Search Results
Found 7 results
510(k) Data Aggregation
(193 days)
Sofsilk Coated Braided Silk Suture
Sofsilk™ Coated Braided Silk Sutures are indicated for use in soft tissue approximation and/or ligation.
Sofsilk™ 6/0 and 7/0 black braided silk sutures are nonabsorbable, sterile, non-mutagenic surgical sutures composed of natural proteinaceous silk fibers called fibroin. This protein is derived from the domesticated silkworm species Bombyx mori of the family bombycidae. The silk fibers are treated to remove the naturally-occurring sericin gum and braided to produce Sofsilk™ surgical silk sutures. The braided sutures are coated uniformly with silicone to reduce capillarity and to increase surface lubricity, which enhances handling characteristics, ease of passage through tissue, and knot run-down properties. Sofsilk™ sutures are colored black with Logwood extract (21CFR73.1410). Sofsilk™ sutures meet all requirements established by the United States Pharmacopeia (USP) and the European Pharmacopeia (EP) for nonabsorbable surgical sutures, Sofsilk™ Coated Braided Silk Suture. Sofsilk™ sutures are indicated for use in soft tissue approximation and/or ligation.
The provided text is a 510(k) summary for the Sofsilk™ Coated Braided Silk Suture. It details the device's characteristics, indications for use, and a comparison to a predicate device. However, it does not contain information about an AI/ML-based device or its performance criteria.
Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, ground truth, or MRMC studies, as these aspects are not discussed in the provided document. The document describes a traditional medical device (surgical sutures) and its regulatory clearance based on substantial equivalence, primarily through bench testing for physical and mechanical performance, and biocompatibility.
The "study" mentioned refers to mechanical and physical bench tests of the suture, not a clinical or AI performance study.
Here's what the document does provide regarding the device's performance evaluation:
-
Acceptance Criteria (implied standards compliance):
- Meets or exceeds requirements of U.S.P. (United States Pharmacopeia) for nonabsorbable surgical sutures.
- Meets or exceeds requirements of E.P. (European Pharmacopeia) for nonabsorbable surgical sutures.
- Compliant with ISO 10993-1:2018 "Biological evaluation of medical devices-Part 1: Evaluation and testing" for biocompatibility.
- Sterility Assurance Level (SAL) of 10^-6 for ETO sterilization.
-
Reported Device Performance (bench tests):
- USP/EP diameter: Tested, found equivalent to predicate.
- USP needle attachment: Tested, found equivalent to predicate.
- USP tensile strength: Tested, found equivalent to predicate.
- Biocompatibility: Evaluated and confirmed compliant with ISO 10993-1 for its intended patient contact profile.
- Shelf life: Stability studies conducted, proposed 5-year shelf life demonstrated.
The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." This further confirms that there was no clinical study, let alone one involving AI/ML performance.
Ask a specific question about this device
(753 days)
SafePath Sulturing System-Silk Suture, SafePath Sulturing System Polyamide Suture (Nylon) SafePath Sulturing
SafePath Suturing System - Silk Suture: The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues.
SafePath Suturing System - Polyamide (Nylon) Suture: The SafePath Suturing System is intended for use in placement of a nylon suture in the skin and subcutaneous tissues.
SafePath Suturing System - Polypropylene Suture: The SafePath Suturing System is intended for use in placement of a polypropylene suture in the skin and subcutaneous tissues.
The silk suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of the organic protein, fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. The silk suture is silicone coated, braided and dyed (black) with logwood extract. The silk suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP size 2-0. The suture is provided as a 36 in. Jength, swaged on a single-armed 1½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
The nylon suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of long chain aliphatic polymers - nylon 6,6. SafePath nylon sutures are dyed with Logwood extract, FDA approval number §73.1410. The nylon suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a singlearmed ½ circle. 26 mm reverse cutting needle. The suture is pre-loaded in a handheld, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
The polypropylene suture provided with the SafePath Suturing System is a monofilament synthetic non-absorbable, sterile surgical suture composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin. The molecular formula is (C3H6)n. The Safepath suture is dyed with [Phthalocyaninato(2-)] copper, FDA approval number §74.3045. The polypropylene suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a single-armed ½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
Finished devices may be packaged individually, or in multi-unit cartons or procedure packs. The SafePath Suturing System is designed to place a suture in skin and subcutaneous tissue.
