Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K970188
    Date Cleared
    1997-02-20

    (34 days)

    Product Code
    Regulation Number
    862.1625
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SEALITE SCIENCES, INC. AQUALITE PROLACTIN

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AquaLite® Prolactin Bioluminescent Immunoassay (BIA) Kit (or the AquaLite® Prolactin assay) is intended to be used in clinical laboratories for the quantitative determination of human prolactin levels in sera and plasma. The AquaLite® Prolactin assay is for in vitro diagnostic use.

    Device Description

    The AquaLite® Prolactin Bioluminescent Immunoassay Kit uses a mixture of mouse monoclonal and rabbit polyclonal with anti-prolactin activity that is pre-coated onto polystyrene tubes (solid phase). Samples (serum or plasma), or appropriate calibrators or controls, are pipetted (25 uL) into the pre-coated tubes. Anti-prolactin conjugate consisting of mouse monoclonal antibody covalently linked to AquaLite® (150µL) is then added to the tubes. The conjugate uses the photoprotein, AquaLite® (recombinant aequorin: U.S. Patent Nos. 5,422,266 and 5,486,455) which is covalently linked to an anti-prolactin polyclonal antibody. Prolactin in the sample simultaneously combines with polyclonal antibody on the solid phase and conjugate antibody to form an immune complex or "sandwich" bound to the solid phase. Complex formation is complete after a 60-minute incubation period at room temperature (18° to 25°C) on a standard orbital shaker. The tubes are then washed to remove unbound conjugate.

    The washed tubes are placed in a luminometer that is capable of reading a triggered. flash-type reaction in 12 x 75 mm tubes. An injected calcium trigger solution causes AquaLite® to oxide its self-contained luciferin molecule. This reaction produces a twosecond flash of light at 469 nm, which is measured by the luminometer. The intensity of the light is directly proportional to the concentration of the prolactin in the sample. To calculate results, the luminometer uses a cubic spline curve fit applied to a logit-log transformation of the light intensity (in relative light units, RLU) of the prolactin calibrators versus prolactin concentration (in ng/mL).

    Note: Samples that generate signals greater than the signal from the highest calibrator are off-scale. These samples must be diluted and re-assayed. Remember to multiply the results from diluted samples by the dilution factor used.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the SeaLite Sciences, Inc. AquaLite® Prolactin Bioluminescent Immunoassay (BIA) Kit:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Metric (Unit)Acceptance CriterionReported Device Performance
    SensitivityDetection Limit (ng/mL)Not explicitly stated (implied to be low)0.01 ng/mL
    SpecificityCross-reactivity (%)Not explicitly stated (implied to be low for non-prolactin hormones)rh:
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1