Search Results
Found 1 results
510(k) Data Aggregation
(88 days)
The RMI Hemoconcentrator Tubing Set is a set of tubing, fittings and connectors that is used to interconnect an RMI hemoconcentrator to: 1) a location in an extracorporeal circuit that will sustain blood flows up to 500 ml/min, 2) either a cardiotomy or venous reservoir, and 3) a graduated ultrafiltrate collection container.
The RMI Hemoconcentrator Tubing Set is a set of tubing, fittings and connectors that is used to interconnect an RMI hemoconcentrator to: 1) a location in an extracorporeal circuit that will sustain blood flows up to 500 ml/min, 2) either a cardiotomy or venous reservoir, and 3) a graduated ultrafiltrate collection container. The RMI Hemoconcentrator Tubing Set is provided sterile and non-pyrogenic.
Here's an analysis of the provided text regarding the RMI Hemoconcentrator Tubing Set, structured to address your specific questions.
Note: The provided text primarily focuses on biocompatibility and basic functional performance testing. It does not contain information related to AI/ML device performance, ground truth establishment, expert consensus, or multi-reader multi-case studies. Therefore, many of your requested points will be marked as "Not Applicable" or "Not Provided in Text."
Acceptance Criteria and Device Performance for RMI Hemoconcentrator Tubing Set
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Biocompatibility | Not cytotoxic | The test article is not cytotoxic (MEM Elution Test). |
| Not a significant sensitizer | Upon rechallenge, none of the test group or negative control animals exhibited scores higher than 1 (response of 1 is not considered significant sensitization) (Guinea Pig Maximization). | |
| Meets USP Intracutaneous Test requirements | The test article met the requirements of the USP Intracutaneous Test. | |
| Meets USP Systemic Injection Test requirements | The test article met the requirements of the USP Systemic Injection Test. | |
| Hemolysis ≤ Acceptable Limit (Implicitly 0% or very low) | The test article had 0.00% Hemolysis. | |
| Functional / Performance Testing | ||
| Flow vs. Pressure (Blood Flow) | Deliver flows exceeding 1000 ml/min with line pressure < 70 mmHg (for blood flows up to 500 ml/min in extracorporeal circuit) | Five (5) product samples tested delivered flows exceeding 1000 ml/min with line pressure under 70 mmHg. (This seems to indicate the tubing can pass 1000 ml/min at <70mmHg line pressure while the device itself is designed for up to 500ml/min, implying sufficient capacity). |
| Overpressure and Leak | Withstand pressure 10 times greater than standard operating pressure | Five (5) unit samples tested demonstrated the product will withstand a pressure ten (10) times greater than the standard operating pressure observed during flow testing. |
| Mechanical Strength | Withstand a 6 times minimum acceptable pull force | Five (5) unit samples tested demonstrated a six (6) times minimum acceptable pull force per Q.O.P. #80138. |
-
Sample sizes used for the test set and the data provenance:
- Biocompatibility Testing: The sample sizes are not explicitly stated as a single number but are implied by tests involving "whole product extract," "tubesheet slices," "test group," and "negative control animals." Specific numbers for animals or extraction replicates are not provided in this summary.
- Functional / Performance Testing:
- Flow vs. Pressure Test: 5 product samples
- Overpressure and Leak Testing: 5 unit samples
- Mechanical Strength Test: 5 unit samples
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). This is a general safety and effectiveness summary, so it's likely part of internal R&D/testing.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This device is a medical tubing set, and its performance is evaluated through physical and chemical tests, not through expert interpretation of medical images or data requiring "ground truth" in the AI/ML sense. The "ground truth" in this context refers to the measured physical and chemical properties matching predetermined specifications.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or for establishing ground truth from expert opinions, which is not relevant for the types of tests described here (biocompatibility, flow, pressure, mechanical strength). The results of these physical tests are quantitative measurements.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a medical device (tubing set), not an AI/ML diagnostic or assistive tool. MRMC studies are not applicable.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For biocompatibility: Ground truth is established by the specified biological responses (e.g., lack of cytotoxicity, minimal sensitization, meeting USP standards for irritation and systemic toxicity, 0.00% hemolysis) as defined by the referenced standards (FDA MATRIX, ISO 10993, USP).
- For functional/performance testing: Ground truth is established by quantitative measurements against predefined engineering specifications (e.g., flow rate, pressure resistance, pull force).
-
The sample size for the training set:
- Not Applicable. This is not an AI/ML device, so there is no "training set."
-
How the ground truth for the training set was established:
- Not Applicable. No training set exists for this device.
Ask a specific question about this device
Page 1 of 1