Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    ), HDR Compatible Tandem and Ovoid Applicators (Henschke), HDR CT Compatible Split Ring Applicator, Segmented
    Vaginal Applicator, HDR Miami Applicator HDR Brachytherapy Applicator, 2/3 Endometrial Applicator Sets

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    2/3 Endometrial Applicator Sets: A manual radionuclide applicator system, a manually operated device intended to apply a radionuclide source into the body or to the surface of the body for radiation therapy.

    Segmented Vaginal Applicator: The use of sealed Radioisotopes to treat tumors within the body has been documented and published since the turn of the century. Modem era Radiation Therapy has developed delivery systems using isotopes of Cesium, Iridium, Iodine, and Gold to name a few examples. Many tumors are now treated by internal exposure to radiation emitted from sealed radioactive sources.

    Two common modalities of this are Low Dose Rate remote afterloading (Brachytherapy). One common use of Brachytherapy is in the treatment of cancer of the vaginal process. The system described in this 510(k) has been developed to function as an applicator for the positioning of sealed sources in the intracavitary treatment of the vagina.

    HDR Miami Applicator: This applicator is designed as an accessory to the Varisource System (Varian Associates K952913) and the Gammamed System (K89113) which uses a single radioactive source of Iridium-192 to treat cancer in a wide range of body sites. The Miami Applicator is placed in the vicinity of the cervix via the vagina just as described for the predicate device (Nucletron Miami Vaginal Applicator, K953946) and different diameter sleeves and intrauterine tubes, can be optimized to best meet the clinical needs of the patient along with minimization of dose to the mucosa.

    HDR Compatible Tandern and Ovoid Applicators: The applicators presented in this 510(k) notification have been developed to function as Applicators for the positioning of HDR Remote After-Loader sealed sources in the intra-cavitary treatment of cancer of the vagina and cervix.

    CT Compatible F/S/D Applicators: CT Compatible F/S/D applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of cancer in the cervix and uterus is an accepted clinical practice.

    HDR CT Compatible Split Ring Applicator: The HDR CT Compatible Split Ring type applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of cancer in the cervix and uterus is an accepted clinical practice.

    CCT HDR Tandem Ring Applicator with Rectal Retractor: The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/ Ring Applicator with Rectal Retractor is indicated for High Dose Rate irradiation of the uterus and cervix.

    Device Description

    2/3 Channel Endometrial Applicator Sets: The 2/3 Channel Endometrial Applicator is designed for the treatment of the endometrium and cervix. The applicator set includes 3 intrauterine tubes that are available in 3 cm and 5 cm configurations. These intrauterine tubes can be assembled using only 2 or all 3 intrauterine tubes, depending on the patient requirements. The only change to the device will be the labeling that the device is now MR Conditional.

    Henschke Applicator Set: The Henschke Applicator is based on "Henschke style" geometry for brachytherapy of the endometrium and cervix. This applicator is similar in design to the Fletcher Applicator with the exception of semispherical ovoids. These ovoids are ideal for patients with narrow vaginas due to their small size, the ovoid tube is closer to the surface on the side of the cervix while maintaining distance within the fornices. The tandem is inserted into the endometrium at a pre-determined depth and secured in place with the cervical stop. The only change to the device will be the labeling that the device is now MR Conditional.

    Fletcher (FSD) Applicator Set: The Fletcher Applicators are based on "Fletcher style" geometry for brachytherapy of the endometrium and cervix. The ovoids are cylindrical to create equal spacing in the vaginal fornices for a symmetrical dose distribution in this area. The tandem is inserted to into the endometrium at a pre-determined depth and secured in place with the cervical stop. The only change to the device will be the labeling that the device is now MR Conditional.

    Miami Applicator Set: The Miami Applicator is designed for treatment of the vagina and cervix and includes an intrauterine tube for the treatment of the endometrium, as required. The design incorporates 7 treatment channels around the periphery of the cylinder body and includes a center channel that can accommodate an intrauterine tube. The intrauterine tubes are available with either a 30° angled tip or in a straight (0° angle) configuration. A stump plug is provided to seal the end of the applicator when an intrauterine tube is not required. Build-up caps are available for the cylinder body to provide added spacing between the mucosa and radioactive source. This applicator provides additional treatment options, with the peripheral channels, to increase the dose to the target area while limiting the dose to normal tissue. The only change to the device will be the labeling that the device is now MR Conditional.

    Ring & Tandem Applicator Sets: The Ring & Tandem Applicator is based on the Stockholm technique for brachytherapy of the endometrium and cervix. The applicator consists of a ring tube and intrauterine tube that, when combined, create a fixed geometry and form a 90° angle. Build up caps of different diameters are provided for the ring tube that provide additional spacing between the radioactive source and mucosa. Including the Vienna Ring and Tandem Accessory Kit. The only change to the device will be the labeling that the device is now MR Conditional.

