    Intended Use

    The ReavillMED product is combination product containing a pressure monitoring system and a peripherally inserted central catheter. The pressure monitoring system is intended for continuous measurement of blood pressures or other physiological pressures. The catheter is intended for short term peripheral access to the central venous system.

    Device Description

    The ReavillMED pressure monitoring device is a combination product consisting of the following components within a single device:

    • . Blood pressure monitoring line and end cap
    • . Disposable pressure transducer
    • Single lumen polyurethane reverse taper catheter .
    • . Catheter hub with suture wing tabs and slip tip cap
      The single lumen catheter is packaged within a pressure monitoring line to maintain a sterile environment. A transducer is attached to the distal end of the monitoring line.
      The product is sold sterile, single use, and non-pyrogenic.
    AI/ML Overview

    The ReavillMED Pressure Monitoring System is a combination product consisting of a blood pressure monitoring line with an end cap, a disposable pressure transducer, a single lumen polyurethane reverse taper catheter, and a catheter hub with suture wing tabs and a slip tip cap. It is intended for continuous measurement of blood pressure or other physiological pressures, and the catheter is for short-term peripheral access to the central venous system.

    1. Table of acceptance criteria and the reported device performance

    Testing PerformedStandardAcceptance CriteriaReported Device Performance
    Sensor TestingANSI/AAMI BP-22, 1994Operating Pressure Range: -50 mmHg to +300 mmHgMeets Requirements
    Sensitivity: 5µV/V/mmHg, +/- 2% (typically 500 hoursMeets Requirements
    Storage Temperature: -25 C to +65 CMeets Requirements
    Defibrillation Withstand: 5 discharges/5 minutes of 360 joules @ 50 Ω loadMeets Requirements
    Natural Frequency: > 20 Hz in salineMeets Requirements
    Phase Shift: