    K Number
    K220290
    Device Name
    Revitalair 430+
    Date Cleared
    2023-05-04

    (457 days)

    Product Code
    Regulation Number
    868.5470
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Revitalair 430+

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Portable Hyperbaric Chamber Revitalair® 430+ is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at health institutions and physician offices.

    Device Description

    The Portable Hyperbaric Chamber Revitalair® 430+ is a hyperbaric semi rigid chamber for low pressures (operating at pressures of no greater than 1.5 atmospheres absolute (ATA)). The operational design pressure of a hyperbaric chamber that encloses a human within its pressure boundary falls within the scope of the American Society of Mechanical Engineers Pressure Vessels Human Occupancy 1 (ASME PVHO 1-2012).

    Revitalair® 430+ consists of 2 parts, the cabin or chamber and the compressor's cabinet. The chamber, weighing 62 pounds, is constructed of an airtight polyester-based plastic fabric joined by aluminum rings, forming a cylinder 900 mm in diameter and 1850 mm in length (Figure 1).

    The chamber is inflated with atmospheric air through an electric compressor. Fittings allow the chamber to be connected to compressed air by means of manually controlled valves. The safety or relief valves are operated at pressures of 1.5 ATM which ensures safe operations. The compressor has an additional safety valve for any obstruction of the supply hose from the compressor to the Revitalair® 430+ has 10 transparent windows to let in light, 360º viewing and enable easy verification of the patient's comfort from the outside. The Revitalair® 430+ can be operated from the exterior. After folding it up, the Revitalair® 430+ is placed in its transportation box or in the optional carrying case.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Revitalair 430+ Hyperbaric Chamber. It primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a study of the device's performance against specific acceptance criteria for an AI/ML-driven diagnostic system.

    The document does not contain information about:

    • Acceptance criteria and reported performance of an AI/ML device.
    • Sample sizes, data provenance, or ground truth establishment for a test set in the context of AI/ML validation.
    • Number and qualifications of experts or adjudication methods for establishing ground truth for AI/ML.
    • MRMC comparative effectiveness studies or standalone AI algorithm performance.
    • Training set details for an AI/ML model.

    This document is a regulatory submission for a physical medical device (a hyperbaric chamber) and relies on demonstrating substantial equivalence to existing predicate devices, rather than validating a novel AI/ML algorithm's analytical and clinical performance. Therefore, the requested information pertaining to acceptance criteria and performance study for an AI/ML device is not applicable and not present in the provided text.

    The closest relevant information is the non-clinical performance data which focuses on safety and engineering standards for the physical device, not an AI/ML system. These include:

    • Electrical safety testing (IEC 60601-1)
    • Electromagnetic Disturbance (EMD) testing (IEC 60601-1-2)
    • Biocompatibility testing (ISO 10993 and ISO 18562)
    • Device Life Report (reliability analysis)
    • Pressure Testing (ASME PVHO-1-2007)

    The document states: "There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device." This further confirms that it's not a study validating an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171899
    Device Name
    Revitalair 430F
    Manufacturer
    Date Cleared
    2019-11-22

    (879 days)

    Product Code
    Regulation Number
    868.5470
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Revitalair 430F

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions.

    Device Description

    The Portable Hyperbaric Chamber Revitalair® 430F is a hyperbaric chamber for low pressures (operating at pressures of no greater than 1.4 atmospheres absolute (ATA)). The operational design pressure of a hyperbaric chamber that encloses a human within its pressure boundary falls within the scope of the American Society of Mechanical Engineers Pressure Vessels Human Occupancy 1 (ASME PVHO 1-2012).

    Revitalair® 430F consists of 2 parts, the cabin or chamber and the compression system or compressor's cabinet. The chamber, weighing 62 pounds, is constructed of an airtight polyester-based plastic fabric joined by aluminum rings, forming a cylinder 900 mm in diameter and 1850 mm in length (Figure 1).

    The chamber is inflated with atmospheric air through an electric compressor. Fittings allow the chamber to be connected to compressed air by means of manually controlled valves.

    The safety or relief valves are operated at pressures above 1.3 ATM which ensures safe operations. The compressor has an additional safety valve for any obstruction of the supply hose from the compressor to the chamber.

    The Revitalair® 430F has 10 transparent windows to let in light, 360º viewing and enable easy verification of the patient's comfort from the outside.

    The Revitalair® 430F can be operated from the interior as well as from the exterior.

    After folding it up, the Revitalair® 430F is placed in its transportation box or in the optional carrying case.

    AI/ML Overview

    This document primarily concerns the 510(k) premarket notification for the Revitalair® 430F Portable Hyperbaric Chamber, seeking to establish substantial equivalence to an existing predicate device. It does not describe a study involving an AI/Machine Learning device, nor does it list acceptance criteria and performance data for such a device.

    The information provided in the document is related to a traditional medical device (a hyperbaric chamber) and its non-clinical performance testing to demonstrate safety and effectiveness.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and the reported device performance: This document does not contain this information for an AI/ML device. It lists non-clinical performance tests for the hyperbaric chamber itself.
    2. Sample sized used for the test set and the data provenance: Not applicable to an AI/ML study from this document.
    3. Number of experts used to establish the ground truth... and qualifications: Not applicable to an AI/ML study from this document.
    4. Adjudication method: Not applicable to an AI/ML study from this document.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: The document explicitly states, "There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device." This indicates no MRMC study was performed.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI/ML device.
    7. The type of ground truth used: Not applicable. The "ground truth" for this device would be its physical performance characteristics (e.g., pressure capabilities, material strength), verified through engineering tests, not expert consensus on medical images or patient outcomes.
    8. The sample size for the training set: Not applicable, as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.

    In summary, the provided text is a 510(k) submission for a non-AI/ML medical device (a hyperbaric chamber) and does not contain the information requested about AI device performance studies.