Search Results
Found 1 results
510(k) Data Aggregation
(249 days)
Royal-Dent Eureka Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including: cemented retained, screw retained or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. Royal-Dent Eureka Dental Implant System is for one and two stage surgical procedures. It's not for immediate load.
Royal-Dent Eureka Dental Implant System is a multiple component device, made from CP Ti and Ti-6Al-4V, which allow the surgeon to build an implant system to fit the patient's requirements. The system is a set of implants designed to be inserted into the upper and/or lower jawbone and serves as a replacement for patient's tooth providing a stable foundation for restorations. The system is provided with simplified main implants, but with a variety of different styles of components. Eureka and Eureka Plus implant series are the sub-types incorporated into Royal-Dent Dental Implant System. The major difference between two implant sub-types is outer design configuration. The design configuration and features of inner hole for the sake of fitness to various screws and abutments are identical. The Implants are provided sterile, the remaining components must be sterilized prior to use.
The provided text describes the 510(k) summary for the "Royal-Dent Eureka Dental Implant System." This document focuses on demonstrating substantial equivalence to predicate devices for regulatory approval, rather than a clinical study evaluating AI performance. Therefore, many of the requested criteria related to AI device performance evaluation, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable or cannot be extracted from this specific document.
The document primarily details the device's characteristics, intended use, and pre-clinical bench and biocompatibility testing to show similarity to existing approved devices.
Here's an analysis of the provided information based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in the format of quantitative thresholds for performance metrics. Instead, "acceptance" is demonstrated through substantial equivalence to predicate devices and adherence to relevant ISO standards for bench and biocompatibility testing. The "reported device performance" is essentially that the device performed acceptably in these pre-clinical tests, supporting its safety and effectiveness.
Criteria Category | Details Provided in Submission |
---|---|
Substantial Equivalence to Predicate Devices | Indication for Use: "Royal-Dent Eureka Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including: cemented retained, screw retained or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. Royal-Dent Eureka Dental Implant System is for one and two stage surgical procedures. It's not for immediate load." (Identical or very similar to predicate devices, noting the subject device allows for one and two-stage procedures while predicates were for one-stage). |
Materials: "CP Ti : dental implant"; "Ti6Al4V: abutments, screws, healing caps, clix." (Similar to predicate devices which also use CP Ti and Ti6Al4V, though predicates also list Gold alloy + plastic for some components).
Use: "Single use" (Same as predicate devices).
Dimensions: Diameter: 3.5, 4.0, 4.3, 4.5, 5.0mm; Length: 8.5, 10,11.5,13,15mm (Similar to predicate devices, which have a slightly wider length range starting from 7.0mm).
Method of Sterilization: "Gamma sterilization: implant; Abutments and accessories : non-sterile" (Same as predicate devices).
Conclusion: "Based on the similarities between the Royal-Dent Eureka Dental Implant System and the predicate devices, the safety and effectiveness of the Royal-Dent Eureka Dental Implant System is substantially equivalent to the predicate devices." |
| Pre-clinical Testing Results | Bench testing: Per ISO 14801 (Implied successful passing of this static and dynamic fatigue strength test for endosseous dental implants).
Biocompatibility Testing:
- Genotoxicity Testing per ISO 10993-3 (Implied acceptable results).
- Cytotoxicity testing per ISO 10993-5 (Implied acceptable results).
- Implantation testing per ISO 10993-6 (Implied acceptable results).
- Sensitization testing per ISO10993-10 (Implied acceptable results).
- System Injection testing per ISO10993-11 (Implied acceptable results).
Conclusion: "Performance Testing were also conducted to demonstrate safety and effectiveness of the proposed devices." (This statement implies the tests met the required standards for safety and effectiveness). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable as this document describes the 510(k) clearance process for a dental implant system, not a study involving a "test set" for an AI device. The testing mentioned (bench and biocompatibility) refers to a mechanical and biological evaluation of the implant material and design, not an AI algorithm's performance on a dataset.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable. There is no "ground truth" for a "test set" in the context of an AI device being evaluated. The evaluations were based on established ISO standards for material and mechanical properties.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable for the reasons stated above.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. This document is for a dental implant, not an AI device, so no MRMC study or AI assistance evaluation was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as the device is a physical dental implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not applicable. The "ground truth" in this context refers to established scientific and engineering principles, and validated test methodologies defined by ISO standards for evaluating the physical and biological properties of medical devices.
8. The sample size for the training set
This information is not applicable as this is not an AI device.
9. How the ground truth for the training set was established
This information is not applicable as this is not an AI device.
Ask a specific question about this device
Page 1 of 1