510(k) Data Aggregation

    K Number
    K113719
    Date Cleared
    2012-04-20

    (123 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The sets are intended to deliver liquid nutrition formulas or water to an enteral access device (a feeding tube).

    Device Description

    The subject of this 510(k) is a modified proximal connector for Abbott Nutrition enteral feeding sets. The proximal connector connects the tubing of the enteral feeding set to the source of the nutrition (i.e., the container of the enteral nutritional formula). The modified design is the Safety Screw Connector and is compliant with the new AAMI/ISO Standard 80369-1 (Small-bore connectors for liguids and gases in healthcare applications – Part 1: General Requirements). This same design is being used for currently marketed competitive enteral nutrition sets.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the "Safety Screw Connector," a component of enteral feeding sets. It describes the device, its intended use, and the performance tests conducted to demonstrate substantial equivalence to existing devices.

    Here's an analysis of the acceptance criteria and study data based on the provided text, structured as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Non-interconnectable CharacteristicsCompliance with ANSI/AAMI/ISO 80369-1:2010 for preventing misconnection with other types of connectors."Mechanical test for verifying non-interconnectable characteristics of Safety Screw Connector and other connectors (ANSI/AAMI/ISO 80369-1:2010)" was conducted. The conclusion states, "The new design reduces the risk of unintentional misconnection to another type of tube." This implies the connector met the standard's requirements for non-interconnectability with other types of connectors.
    Delivery Accuracy & Feeding System PerformanceThe device should not negatively impact the accurate delivery of liquid nutrition formulas or water."Delivery accuracy and feeding system performance assessment testing of Safety Screw Connector" was conducted. The conclusion states, "Performance testing showed that the new designs are equivalent to currently marketed devices." This implies that the device maintained delivery accuracy and overall feeding system performance comparable to existing systems.
    BiocompatibilityThe device materials should be biocompatible and not cause adverse biological reactions when in contact with fluids."Biocompatibility testing for the Safety Screw Connector and the RTH Adapter Cap" was conducted. While no specific results are detailed, the general conclusion that "new designs are equivalent to currently marketed devices" and the statement "the new connector does not raise new questions of safety and effectiveness" suggest the biocompatibility criteria were met.
    Tensile Strength (Bond Junction)The bond between the Safety Screw Connector and the tubing should be strong enough to withstand expected forces without separation."Tensile strength testing for the bond junction between the Safety Screw Connector and tubing" was conducted. Although no specific numerical result is provided, the conclusion "new designs are equivalent to currently marketed devices" indicates that the tensile strength of the bond junction met the required performance standards for safety and effectiveness, similar to predicate devices.
    Overall Safety and EffectivenessThe device should be as safe and effective as legally marketed predicate devices and introduce no new safety concerns.The document explicitly states: "The new design reduces the risk of unintentional misconnection to another type of tube. The risk analysis showed that the new connector does not raise new questions of safety and effectiveness. Performance testing showed that the new designs are equivalent to currently marketed devices." This is the overarching acceptance criterion, concluding that the device is substantially equivalent to predicate devices, thus meeting safety and effectiveness expectations.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each test (mechanical, delivery accuracy, biocompatibility, tensile strength). Nor does it specify the data provenance (e.g., country of origin, retrospective or prospective) for these performance tests. It implies these tests were conducted as part of the device development and validation by Abbott Nutrition/Abbott Laboratories.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not provided in the given text. The tests described (mechanical, delivery accuracy, biocompatibility, tensile strength) are typically objective engineering and laboratory tests, not requiring expert "ground truth" in the way a diagnostic AI might. The "ground truth" for these tests would be the established specifications and standards (e.g., ANSI/AAMI/ISO 80369-1:2010), and the results are quantitatively measured and compared against these benchmarks.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. The performance tests are objective laboratory and engineering tests, not involving subjective interpretation or adjudication by multiple human observers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not applicable as this device is a physical connector, not an AI-powered diagnostic or assistive tool for human readers. Therefore, no information on human reader improvement with or without AI assistance is provided or relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable as the device is a physical connector, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The ground truth for the performance tests was based on established engineering standards and specifications (e.g., ANSI/AAMI/ISO 80369-1:2010) against which the device's measured performance was compared. For biocompatibility, the ISO 10993 series would typically serve as the ground truth framework. For delivery accuracy and tensile strength, the ground truth would be the performance of legally marketed predicate devices and internal/industry specifications deemed safe and effective.

    8. The Sample Size for the Training Set

    This question is not applicable. The Safety Screw Connector is a physical medical device, not a machine learning or AI model. Therefore, there is no "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no training set for this type of device.