    Segmented Vaginal Applicator Set: The Segmented Vaginal Applicator is designed for treatment of the vagina and cervix and includes an intrauterine tube for the treatment of the endometrium, as required. The intrauterine tubes are available in different angles and lengths to meet anatomical requirements. The four individual segment design allows the applicator length to be adjustable based on the treatment volume. Additionally, the segments are available in different diameters to match the patient's anatomy. The only change to the device will be the labeling that the device is now MR Conditional.

    Split Ring Applicator Sets: The Splicator's patented design combines the benefits of several other intracavitary applicators. The ring tubes can be configured as a closed/complete ring or separated/" split' into four different distances, symmetrically or asymmetrically with the spacing bracket. Build-up caps of different diameters fit over the ring tubes to provide additional tissue spacing distance between the radioactive source and the cervix. The only change to the device will be the labeling that the device is now MR Conditional.

    Vienna System: The Mick Radio-Nuclear Instruments, Inc. Vienna System is intended to be used as an accessory to the Mick CT HDR Tandem I Ring Applicator and is indicated for High Dose Rate irradiation of the uterus and cervix. The Vienna System consists of perforated Build-Up Caps and complementary Needle Collectors which connect to the Mick CT HDR Tandem Ring Applicator. Pre-bent interstitial needles are intended to be used with the Vienna System but they are not manufactured by Mick Radio-Nuclear Instruments. Inc. and are not part of this submission.

    Build-Up Caps: When used with the CT HDR Tandem/Ring Applicator, the Vienna Build-Up Caps enable the introduction of up to nine (9) interstitial needles around the circumference of the ring to enhance the standard HDR treatment. The Vienna System is not designed to be used with any Rectal Retractor due to the introduction of the interstitial needles.

    Needle Collectors: When used in conjunction with the Vienna Build-Up Caps, the Needle Collectors will maintain and control the positioning of an array of up to nine (9) interstitial needles. The Needle Collectors are part of the Vienna System and as such, are not designed to be used with any Rectal Retractor.

    The Vienna System is designed to be used as an accessory with the Applicator and this does not alter the indications for use and is MR Conditional.

    AI/ML Overview

    The provided text describes a Special 510(k) Premarket Notification for various radionuclide applicator systems. The core purpose of this submission is to add MR Conditional labeling to existing devices, meaning the devices are safe for use in specific MR environments. The letter explicitly states that there have been no changes to the devices themselves or their intended use, only to their labeling.

    Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are related to demonstrating that the existing devices are indeed MR Conditional and comply with relevant safety standards for MR environments, specifically regarding radiofrequency-induced heating.

    Based on the provided information, here's an analysis to address your request:

    Acceptance Criteria and Study to Prove MR-Conditional Status

    The primary acceptance criteria for this submission is to demonstrate that the various applicator sets are "MR Conditional" as per relevant standards (which are not explicitly detailed in the document but implied by the testing). This typically means ensuring that the device does not pose a safety hazard (e.g., excessive heating, significant artifact, or device malfunction) when exposed to a specified MR environment.

    The study that proves the device meets the acceptance criteria for MR Conditional labeling is "Computational Modeling of Mick Radio-Nuclear Instruments HDR Brachytherapy Intracavitary Applicators for Radiofrequency-Induced Heating Evaluation Final Report."

