Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221918
    Date Cleared
    2022-09-29

    (90 days)

    Product Code
    Regulation Number
    882.5550
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ReFlow System Mini, used as components of a shunt system, is for use in the treatment of patients with hydrocephalus or conditions where draining of cerebrospinal fluid (CSF) is medically indicated. The miniaturized ReFlow Flusher may be used by a qualified clinician as a tool to facilitate a noninvasive retrograde fluid flush of the ReFlow Ventricular Catheter to unblock inlet holes or open its relief membrane to restore, increase, or maintain CSF flow. The ReFlow System Mini is not intended to change the diagnosis, treatment, or follow-up of patients with proximal catheter occlusions. Under the care, direction of the treating physician, the ReFlow System Mini may be used as directed for noninvasive flushing by a trained healthcare professional in-clinic or by a trained caregiver or adult patient in a non-clinical environment.

    Device Description

    The ReFlow System Mini (RSM), consists of the ReFlow Mini Flusher and ReFlow Ventricular Catheter, both implantable components of a cerebrospinal fluid (CSF) shunt system used in the treatment of patients with hydrocephalus and other CSF disorders when draining of CSF is medically indicated. The catheter is surgically implanted in the ventricle of the brain and connected distally to the Mini Flusher. The Mini Flusher is then connected to a commercially available flow/pressure regulating valve (not provided as part of the ReFlow System Mini).

    The flusher contains a fluid reservoir and when implanted as part of a shunt system, may be used at the treating physician's discretion as a tool to non-invasively facilitate a one-way retrograde pulse of fluid through the ventricular catheter to potentially restore or maintain flow in the shunt system by unblocking the proximal catheter. The flusher is noninvasively actuated by depressing the flusher dome, which is palpable underneath the patient's scalp, to send a controlled and limited pulse of a consistent volume of fluid (patient's CSF or sterile saline introduced from the priming procedure before implant) within the closed system. In cases where inlet holes are not able to be unblocked by a flush actuation, the flush will open the ReFlow Ventricular Catheter emergency relief membrane to provide a secondary fluid pathway to restore or increase CSF flow in the shunt system.

    After flushing the device, palpation to confirm the dome has returned to its original shape may be used to determine that the reservoir has refilled prior to flushing a second time. Refilling of the flusher reservoir can be confirmed by palpation of the dome, indicating that CSF from the ventricles is able to flow through the ventricular catheter to the ReFlow Mini Flusher. The clinician must use his/her medical judgment and standard practice at his/her institution to care for the patient pre, during, and post utilization of the ReFlow System Mini.

    The ReFlow System Mini does not regulate the flow/pressure of the shunt system. A flow regulating shunt valve is not provided with the ReFlow System Mini. During passive flow, fluid from the ventricular catheter flows freely through the Mini Flusher to the regulating valve. The ReFlow System Mini is designed to be compatible with most commercially available flow/pressure requlative valves with standard inlet connectors. Additionally, the Reflow Mini Flusher component utilizes a common barbed connector designed to be compatible with most commercially available ventricular catheters in addition to the ReFlow Ventricular Catheter.

    AI/ML Overview

    This document describes the ReFlow System Mini, a device used in the treatment of hydrocephalus. The device consists of a miniaturized flusher and a ventricular catheter, designed to facilitate non-invasive retrograde fluid flushing to restore or maintain CSF flow. The submission (K221918) seeks to demonstrate substantial equivalence to a predicate device (ReFlow Ventricular Catheter and Flusher System, K172006).

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document summarizes performance data rather than providing a detailed table of acceptance criteria and specific numerical performance metrics. However, based on the provided text, the acceptance criteria are generally "met specifications for intended use" and "substantially equivalent to the predicate device."

    Test CategoryTest Method SummaryReported Device Performance and Conclusion
    Particulate TestingUnits undergo particulate testing to evaluate that the number of particulates generated from both exterior and interior lumen of the ReFlow Mini Flusher, under simulated use conditions, does not pose any safety risks.All results met acceptance criteria and demonstrate that the RSM performance is suitable for its intended use and is substantially equivalent to the predicate device.
    Design Verification Bench TestingConducted testing and reported results in accordance with ISO 7197:2006 (E), Neurosurgical Implants - Sterile, single-use hydrocephalus shunts and components and ASTM F647: 94(2014), Standard Practice for Evaluating and Specifying Implantable Shunt Assemblies for Neurosurgical Application.All results met acceptance criteria and demonstrate the RSM is suitable for its intended use and is substantially equivalent to the predicate device.
    Reliability and System Level Functional TestTesting was conducted to verify the reliability of the ReFlow System Mini to function during its expected useful life and meet requirements of ISO 7197:2006 (E), Neurosurgical Implants - Sterile, single-use hydrocephalus shunts and components and ASTM F647: 94(2014), Standard Practice for Evaluating and Specifying Implantable Shunt Assemblies for Neurosurgical Application.All results met acceptance criteria and demonstrate the RSM is suitable for its intended use and is substantially equivalent to the predicate device.
    Usability and Human Factors TestingHuman Factors and Usability evaluations for critical tasks according to ReFlow Mini labeling were conducted in accordance with ANSI/AAMI HE 75, Human Factors Engineering Design of Medical Devices, IEC 62366:2007, Medical Devices - Application of Usability Engineering to Medical Devices, and FDA Guidance (Jun 22, 2011) Human Factors Draft Guidance.RSM is deemed to meet its specifications for its intended use and is substantially equivalent to the predicate device. In all cases, the acceptance criteria were met, and the device performed as expected according to its specifications and in compliance with applicable recognized standards. In all instances, the RSM functioned as intended and were as expected. These test results confirm that RSM complies with the recognized standards, the design specifications and performance requirements for the intended use, and is substantially equivalent to the predicate.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    The document mentions that usability and human factors testing demonstrated that flushing can be performed by "a trained healthcare provider (other than the treating physician), a caregiver, or adult patient" without special skills or education. However, it does not specify the sample size for this human factors test set, nor the country of origin of the data, or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The document states that human factors and usability performance testing demonstrated that the device could be used by "a trained healthcare professional in-clinic or by a trained caregiver or adult patient in a non-clinical environment" under the direction of the treating physician. However, it does not specify the number of experts used to establish a 'ground truth' for the test set or their qualifications, as the testing seems focused on user performance rather than expert assessment of outcomes in a clinical trial.

    4. Adjudication Method for the Test Set:

    The document does not mention any adjudication method (e.g., 2+1, 3+1, none) for the test set. The performance data focuses on technical and functional compliance rather than diagnostic or clinical effectiveness that would typically require such adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The evaluation focuses on the safety and performance of the device itself, and its equivalence to a predicate device, rather than comparing human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    The device described is a physical medical device (shunt system components) and not an algorithm or AI-powered system. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" is not applicable here. The device's function involves human interaction for actuation (flushing).

    7. The Type of Ground Truth Used:

    The ground truth used for these tests is based on engineering and performance standards (ISO 7197:2006, ASTM F647: 94(2014), ANSI/AAMI HE 75, IEC 62366:2007, and FDA Guidance for Human Factors) and whether the device operates as designed to restore or maintain CSF flow, without posing safety risks. The "ground truth" here is the successful demonstration that the device's functional and safety characteristics meet these established engineering and usability requirements.

    8. The Sample Size for the Training Set:

    This information is not provided in the document, as the device is a physical medical device, not a machine learning model that would typically have a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as the device is not a machine learning model, and therefore, there is no "training set" or ground truth for such a set in the context of this submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1