This document discusses the FDA clearance for the SafePath Suturing System, which includes silk, polyamide (nylon), and polypropylene sutures. However, the provided text does not contain information about a study based on AI/ML. All the performance criteria and testing mentioned relate to the physical and biological properties of the sutures and the suturing system, not to an AI component.
Therefore, I cannot answer your request with the specific details about acceptance criteria and studies related to AI/ML performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.
The document highlights the following non-clinical performance criteria and testing for the sutures:
1. A table of acceptance criteria and the reported device performance:
Acceptance Criteria (General) | Reported Device Performance |
---|---|
USP Nonabsorbable Sutures - Diameter | Met or exceeded USP criteria |
USP Sutures - Needle Attachment | Met or exceeded USP criteria |
USP Sutures - Tensile Strength | Met or exceeded USP criteria |
Biocompatibility - Cytotoxicity | Met test criteria |
Biocompatibility - Sensitization | Met test criteria |
Biocompatibility - Irritation | Met test criteria |
Biocompatibility - Systemic Toxicity | Met test criteria |
Biocompatibility - Hemocompatibility | Met test criteria |
Biocompatibility - Pyrogenicity | Met test criteria |
Biocompatibility - Endotoxicity | Met test criteria |
2. Sample size used for the test set and the data provenance: Not applicable as this is a non-AI/ML device. The testing was physical and biological, not based on data sets.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is a non-AI/ML device. Ground truth as typically understood in AI/ML is not relevant here.
4. Adjudication method for the test set: Not applicable as this is a non-AI/ML device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is a non-AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as this is a non-AI/ML device.
7. The type of ground truth used: For physical testing, ground truth would be established by the specifications of the USP standards. For biocompatibility, it's about meeting pre-defined safety thresholds as per recognized standards (e.g., ISO 10993).
8. The sample size for the training set: Not applicable as this is a non-AI/ML device.
9. How the ground truth for the training set was established: Not applicable as this is a non-AI/ML device.
The study described is a non-clinical performance testing study that evaluated the physical and biocompatibility characteristics of the sutures and the suturing system. The conclusions state that the device met or exceeded USP criteria for physical properties and met the test criteria for a battery of biocompatibility evaluations. The safety evaluation for the nylon and polypropylene sutures leveraged information from a previously cleared device (K180701) and used identical raw materials, packaging, and sterilization conditions as the predicate devices.
Ask a specific question about this device
(162 days)
Non absorbable Surgical Silk Suture
Non absorbable Surgical Silk Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic procedures, cardiovascular and neurological procedures. The device is limited to use where short term wound support (7-10 days) is required and can be left in place for a maximum of 10 days.
Non absorbable Surgical silk Suture is a nonabsorbable, braided surgical suture which is supplied sterile, surgical suture composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori of the family Bombycidae. Silk for braided material is processed to remove the natural waxes and gums. Silk suture is dyed black(Logwood extract) and coated with silicone. The quantity of color additive is not to exceed 1.00% by weight of the suture. The device will be offered in diameters ranging from USP size 6-0 through 2 and available in length varying from 45cm to 150cm with or without needles attached. Silk suture meets all the requirements of USP for nonabsorbable surgical suture. The suture needle are made from medical stainless steel, the grade of medical stainless steel is SUS 420 J2.
The provided document is a 510(k) premarket notification letter from the FDA regarding a Non-absorbable Surgical Silk Suture. This type of device is a physical medical product, not an AI/software device. Therefore, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" as it applies to AI/software performance metrics (like sensitivity, specificity, AUC, human reader improvement with AI assistance, etc.) is not applicable here.
Instead, for a physical medical device like a surgical suture, the "acceptance criteria" and "proof" relate to:
- Material properties: Does the suture material meet established standards (e.g., USP requirements for diameter, tensile strength)?
- Biocompatibility: Is the material safe for use in the human body (e.g., non-toxic, non-irritating, non-hemolytic)?
- Sterilization: Is the device effectively sterilized and free from harmful residuals?
- Packaging integrity: Is the packaging designed to maintain sterility until use?
- Performance: Does the device perform as intended (e.g., needle attachment strength)?
The document details the non-clinical tests performed to demonstrate substantial equivalence to a predicate device, as required for 510(k) clearance.
Here's a breakdown of the relevant information from the document, reinterpreting "acceptance criteria" for a physical device:
Acceptance Criteria and Device Performance for Non-absorbable Surgical Silk Suture (K232355)
The acceptance criteria for this medical device are based on compliance with various established standards and the demonstration of substantial equivalence to a legally marketed predicate device (K161633, REXSIL). The performance is evidenced by non-clinical testing.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Standard/Requirement | Reported Device Performance/Conclusion |
---|---|---|
Material Properties | USP Sutures - Diameter | "Comply with USP " |
USP Tensile Strength | "Comply with USP " | |
Suture Length | "Not less than 95.0% of the length stated on the label" | |
Performance (Device Assembly) | USP Sutures - Needle Attachment | "Comply with USP " |
Biocompatibility | ISO 10993-3:2014 (Genotoxicity, carcinogenicity, reproductive toxicity) | "The test result showed that the material and colorant of the proposed device will not have any adverse effects when used for up to 10 days." (General statement for ISO 10993 tests) |
ISO 10993-5:2009 (In vitro Cytotoxicity) | "The test result showed that the material and colorant of the proposed device will not have any adverse effects when used for up to 10 days." | |
ISO 10993-6:2016 (Local effects after implantation) | "The test result showed that the material and colorant of the proposed device will not have any adverse effects when used for up to 10 days." | |
ISO 10993-10:2010 (Irritation and skin sensitization) | "The test result showed that the material and colorant of the proposed device will not have any adverse effects when used for up to 10 days." | |
ISO 10993-11:2017 (Systemic Toxicity) | "The test result showed that the material and colorant of the proposed device will not have any adverse effects when used for up to 10 days." | |
ASTM F756-17 (Assessment of Hemolytic Properties) | "The test result showed that the material and colorant of the proposed device will not have any adverse effects when used for up to 10 days." | |
Sterilization & Endotoxins | ISO 10993-7:2008 (Ethylene oxide sterilization residuals) | Compliance stated in "Non-Clinical Test Conclusion" section. |
USP Pyrogen Test (USP Rabbit Test) | Compliance stated in "Non-Clinical Test Conclusion" section. | |
USP Bacterial Endotoxins Test | Compliance stated in "Non-Clinical Test Conclusion" section. | |
Packaging & Shelf Life | ASTM F88/F88M-15 (Seal Strength of Flexible Barrier Materials) | Compliance stated in "Non-Clinical Test Conclusion" section. |
ASTM F1929-15 (Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration) | Compliance stated in "Non-Clinical Test Conclusion" section. | |
Overall Substantial Equivalence (Non-Clinical) | Comparison to predicate device (K161633) characteristics and performance based on standards. | "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device." |
The following sections (2-9) are not applicable as they relate to AI/software performance studies, which are not relevant for this physical surgical suture device. The information provided in the document confirms that this is a traditional medical device submission for which clinical trials (and thus clinical data for training/testing or expert adjudication for ground truth) were not required.
2. Sample size used for the test set and the data provenance: Not applicable for this physical device. Testing would involve a statistically significant number of physical samples for each test (e.g., tensile strength, diameter, biocompatibility). Data provenance would be from laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical properties and biocompatibility is established by adherence to recognized international and national standards (e.g., ISO, ASTM, USP) and laboratory measurements by qualified technicians.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical device, not an AI algorithm.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this physical device, the "ground truth" is adherence to established engineering specifications, material standards, and biological safety standards (e.g., precise diameter measurements, specific tensile strength values, confirmed absence of cytotoxicity, etc.). These are objectively measured in laboratories according to validated test methods.
8. The sample size for the training set: Not applicable. This is not an AI/ML device requiring a training set.
9. How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(88 days)
NON-ABSORBABLE SILK SUTURE
The non-absorbable silk suture is intended for use in general soft tissue approximation, excluding use in cardiovascular, ophthalmic and neurological tissues.
