Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182272
    Manufacturer
    Date Cleared
    2019-04-15

    (236 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Reprocessed HARMONIC ACE+ 7 Shears with Advanced Hemostasis

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilmed Reprocessed HARMONIC ACE®+ 7 Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter using the Advanced Hemostasis hand control button.

    Device Description

    The Sterilmed Reprocessed HARMONIC ACE+7 Shears with Advanced Hemostasis are used for coagulation and mechanical transection of soft tissue during laparoscopic and open procedures. The devices allow the surgeon to grasp, coagulate, and transect soft tissue with a single instrument. The devices are hand-actuated with a shaft and tissue effector that can be rotated. The energy delivery can be activated with hand activation or with an optional generator foot switch. The two models are identical to each other in materials, form, fit and function except for the length of the shafts which are 23cm and 36 cm.

    AI/ML Overview

    This document describes the validation of reprocessed HARMONIC ACE®+ 7 Shears with Advanced Hemostasis, demonstrating their substantial equivalence to the original equipment manufacturer (OEM) predicate device. Given that the provided text is a 510(k) summary for a reprocessed medical device (surgical shears), the acceptance criteria and study design are focused on demonstrating that the reprocessed device performs as safely and effectively as the original device. This is a non-clinical submission, thus the information regarding AI/human reader studies, expert consensus on images, etc., is not applicable here.

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for a reprocessed device are fundamentally centered on demonstrating that it performs equivalently to the new, original device in terms of safety and functional effectiveness. The performance of the reprocessed device is compared against the OEM predicate device. Specific quantitative acceptance criteria or threshold values are not explicitly listed in this summary, but the general principle is that the reprocessed device "performs as originally intended" and is "substantially equivalent" to the predicate.

    Test Parameter / Acceptance CriteriaReported Device Performance (Reprocessed Device)
    Functional Performance
    Simulated use, visual inspection, fatigue testing. Must perform as originally intended."Performance testing shows the Reprocessed HARMONIC ACE+ 7 Shears with Advanced Hemostasis performs as originally intended."
    Cleaning Validation
    Validation of cleaning efficacy."Process validation testing was performed to validate cleaning..."
    Sterilization Verification
    Validation of sterilization efficacy."...and sterilization..."
    Ethylene Oxide Residual Testing
    (ISO 10993-7 compliant for residuals)"...ethylene oxide residual testing (ISO 10993-7)..."
    Packaging Validation
    (ASTM D4169, ASTM F88, ASTM F2096 compliant)"...packaging validation (ASTM D4169, ASTM F88, ASTM F2096)..."
    Shelf-Life Validation
    (ASTM 1980-07 compliant)"...and shelf-life validation (ASTM 1980-07)."
    Electrical Safety
    Must meet safety standards."Testing performed: Electrical Safety"
    Vein Seal Burst (1-7mm) & Tissue Adhesion
    Equivalent sealing performance to predicate."Vein Seal Burst (1-7mm) and Tissue Adhesion" (Implied equivalent to predicate)
    Artery Seal Burst (1-7mm) & Tissue Adhesion
    Equivalent sealing performance to predicate."Artery Seal Burst (1-7mm) and Tissue Adhesion" (Implied equivalent to predicate)
    Lymphatics Seal Performance
    Equivalent sealing performance to predicate."Lymphatics Seal Performance" (Implied equivalent to predicate)
    Vessel Seal Thermal Spread
    Minimize thermal injury, equivalent to predicate."Vessel Seal Thermal Spread" (Implied equivalent to predicate)
    Drop Fluid Ingress
    Maintain integrity against fluid ingress."Drop Fluid Ingress" (Implied satisfactory)
    Pad Retention, Pad Life, Grasping Force
    Maintain mechanical integrity and function."Pad Retention, Pad Life, Grasping Force" (Implied satisfactory)
    Biocompatibility
    (ISO 10993-1 compliant: Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity)"Biocompatibility testing included: Cytotoxicity Sensitization Irritation/Intracutaneous Reactivity Acute Systemic Toxicity Pyrogenicity" (Implied compliant)
    Overall Comparison to Predicate Device
    Substantially equivalent with respect to safety and effectiveness."Results demonstrated substantial equivalence to the predicate devices with respect to safety and effectiveness."

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document states that "representative samples of reprocessed devices were tested," and for functional testing, "all products produced" undergo visual and validated functional testing. However, specific numerical sample sizes for each test (e.g., how many devices were tested for vein seal burst, or fatigue testing) are not provided in this summary.
    • Data Provenance: The testing was conducted by Sterilmed, Inc., a reprocessor based in Plymouth, MN, and Maple Grove, MN, USA. The data is from prospective bench testing and validation studies of their reprocessed devices. It is not human clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not applicable as this is a submission for reprocessed surgical shears based on non-clinical, bench testing, and engineering validation, not an AI or imaging-based device. No "experts" in the sense of medical professionals interpreting images were used to establish ground truth for the device's functional performance. Ground truth was established through validated engineering and laboratory test methods, often comparing performance against the OEM predicate.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving multiple human readers interpreting data, typically in diagnostic imaging. For a reprocessed surgical device, the "adjudication" is inherent in the robust, validated test methods and comparison to established predicate performance. If a test failed, it would need to be investigated and resolved, but there isn't a "reader-based" adjudication process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. MRMC studies are specific to diagnostic imaging devices, often involving AI, where human readers evaluate cases. This submission is for reprocessed surgical shears, which do not involve human readers for diagnostic interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This device is a mechanical/ultrasonic surgical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is established by:

    • Engineering specifications and performance characteristics of the OEM predicate device (Ethicon HARMONIC ACE®+7 Shears with Advanced Hemostasis, K132612).
    • Validated industry standards and test methods (e.g., ISO 10993, ASTM standards for packaging, electrical safety).
    • Successful functional performance benchmarks (e.g., vein/artery seal burst pressure, thermal spread, grasping force), which show the device "performs as originally intended."
      In essence, the ground truth is that the reprocessed device must meet or exceed the safety and functional performance of the new, original device.

    8. The sample size for the training set:

    This is not applicable. This is not an AI/machine learning device; there is no "training set" in the computational sense. The "training" for the manufacturing process involves process validation and quality control measures to ensure consistent reprocessing.

    9. How the ground truth for the training set was established:

    This is not applicable for the same reason as point 8. The "ground truth" for the overall manufacturing and reprocessing process is established through rigorous validation of cleaning, sterilization, packaging, and functional testing protocols to ensure the device is consistently reprocessed to meet its intended performance specifications.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1