Search Results
Found 1 results
510(k) Data Aggregation
(29 days)
SAFECARE**®** THC Urine Strip Test
SAFECARE® THC Urine Strip Test is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Marijuana in human urine at the cutoff concentrations of 50 ng/mL.
The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
The test is intended for over-the-counter use.
SAFECARE® THC Urine Strip Test devices are immunochromatographic assays for the qualitative detection of 11-nor-A9-THC-9 COOH (target analyte) in human urine. The product is a single-use in vitro diagnostic device. It contains a Test Device and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.
The provided document outlines the acceptance criteria and performance data for the SAFECARE® THC Urine Strip Test, which is a qualitative lateral flow immunoassay for detecting Marijuana in human urine.
Here's the breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" for precision or accuracy using specific numerical thresholds (e.g., >95% agreement). Instead, it presents the results of various performance studies. The overall acceptance is inferred from the device being deemed "substantially equivalent" to a predicate device.
However, based on the performance characteristics, we can infer performance goals for accuracy at different concentrations relative to the cutoff.
Acceptance Criteria Category | Specific Metric (Inferred) | Acceptance Threshold (Inferred/Observed) | Reported Device Performance | Comments on Performance |
---|---|---|---|---|
Analytical Performance | ||||
Precision | Agreement at various | High agreement expected, especially | ||
concentrations | at concentrations further from cutoff. | |||
-100% Cut-off | 100% Negative (0% Positive) | Lot 1, 2, 3: 50-/0+ | Meets expectation | |
-75% Cut-off | 100% Negative (0% Positive) | Lot 1, 2, 3: 50-/0+ | Meets expectation | |
-50% Cut-off | 100% Negative (0% Positive) | Lot 1, 2, 3: 50-/0+ | Meets expectation | |
-25% Cut-off | 100% Negative (0% Positive) | Lot 1, 2, 3: 50-/0+ | Meets expectation | |
Cut-off (50 ng/mL) | ~50% Positive, ~50% Negative | Lot 1, 2: 25-/25+; Lot 3: | Meets expectation | |
26-/24+ | ||||
+25% Cut-off | 100% Positive (0% Negative) | Lot 1, 2, 3: 50+/0- | Meets expectation | |
+50% Cut-off | 100% Positive (0% Negative) | Lot 1, 2, 3: 50+/0- | Meets expectation | |
+75% Cut-off | 100% Positive (0% Negative) | Lot 1, 2, 3: 50+/0- | Meets expectation | |
+100% Cut-off | 100% Positive (0% Negative) | Lot 1, 2, 3: 50+/0- | Meets expectation | |
Interference | No interference from | No false positives or negatives due to | Compounds at 100µg/mL | No interference reported |
common substances | specified interfering substances when | (list provided) showed no | ||
drug is absent or present above cutoff | interference. | |||
Specificity | Cross-reactivity with | Low or no cross-reactivity with | Provided specific cross- | Demonstrated specificity |
related compounds | non-target compounds. | reactivity percentages for | ||
related cannabinoids. | ||||
Urine Specific Gravity | Correct results across | Correct classification (negative for | Samples from 1.000 to 1.035 | Device performs reliably |
range of SG | -25% cutoff, positive for +25% cutoff). | with THC at +/-25% cutoff | across range of urine SG | |
showed expected results. | ||||
Urine pH | Correct results across | Correct classification (negative for | Samples from pH 4 to 9 with | Device performs reliably |
range of pH | -25% cutoff, positive for +25% cutoff). | THC at +/-25% cutoff | across range of urine pH | |
showed expected results. | ||||
Comparison Studies | Agreement with LC/MS | High agreement between device results | ||
for clinical samples | and LC/MS, particularly for samples | |||
not near the cutoff. | ||||
Overall Agreement | Not explicitly quantified, but inferred | Viewer A: 78/80 (97.5%) | High agreement for | |
from discordant results. | Viewer B: 78/80 (97.5%) | human readers | ||
Viewer C: 79/80 (98.75%) | ||||
Lay-User Study | Ease of use | Instructions are easily understood. | All lay users indicated | Instructions are clear |
instructions were easy to | ||||
Accuracy of self-testing | High percentage of correct results, | |||
decreasing near the cutoff. | ||||
-100% Cutoff | 100% agreement | 100% Correct | Excellent | |
-75% Cutoff | 100% agreement | 100% Correct | Excellent | |
-50% Cutoff | 100% agreement | 100% Correct | Excellent | |
-25% Cutoff | >90% agreement, with expected | 90% Correct | Acceptable, slight dip | |
increase in discordance. | near cutoff. | |||
+25% Cutoff | >90% agreement, with expected | 95% Correct | Acceptable, slight dip | |
increase in discordance. | near cutoff. | |||
+50% Cutoff | 100% agreement | 100% Correct | Excellent | |
+75% Cutoff | 100% agreement | 100% Correct | Excellent |
2. Sample Size Used for the Test Set and Data Provenance
- Precision Study:
- Sample Size: 9 different concentrations, with 50 tests performed for each concentration per lot (2 runs per day for 25 days). Since 3 lots were tested, this amounts to 9 concentrations * 50 tests/lot * 3 lots = 1350 tests in total for the precision study itself. The samples were prepared by spiking drug in "negative samples." The provenance is not explicitly stated but implies laboratory-prepared samples.
