510(k) Data Aggregation

    K Number
    K993063
    Manufacturer
    Date Cleared
    1999-12-09

    (87 days)

    Product Code
    Regulation Number
    876.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rüsch Brillant Two Way Silicone Foley is used to drain fluids to and from the urinary tract.

    The Rüsch Brillant Three Way Silicone Foley is used to drain fluids to and from the urinary tract, and to flush and irrigate the urinary tract.

    Device Description

    The Rusch Brillant Silicone catherer is of the balloon retention type, commonly called a Foley catheter. The device is single use, disposable, and sterile. The central lumen is for draining fluids to and from the urinary tract. The second lumcn is to inflate and deflate the balloon with sterile water. On models with a third lumen (commonly called a Threeway Catheter), it is used in conjunction with the central lumen for flushing or irrigating the urinary tract.

    The distal end has eye holes, which are for drainage. On the opposing end of the shaft, are a connecting funnel (for a drainage bag, purchased separately) and a Luer activated valve.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving that the device meets those criteria. This document is a 510(k) summary for a Foley catheter, primarily focusing on establishing substantial equivalence to previously marketed devices. It describes the device, its intended use, and its technological characteristics, but it does not detail any performance studies, acceptance criteria, or specific metrics.

    Therefore, I cannot extract the requested information from the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972184
    Manufacturer
    Date Cleared
    1998-01-08

    (212 days)

    Product Code
    Regulation Number
    876.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rüsch Brillant Silicone Foley is used to drain fluids to and from the urinary tract.

    Device Description

    The Rusch Brillant Silicone catheter is of the retention type, commonly called a Foley catheter. The device is a single use, disposable, sterile with retention balloon, which is attached to the silicone shaft. One lumen is for draining fluids to and from the urinary tract. The second lumen is to inflate and deflate the balloon with sterile water. On models with a third lumen, it is used in conjunction with the first lumen for flushing the urinary tract. Sterile water is used for inflation and deflation of the balloon. The distal end has two opposite eye holes, which are used for drainage. On the opposing end of the shaft, are a connecting funnel and a Luer activated valve. Pediatric sizes of this product are used with a stylet that is inserted prior to the catheter placement. This product is available in sizes 8 Fr to 26 Fr.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and supporting studies, formatted as requested.

    Based on the provided 510(k) summary for the Rüsch Brillant Silicone Foley catheter, there is no information about a study proving the device meets specific performance-based acceptance criteria in the context of AI/software device evaluation.

    The document is a traditional 510(k) for a physical medical device (a catheter), which focuses on substantial equivalence to a predicate device rather than performance metrics of an AI algorithm. Therefore, many of the requested fields are not applicable in this context.

    Here's an attempt to answer the questions based on the available information, noting where information is absent or irrelevant to AI/software performance studies:

    **1. A table of acceptance criteria and the reported device performance**
    
    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Material & DesignEquivalent to predicate devices (Porgés Foley Catheter) in:The silicone catheter with silicone balloon, ported distal tip, and silicone funnel is equivalent to predicate devices.
    - Catheter materialSilicone
    - Balloon materialSilicone
    - Tip designPorted distal tip
    - Funnel designSilicone funnel
    Intended UseUsed to drain fluids to and from the urinary tract.The device is described for this intended use.
    SterilitySterileDevice is described as sterile.
    Single Use/DisposableSingle use, disposableDevice is described as single use, disposable.
    Mechanical Features- Retention balloon mechanismPresent (inflates/deflates with sterile water).
    - LumensOne for drainage, second for balloon inflation/deflation. (Some models have a third for flushing).
    - Eye holesTwo opposite eye holes at distal end for drainage.
    - ConnectorsConnecting funnel and Luer activated valve at opposing end.
    SizingAvailable in sizes 8 Fr to 26 Fr.Available in sizes 8 Fr to 26 Fr.
    Pediatric UseStylet used with pediatric sizesPediatric sizes use a stylet.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This submission is for a physical medical device (catheter) and does not involve a "test set" of data for evaluating an AI or software algorithm. The evaluation is based on substantial equivalence to a predicate device, material properties, and manufacturing controls.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. As above, there is no "test set" or "ground truth" establishment in the context of an AI/software performance study for this device. The regulatory review focuses on engineering, materials, and clinical equivalence rather than expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. Any "ground truth" related to this physical device would refer to established engineering standards, material specifications, and clinical performance of the predicate device, not ground truth for an AI assessment.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/machine learning device.