) for the stated indications. The new device is deemed safe and effective because its performance is equivalent to these cleared devices.

    8. Sample Size for the Training Set

    Not applicable. This device is a laser system, not a machine learning or artificial intelligence model, therefore there is no "training set" in the context of algorithm development.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set mentioned or implied for an AI/ML algorithm, the concept of establishing ground truth for it does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092660
    Date Cleared
    2009-12-03

    (97 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    SIROLASER ADVANCE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue.

    Indications for use are the following applications: marginal and interdental gingiva and epithelial lining of free gingiva, frenectomy frenotomy, biopsy, operculectomy, Implant recovery, gingivectomy, gingivoplasty, gingival troughing crown lengthening, hemostasis of donor site removal of granulation tissue, laser assisted flap surgery debridement of deseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasty, excision of lesions, exposure of unerupted/partially erupted teeth, removal of hyperplastic tissues, treatment of aphthous ulcers, leukoplakia, sulcular debridement (removal of diseased or inflamed soft tissue, in the periodontal pocket), pulpotomy, pulpotomy as adjunct to root canal therapy.

    Device Description

    The SIROLaser Advance is a Diode Laser System for dental soft tissue surgery. The system is comprised of the laser unit with handpiece, laser fibers of different diameters, bendable tips, and a wireless foot control.

    The SIROLaser Advance has the following basic functions

    • Ability to emit laser radiation either in continuous wave mode (cw) or . chopped mode
    • Setup and display of treatment parameters: .
      • o power,
      • o treatment time,
      • chop frequency o
      • o duty cycle
    • Selection of predetermined settings for different indications ●
    • Upgradeable firmware .
    AI/ML Overview

    This document describes the Sirona Dental Systems SIROLaser Advance, a diode laser system for dental soft tissue surgery, and its substantial equivalence to predicate devices. However, it does not contain information about specific acceptance criteria, a study proving the device meets those criteria, or details regarding sample sizes, expert ground truth establishment, or clinical performance metrics as requested in your prompt.

    The 510(k) summary primarily focuses on:

    • Identification of the device: SIROLaser Advance
    • Intended Use: Various intra- and extra-oral soft tissue surgical applications.
    • Predicate Devices: SIROLaser, Kavo Gentle Ray 980, Ivoclar Odyssey Navigator, Ceramoptec Ceralas D15.
    • Device Description: Functions include continuous/chopped wave modes, parameter setup, predetermined settings, and upgradeable firmware.
    • Basis for Substantial Equivalence: Stated as performance testing for electrical safety, electromagnetic compatibility, and validation of hardware and software functions.

    Therefore, I cannot fulfill your request for the tables and specific details regarding acceptance criteria, study methodologies, sample sizes, and expert qualifications, as this information is not present in the provided text.

    The document is a 510(k) summary which claims substantial equivalence based on in vitro engineering and safety testing, not clinical performance studies with human subjects or a comparison of AI assistance. It falls under the regulatory pathway for medical devices (Class II) where performance is often demonstrated through engineering testing and comparison to existing devices, rather than clinical efficacy trials that would typically involve the detailed study design elements you've asked about.