The Arrow Non-Absorbable Silk Suture has the following characteristics: Black, braided silk suture thread of '000' (Metric 2) diameter, 30" (75 cm) suture thread length, Pre-attached, straight, stainless steel needle, 2-3/8" (60 mm) needle length
The provided document is a 510(k) summary for a Non-Absorbable Silk Suture. This type of device approval primarily relies on demonstrating substantial equivalence to a predicate device, rather than explicit acceptance criteria and performance studies for a novel AI or diagnostic device. As such, many of the requested categories are not applicable or cannot be extracted from this specific document.
Here's an attempt to address the questions based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in terms of specific numerical thresholds for performance metrics. Instead, "performance tests" were conducted to demonstrate substantial equivalence to predicate devices. The implicit acceptance criterion is that the new device's performance is comparable to or "favorable" compared to the predicate devices in these tests.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Suture Tensile Strength | Passed USP 29:2006 compliance; safe, effective, and performs favorably compared to predicates. |
Needle Attachment | Passed USP 29:2006 compliance; safe, effective, and performs favorably compared to predicates. |
Suture Diameter | Passed USP 29:2006 compliance; safe, effective, and performs favorably compared to predicates. |
Biocompatibility | Passed ISO 10993-1:2003 compliance; safe, effective, and performs favorably compared to predicates. |
2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes for the tensile strength, needle attachment, suture diameter, or biocompatibility tests. It also does not mention what would typically be considered "test sets" in the context of AI or diagnostic device studies, nor does it refer to "data provenance" (country of origin, retrospective/prospective data). These are bench tests performed on the physical device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to a submission for a physical medical device like a suture, which relies on standardized bench testing and biocompatibility assessments, not expert-derived ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Bench tests do not involve expert adjudication in the same way as clinical or AI performance studies.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a physical medical device, not an AI or diagnostic tool designed to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the performance tests, the "ground truth" is defined by the standards themselves (e.g., USP 29 and ISO 10993-1). The device's characteristics are measured against the specifications outlined in these standards and against the performance of predicate devices. For biocompatibility, it's about whether the material elicits an adverse biological response according to the ISO standard.
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device; there is no "training set."
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
Ask a specific question about this device
(79 days)
SILK SUTURES
Aurolab nonabsorbable Silk suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
This Silk suture is a nonabsorbable, sterile, surgical sutures composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae.Those dyed black are dyed with Hematein (logwood) black and the logwood extract conforms with 21 CFR 73.1410 and does not exceed 1.0% (W/W) of Suture.
This nonabsorbable suture is compossed of silk filaments that are braided or twisted in a suitable construction for the intended size to meet current USP specification.
This suture may be uncoated or have a silicone coating , a paraffin wax coating , or a natural gum coating ( Virgin Silk) . This sutures come with needles attached.
The provided document is a 510(k) summary for Aurolab's nonabsorbable Silk sutures. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed acceptance criteria and performance data as typically seen in AI/ML device submissions.
Here's an analysis based on the provided text, addressing the requested points where information is available:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria (Based on USP Standards) | Reported Device Performance (as stated) |
---|---|---|
Material Properties | Diameter meets USP specifications | Meets USP standards |
Mechanical Properties | Tensile strength meets USP specifications | Meets USP standards |
Device Integration | Suture-needle attachment meets USP specifications | Meets USP standards |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly mentioned. The document states that "tests according to methods presented in United States Pharmacopoeia (U.S.P) were conducted for diameter, tensile strength and Suture-needle attachment." USP monographs specify testing methods, which include sample sizes (e.g., 10 suture strands for specific tests), but the exact number of samples tested by Aurolab is not provided in this summary.
- Data Provenance: The tests were conducted internally by Aurolab in India. The data is retrospective in the sense that the summary is presenting results of tests already completed.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This is not applicable to the provided document. The "ground truth" for this medical device (nonabsorbable silk suture) is established by adherence to recognized industrial standards, specifically the United States Pharmacopoeia (USP). Experts in this context would be laboratory technicians and quality control personnel performing the physical and mechanical tests, not clinical experts establishing a medical "ground truth."
4. Adjudication Method
Not applicable. Performance is determined by objective measurements against predefined USP specifications, not by expert adjudication of clinical outcomes or images.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic or AI/ML devices where human interpretation is involved. For a surgical suture, the effectiveness is demonstrated through mechanical and material property testing and clinical use experience (not detailed in this summary).
6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study
Not applicable. This is a medical device (surgical suture), not an AI algorithm.
7. Type of Ground Truth Used
The "ground truth" for this device is based on United States Pharmacopoeia (USP) standards and specifications for nonabsorbable silk sutures. These are objective, measurable physical and chemical properties.
8. Sample Size for the Training Set
Not applicable, as this is not an AI/ML device.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as this is not an AI/ML device.
Summary of Acceptance Criteria and Study Design for the Suture:
The study conducted to demonstrate the substantial equivalence of the Aurolab nonabsorbable Silk sutures was a bench testing study focused on characterizing the physical and mechanical properties of the suture.
- Acceptance Criteria: The primary acceptance criteria were that the device's diameter, tensile strength, and suture-needle attachment met the specifications outlined in the United States Pharmacopoeia (USP) for nonabsorbable silk sutures.
- Study Design:
- Objective: To establish technical equivalency with predicate devices (Ethicon silk nonabsorbable sutures).
- Methodology: Performance tests were conducted in accordance with USP methodologies.
- Results: The tests performed showed that Aurolab devices meet USP standards and are technically equivalent to the predicate devices.
- Conclusion: Based on these tests, the device was deemed substantially equivalent to legally marketed predicate devices.
Ask a specific question about this device
(60 days)
SUTRASILK SILK SUTURE
SutraSilk nonabsorbable silk sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
SutraSilk silk suture is a nonabsorbable, sterile, surgical suture composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. Those dyed black are dyed with Hematein (logwood) black and the logwood extract conforms with 21 CFR 73.1410 and does not exceed 1.0% (W/W) of suture.
This non-absorbable suture is composed of silk filaments that are braided or twisted in a suitable construction for the intended size to meet current USP specifications.
The suture may be uncoated or have a silicone coating, a paraffin wax coating, or a natural gum coating (Virgin silk). The sutures come with needles attached.
Here's a breakdown of the requested information regarding the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion (Test) | Standard Met | Reported Device Performance |
---|---|---|
Diameter | United States Pharmacopeia (U.S.P.) standards | Meets USP standards |
Tensile Strength | United States Pharmacopeia (U.S.P.) standards | Meets USP standards |
Suture-Needle Attachment Strength | United States Pharmacopeia (U.S.P.) standards | Meets USP standards |
2. Sample size used for the test set and the data provenance
- Sample Size: The document does not specify the exact sample size used for the tests. It only states "tests according to methods presented in United States Pharmacopia (U.S.P.) were conducted".
- Data Provenance: Not specified. The document does not mention the country of origin of the data or whether it was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This information is not applicable as the study described is a performance test against established industry standards (USP), not a diagnostic device requiring expert interpretation for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. This study involves direct physical property measurements against predetermined standards, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is not an AI-powered diagnostic device, nor is it a study involving human readers or comparative effectiveness in that context. It's a performance study for a surgical suture.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an algorithm or AI device. It's a physical medical device (suture).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for the performance tests was the United States Pharmacopeia (U.S.P.) standards for diameter, tensile strength, and suture-needle attachment.
8. The sample size for the training set
- Not applicable. This is a performance study for a physical device, not an AI/machine learning model that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
Ask a specific question about this device
(260 days)
SILK SUTURE
Silk sutures are indicated for use as nonabsorbable sutures in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neural tissue.
Silk Non-Absorbable Surgical Sutures are composed of the organic protein fibroin which is derived from the domesticated species Bombyx mori (B.mori) of the family Bombycidae. Silk sutures are braided or twisted and available in white or dyed (black or blue). Braided sutures (except for size 8/0) are coated with silicone. Silk sutures are also available in various lengths, diameters and quantities with or without surgical needles. SILK sutures are not absorbed, progressive degradation of the proteinaceous silk fiber in-vivo may result in gradual loss of all the suture's tensile strength over time. The proposed Silk suture needle will be manufactured using an automated end stiffening process of applying the medical grade adhesive to a portion of the silk strand which is then cured and cut to length.
