Search Results
Found 1 results
510(k) Data Aggregation
(296 days)
S.I.N. Implant System
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7 mm are intended for delayed loading only.
The subject device consists of four Morse taper connection product lines: Unitite, Unitite Compact, and Strong SW CM. All implants are threaded cylindrical root-form designs. The Unitite provided with an acid-etched and HA surface, the Strong SW CM is provided with an acid-etched surface, Each product line has abutments available in multiple designs (temporary, cementable, angled, UCLA, ball, multi-unit). Subject device abutments are straight only.
Unitite implants are available in six implant body diameters (2.9. 3.5. 4.0. 4.3. 5.0 and 6.0 mm) and eight lengths (5, 6, 7, 8.5, 10, 11.5, 13, and 15 mm). Strong SW CM implants are available in four implant diameters (3.5, 3.8, 4.5.5.0 mm) and five lengths (8.5. 10. 11.5. 13. and 15 mm). Unitite abutments are available in five prosthetic platform diameters (3.3, 3.5, 4.0, 4.5, and 4.8 mm). Strong SW CM abutments are available in four prosthetic platform diameters (3.3, 3.5, 4.5 and 4.8 mm). Subject Device components are made of commercially pure titanium, titanium alloy, cobalt-chromium, or polycarbonate.
The provided text is a 510(k) Summary for the S.I.N. Dental Implant System (K170392). It details the device, its intended use, and a comparison to predicate devices, but does not contain any information about clinical studies with acceptance criteria, sample sizes, expert ground truth establishment, or multi-reader multi-case studies related to device performance.
The document states: "No clinical data were included in this submission."
Therefore, based on the provided text, I cannot complete the requested information. However, I can extract the information provided about performance data (non-clinical) and the device's characteristics.
Here's what I can provide based on the given text:
1. A table of acceptance criteria and the reported device performance
The document does not specify "acceptance criteria" in the traditional sense of performance metrics for a clinical study with a device output. Instead, it describes non-clinical testing performed to demonstrate substantial equivalence to predicate devices.
Non-Clinical Test Category | Reported Performance (Demonstrated Equivalence) |
---|---|
Sterilization Validation | Radiation sterilization validated according to ISO 11137-1 and 11137-2. Steam sterilization according to ISO 17665-1 and ISO 17665-2. Both demonstrating a sterility assurance level (SAL) of 10-6. |
Endotoxin Testing | Limulus amebocyte lysate (LAL) testing according to AANSI/AAMI ST 72. |
Shelf Life Testing | According to ASTM F1980, ASTM F1929, and ASTM F88/F88M. |
Biocompatibility Testing | According to ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-6 (implantation), ISO 10993-10 (sensitization and irritation), and ISO 10993-11 (systemic toxicity), demonstrating acceptable biocompatibility. |
Surface Area Analysis | Showing substantial equivalence to a predicate device. |
Performance Testing | For insertion, showing substantial equivalence to the predicate device. |
2. Sample sized used for the test set and the data provenance
Not applicable. No clinical test set. Non-clinical tests were conducted; however, specific sample sizes for these tests are not provided in the summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. No clinical test set or ground truth established by experts is mentioned.
4. Adjudication method for the test set
Not applicable. No clinical test set or adjudication process is mentioned.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done. The device is a dental implant system, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. The device is a physical dental implant system, not an algorithm or AI.
7. The type of ground truth used
Not applicable. No clinical ground truth is mentioned.
8. The sample size for the training set
Not applicable. No training set is mentioned as this device is not an AI/ML algorithm.
9. How the ground truth for the training set was established
Not applicable. No training set or ground truth for it is mentioned.
Summary of what the document implies about meeting acceptance criteria:
The document demonstrates that the "S.I.N. Dental Implant System" meets the requirements for substantial equivalence to its predicate devices through a series of non-clinical tests. These tests cover aspects like sterilization, shelf-life, biocompatibility, and physical performance (e.g., insertion). The "acceptance criteria" here implicitly refer to the successful completion of these standard tests demonstrating that the device is as safe and effective as the legally marketed predicate devices, and that any differences do not raise new questions of safety or effectiveness. The absence of clinical data in the submission indicates that the FDA deemed the non-clinical evidence sufficient for a finding of substantial equivalence.
Ask a specific question about this device
Page 1 of 1