- Comparison Studies (Clinical Samples):
- Sample Size: 80 unaltered clinical samples (40 negative and 40 positive).
- Data Provenance: The document states "unaltered clinical samples." The country of origin for these clinical samples is not specified. It is a retrospective analysis as the samples were collected and then tested.
- Lay-User Study:
- Sample Size: 140 lay persons tested individual samples. There were 7 concentration levels, with 20 samples per concentration. So, 7 concentrations * 20 samples/concentration = 140 samples in total.
- Data Provenance: The samples were "spiked drug THC into drug free-pooled urine specimens." The country of origin is not specified. It is a prospective study in the sense that the lay users tested the devices with prepared samples, but the samples themselves were laboratory-prepared.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Precision Study: The ground truth was established by laboratory preparation of samples with known concentrations confirmed by LC/MS. No human experts are explicitly mentioned for ground truth.
- Comparison Studies (Clinical Samples): The ground truth was established by LC/MS results. "LC/MS is the preferred confirmatory method." The number of LC/MS operators or analysts is not specified, nor are their qualifications.
- Lay-User Study: The ground truth for the prepared urine samples was established by LC/MS. The number of LC/MS operators or analysts is not specified, nor are their qualifications.
4. Adjudication Method for the Test Set
- Precision Study: Not applicable, as results were quantitative (known concentrations) rather than subjective interpretations needing adjudication.
- Comparison Studies (Clinical Samples): The document mentions "three different laboratory assistants" performing the tests. Their results were compared to LC/MS. There is no explicit adjudication method stated for discrepancies between human readers or between human readers and LC/MS. The discordant results table simply lists them.
- Lay-User Study: Not explicitly stated. Each lay person provided their own result. Their individual results were then compared to the LC/MS confirmed concentration.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
- No, an MRMC comparative effectiveness study was not done.
- This device is a standalone diagnostic strip test intended for visual interpretation, not an AI-assisted diagnostic tool. Therefore, there is no AI component or human readers improving with/without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
- Yes, in essence, the "device performance" in all studies represents a standalone performance relative to the ground truth (LC/MS). The device itself (the strip test) produces a visual result. While a human reads the result, the performance studies assess the accuracy of the device's output (presence/absence of a line) against known concentrations.
- For the "Comparison Studies," the performance of the "Safecare THC Urine Strip Test" is measured for each viewer by comparing their reading of the device against LC/MS. This measures the combined device-human performance.
- For the "Lay-User Study," it also measures the combined device-lay user performance.
7. The Type of Ground Truth Used
- For all performance studies (Precision, Comparison, Lay-User), the ground truth was primarily established by LC/MS (Liquid Chromatography-Mass Spectrometry). This is a highly accurate and quantitative laboratory method.
- For the precision study, it specifies "THC drug concentration was confirmed by LC/MS."
- For the comparison studies, it states "The samples were blind labeled and compared to LC/MS results. LC/MS is the preferred confirmatory method."
- For the lay-user study, it states "The concentrations of the samples were confirmed by LC/MS."
8. The Sample Size for the Training Set
- Based on the provided document, this is a 510(k) submission for a diagnostic device. Such submissions typically focus on analytical and clinical performance validation rather than explicitly detailing a "training set" for an algorithm.
- There is no information provided regarding a "training set" as this is not an AI/machine learning device. The immunoassay device relies on chemical reactions, not on data training.
9. How the Ground Truth for the Training Set Was Established
- As there is no mention of a "training set" or an algorithm that requires training, this question is not applicable based on the provided document.
Ask a specific question about this device
Page 1 of 1