    Here's how the information provided aligns with your specific points:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implicit for MR Conditional Labeling)Reported Device Performance and Compliance
    Device is safe for use in a specified MR environment (no significant radiofrequency-induced heating).Results = MR Conditional. The submission states, "Computational Modeling of Mick Radio-Nuclear Instruments HDR Brachytherapy Intracavitary Applicators for Radiofrequency-Induced Heating Evaluation Final Report. Results = MR Conditional." This indicates the computational modeling demonstrated satisfactory performance regarding RF-induced heating.
    Device integrity and function are maintained in specified MR environment.Not explicitly detailed in the "Performance Test Results" section, but implicitly covered by the "MR Conditional" result. The document states "no new issues of safety or effectiveness are raised."
    Minimal or acceptable image artifact in MR environment.Not explicitly detailed in the "Performance Test Results" section; the focus of the reported study was "Radiofrequency-Induced Heating Evaluation." This is a common aspect of MR Conditional testing but not explicitly stated as having been evaluated in the provided text.
    Proper labeling and instructions for use reflecting MR Conditional status.The purpose of the Special 510(k) is to "add MR Conditional labeling" and modify "Instructions for Use for each of the listed devices to add MR Conditional labeling to each of them." This implies the acceptance criteria for labeling will be met by including this information.
    No new issues of safety or effectiveness are raised by the modification (i.e., adding MR labeling).The submission explicitly states, "No new issues of biocompatibility are raised with regard to the modification of the devices for MR Conditional labeling." and "By introducing this device, no new issues of safety or effectiveness are raised." This serves as a key acceptance criterion.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: This was a computational modeling study, not a study involving a "test set" of patients or physical devices in the traditional sense for performance evaluation (e.g., clinical images). The "sample size" would refer to the computational models of the various applicator sets. The document references the "GYN 1 Applicator Family" and "GYN II Applicator Family" as the devices being re-labeled, which includes multiple distinct applicators (e.g., CT/MR Fletcher, CT/MR Henschke FSD, CT/MR Split Ring, CT/MR Segmented Vaginal, CT/MR Miami, CT/MR 2/3 Channel Endometrial, CT/MR Ring & Tandem). Each of these would have been computationally modeled.
    • Data Provenance: The study was a "Computational Modeling" study which is a form of in silico (computer simulation) analysis. The document does not specify the country of origin of the modeling data or if it was retrospective/prospective in a clinical sense. Given it's a 510(k) submission to the FDA, the assessment was likely conducted to meet US regulatory requirements.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • This was a computational modeling study for MR safety, not a study relying on expert human readers to establish ground truth for clinical diagnostic purposes. Therefore, the concept of "experts establishing ground truth for the test set" (e.g., radiologists) is not applicable here. MR safety testing (including computational modeling) is typically conducted by engineers and physicists specializing in MR compatibility, following established standards (e.g., ASTM, ISO). The document does not specify the number or qualifications of the individuals who performed or validated the computational modeling.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. As noted, this was a computational modeling study focused on MR safety (RF-induced heating). Adjudication methods like 2+1 or 3+1 are used in clinical image interpretation studies where human reviewers disagree on findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This was a submission for MR Conditional labeling of medical devices (applicators), not an AI-assisted diagnostic tool. Therefore, an MRMC study and AI performance metrics are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable as this is not an algorithm/AI device. The "standalone" performance here relates to the inherent physical properties of the device in an MR environment, specifically its heating characteristics, which was assessed via computational modeling.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for MR Conditional status in a computational modeling study is derived from established physics principles of electromagnetic interaction with materials, validated computational methods, and adherence to relevant international standards for MR safety testing (e.g., ASTM F2182 for RF-induced heating, ISO/TS 10974). The "Final Report" of the computational modeling likely presents this "ground truth" in terms of calculated temperature increases within the simulated device and surrounding tissue.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of this device and its assessment for MR Conditional labeling. This is not a machine learning or AI device.

    9. How the ground truth for the training set was established

    • Not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K993472
    Date Cleared
    2000-01-12

    (90 days)

    Product Code
    Regulation Number
    892.5700
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SEGMENTED VAGINAL APPLICATOR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mick Radio-Nuclear Instruments, Inc. Segmented Vaginal Applicators are intended for use in Brachytherapy. The system described in this 510(k) has been developed to function as an applicator for the positioning of sealed sources in the intracavitary treatment of the vagina.

    Device Description

    The Mick Radio-Nuclear Segmented Vaginal Applicator meets these requirements by providing an system that can be adjusted in length and diameter to meet the dimensions of the treatment volume and by utilizing radio opaque markers for visualization.

    AI/ML Overview

    This submission K993472 is for a Segmented Vaginal Applicator, which is a medical device used in brachytherapy. The submission asserts substantial equivalence to a predicate device and does not involve a study to demonstrate performance against specific acceptance criteria.

    Therefore, most of the requested information cannot be extracted from the provided text as the application did not involve a performance study with acceptance criteria.

    Here's what can be stated based on the given document:

    1. A table of acceptance criteria and the reported device performance: Not applicable. The submission asserts substantial equivalence based on design, construction, materials, intended use, and performance characteristics being similar to a predicate device, rather than providing specific performance metrics against pre-defined acceptance criteria for a new study.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. No test set or data provenance is mentioned as this was not a clinical performance study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. No ground truth establishment for a test set is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No test set adjudication is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an applicator for radiation therapy, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. No ground truth was established for a performance study. The "ground truth" for this submission is the established safety and effectiveness of the predicate device to which it claims substantial equivalence.

    8. The sample size for the training set: Not applicable. This device is not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established: Not applicable.

    The core argument for the device's marketability lies in its substantial equivalence to a previously cleared device, not in a new performance study with specific acceptance criteria. The document states:
    "This device is similar in design and construction, utilizes the identical materials, and has the same intended use and performance characteristics to the predicate devices. No new issues of safety or effectiveness are introduced by using this device."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1