Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K963540
    Date Cleared
    1996-09-27

    (23 days)

    Product Code
    Regulation Number
    862.1160
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Both electrolyte buffers are used to measure sodium, chloride, potassium, and carbon dioxide concentrations in serum or plasma, sodium, potassium, and chloride concentrations in urine, and chloride concentrations in CSF on the SYNCHRON CX 3 System

    Device Description

    The Sigma Diagnostics methods use ion selective electrodes for determining sodium, potassium, and chloride and rate of pH change for determing carbon dioxide on the SYNCHRON CX 3 System.

    AI/ML Overview

    The provided text describes the summary of safety and effectiveness for the Sigma Diagnostics CX®3 Electrolyte Buffer, Procedure Number E3766. It focuses on demonstrating substantial equivalence to a predicate device (Beckman Electrolyte Buffer Kit, Part No. 443325) through comparison studies, precision, and linearity.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to a predicate device, focusing on correlation coefficients, regression equations indicating agreement, and acceptable levels of precision and linearity.

    Test AspectAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Correlation to Predicate (Serum)High correlation coefficients (e.g., typically >0.95 for quantitative assays)Sodium: 0.963Potassium: 0.999Chloride: 0.973Total CO₂: 0.992
    Regression to Predicate (Serum)Regression equation close to y = 1x + 0 (i.e., slope near 1, intercept near 0)Sodium: y = 0.90x + 13.48Potassium: y = 0.98x + 0.08Chloride: y = 0.91x + 10.52Total CO₂: y = 0.95x + 1.54
    Correlation to Predicate (Urine)High correlation coefficients (e.g., typically >0.95 for quantitative assays)Sodium: 0.997Potassium: 0.998Chloride: 0.997
    Regression to Predicate (Urine)Regression equation close to y = 1x + 0Sodium: y = 0.95x - 0.10Potassium: y = 0.97x + 0.49Chloride: y = 0.97x + 2.06
    Correlation to Predicate (CSF)High correlation coefficient (e.g., typically >0.95)Chloride: 0.891
    Regression to Predicate (CSF)Regression equation close to y = 1x + 0Chloride: y = 0.98x + 4.00
    Precision (Serum)Low %CVs (e.g., typically <5%, often <2% for electrolytes)Sodium (%CV): <1.1%Potassium (%CV): <1.0%Chloride (%CV): <1.3%
    Precision (Urine)Low %CVsSodium (%CV): <1.2%Potassium (%CV): <1.6%Chloride (%CV): <1.9%
    Precision (CSF)Low %CVsChloride (%CV): <1.8%
    Linearity (Serum)Accurate measurement across a clinically relevant rangeSodium: 100.0 - 200.0 mmol/LPotassium: 1.00 - 15.00 mmol/LChloride: 50.0 - 200.0 mmol/LTotal CO2: 5.0 - 40.0 mmol/L
    Linearity (Urine)Accurate measurement across a clinically relevant rangeSodium: 10 - 200 mmol/LPotassium: 2 - 200 mmol/LChloride: 15 - 300 mmol/L
    Linearity (CSF)Accurate measurement across a clinically relevant rangeChloride: 50 - 200 mmol/L

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not explicitly state the sample size for the comparison studies. It only presents the correlation coefficients and regression equations, implying a dataset was used without specifying its size.
    • Data Provenance: Not specified. It's unclear if the data is from a particular country, or if it's retrospective or prospective. Given the context of a 510(k) submission in 1996 for a U.S. company, it's likely that the data would be derived from U.S. sources, but this is an inference, not a stated fact.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this type of device and study. The "ground truth" for an electrolyte analyzer is typically established by measurements from a reference method or a legally marketed predicate device, not by expert consensus in the same way an imaging or diagnostic AI interpretation device would. In this case, the predicate device (Beckman Electrolyte Buffer Kit, Part No. 443325) served as the reference for comparison.

    4. Adjudication method for the test set

    Not applicable. As noted above, this study is comparing a new device to a predicate device, not involving human interpretation or adjudication in the typical sense.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a study for an in vitro diagnostic (IVD) reagent, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in a sense, the entire study is "standalone" in that it assesses the performance of the Sigma Diagnostics Electrolyte Buffer on the SYNCHRON CX 3 System as an automated measurement system, without human intervention in the analytical measurement itself. The results (correlation, precision, linearity) directly reflect the algorithm's/device's performance.

    7. The type of ground truth used

    The "ground truth" was established by measurements obtained from a legally marketed predicate device, the Beckman Electrolyte Buffer Kit, Part No. 443325. This is a common and accepted method for demonstrating substantial equivalence for IVD devices.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML device that requires a "training set" in the conventional sense. The "training" or development of the reagent and its methods would have occurred internally at Sigma Diagnostics, but that's not a data set used to train an algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for an AI/ML algorithm in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1