The provided text describes a 510(k) submission for a Silk Non-Absorbable Surgical Suture with UV-10 adhesive, focusing on the substantial equivalence to a previously cleared device. The key change is an automated needle-end stiffening process compared to a manual one.
Here's an analysis of the acceptance criteria and study information, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance | Comments |
---|---|---|
Biocompatibility (tested per ANSI/AAMI 10993 and USP guidelines) | Passed all specified tests: Cytotoxicity, Hemolysis, Sensitization, Acute Systemic Toxicity, Irritation, Intracutaneous. | Indicates the material and design are safe for biological contact. |
Material Composition (Organic protein fibroin derived from Bombyx mori) | Composed of organic protein fibroin from Bombyx mori. | Confirms the fundamental material is unchanged and consistent with predicate. |
Sterility | Stated to be sterile sutures. | Implies the sterilization process meets standards. |
Indications for Use (General soft tissue approximation and/or ligation, including cardiovascular, ophthalmic, and neural tissue) | Same indications for use as the predicate device. | Confirms the intended medical applications remain unchanged. |
Physical Characteristics (Lengths, diameters, quantities, colors, availability with/without needles) | Available in the same various lengths, diameters, quantities, colors and sizes, with and without surgical needles. | Confirms physical attributes are consistent with predicate. |
Packaging (Identical Tyvek/Mylar packaging) | Packaged in identical Tyvek/Mylar packaging. | Confirms packaging remains consistent. |
Tensile Strength Loss (in-vivo degradation) | "Progressive degradation of the proteinaceous silk fiber in-vivo may result in gradual loss of all the suture's tensile strength over time." | This is a known characteristic of silk sutures and is consistent with the predicate. It's not an "acceptance criterion" in the sense of a target value, but rather a descriptive characteristic. The acceptance is likely that this behavior is comparable to the predicate. |
Performance of automated needle end stiffening process (Implicit: should not negatively impact device performance compared to manual process) | Not explicitly detailed but implied by the "substantially equivalent" claim and the lack of specific performance testing beyond biocompatibility. | The biocompatibility testing likely serves as the primary evidence that the change in manufacturing process for the needle-end stiffening does not introduce new risks or alter fundamental material properties. There is no mention of comparative mechanical testing (e.g., knot strength, needle attachment strength) as an explicit acceptance criterion or reported performance. |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of a typical clinical study or a performance study with a defined sample size for the evaluation of the automated needle-end stiffening process. The primary evidence presented is biocompatibility testing.
- Sample size: Not explicitly stated for biocompatibility tests. It's usually a small number of samples required to demonstrate biological compatibility according to standards.
- Data provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, biocompatibility testing is typically conducted in a laboratory setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device submission primarily relies on biocompatibility testing and demonstrating substantial equivalence to a predicate device. There is no "ground truth" derived from expert review of cases in the way that might apply to a diagnostic imaging AI algorithm. The "ground truth" for biocompatibility is established by adherence to recognized standards (ANSI/AAMI 10993 and USP guidelines) and passing the specified tests.
4. Adjudication method for the test set
Not applicable. There is no "test set" that requires adjudication (e.g., clinical cases reviewed by experts).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a surgical suture, not a diagnostic device with an AI component.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a surgical suture, not a diagnostic device with an AI component.
7. The type of ground truth used
For the safety assessment, the "ground truth" is based on:
- Biological standards: Adherence to ANSI/AAMI 10993 and USP guidelines for biocompatibility.
- Predicate device characteristics: The known safety and performance profile of the legally marketed predicate Silk Non-Absorbable Surgical Suture (K930590). The new device is considered "substantially equivalent" if it shares the same safety and effectiveness profile.
8. The sample size for the training set
Not applicable. This is a physical medical device (suture), not a machine learning model.
9. How the ground truth for the training set was established
Not applicable. This is a physical medical device (suture), not a machine learning model.
Ask a specific question about this device
Page 1 of 1