    K Number
    K080471
    Date Cleared
    2008-03-05

    (13 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    RADIO BIOAQUACARE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Radio BioAquaCare™ is a hydrogel wound dressing intended for the management of wounds and provides a moist environment that supports the autolytic debridement of areas of the wound that are necrotic. Radio BioAquaCare™ wound dressing provides an effective barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing. It is intended for use with dry, light, and moderately exuding partial and full-thickness wounds such as minor burns, superficial injuries, lacerations, cuts, abrasions, incisions and surgical wounds, and skin tears. The dressing also protects against abrasion, desiccation, and external contamination. The moist environment has a cooling effect that may reduce pain.

    Under the direction of a health care professional, Radio BioAquaCare™ is indicated for the local management of: (1) chronic wounds such as pressure ulcers, lower extremity ulcers, venous ulcers, arterial ulcers, diabetic ulcers, radiation dermatitis, and (2) 2nd degree burns and donor sites.

    Radio BioAquaCare™ is intended for single use only. Radio BioAquaCare™ is not intended for treatment of third degree burns.

    Device Description

    Radio BioAquaCare™ wound dressing is composed of the following ingredients: 1) PEG.Soy protein; 2) EDTA; 3) sodium phosphate; 4) sodium chloride; 5) water; and 6) Liquid Germall Plus® as a preservative. The preservative contained in Radio BioAquaCare™ aids in providing an effective barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing. Radio BioAquaCare™ contains up to 85% water which enables the dressing to provide a moist environment even if the wound bed is dry. Radio BioAquaCare™ will not breakdown during use, so all of the Radio BioAquaCare™ components are removed from the wound when the dressings are removed. Radio BioAquaCare™ does not include any medicinal substance, animal/human tissue, or biologics. Radio BioAquaCare™ is supplied with gauze for mechanical support, and with a backing which serves to control the rate of water evaporation from the wound dressing upon application to the wounds.

    Radio BioAquaCare™ is packaged in either a single use blister packages or in plastic bag. Radio BioAquaCare™ is supplied in a variety of sizes (i.e., 5 cm x 5 cm, 8 cm x 10 cm, 8 cm x 20 cm, 8 cm x 27 cm, 8 cm x 30 cm, 8 cm x 40 cm, 8 cm x 50 cm, 10 cm x 10 cm, 10 cm x 20 cm, 10 cm x 27 cm, 10 cm x 30 cm, 10 cm x 40 cm, 10 cm x 50 cm, 15 cm x 10 cm, 15 cm x 20 cm, 15 cm x 30 cm, 15 cm x 40 cm, or 15 cm x 50 cm). Radio BioAquaCare™ is manufactured under aseptic conditions in a controlled class 100 lean room environment.

    AI/ML Overview

    Here is the breakdown of the acceptance criteria and study information for BAGTech's Radio BioAquaCare™, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state specific quantitative acceptance criteria for the Radio BioAquaCare™ device. Instead, it relies on demonstrating technological identity and substantial equivalence to previously cleared predicate devices. The performance is described in qualitative terms, asserting that the device "functioned as intended and the performance observed was as expected" based on the validation of prior studies.

    Acceptance CriterionReported Device Performance
    BiocompatibilityFunctioned as intended; performance as expected based on prior studies of BioAquaCare™.
    Preservative System EffectivenessFunctioned as intended; performance as expected based on prior studies of BioAquaCare™. Ensures "effective barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing."
    Effectiveness in Wound ManagementFunctioned as intended; performance as expected based on prior studies of BioAquaCare™. Provides "moist environment that supports the autolytic debridement of areas of the wound that are necrotic."
    Maintenance of Integrity During Use"Will not breakdown during use, so all of the Radio BioAquaCare™ components are removed from the wound when the dressings are removed." (This is a design characteristic, not a measured performance metric from the study described).

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" with a defined sample size for Radio BioAquaCare™. The performance data presented for Radio BioAquaCare™ is based on extrapolation from studies conducted on its predecessor, BioAquaCare™. The nature of these studies (in vitro and animal studies) is mentioned, but no sample sizes or data provenance (country of origin, retrospective/prospective) are provided for these underlying studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The filing does not describe a test set or the involvement of experts for ground truth establishment. The device's performance is asserted based on existing studies of a prior, technologically identical device.

    4. Adjudication method for the test set

    Not applicable. There is no described test set or adjudication process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a hydrogel wound dressing, not an AI-assisted diagnostic or imaging device, so MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a hydrogel wound dressing, not an algorithm.

    7. The type of ground truth used

    For Radio BioAquaCare™ itself, the "ground truth" is established through technological identity to the predicate device, BioAquaCare™. The assumption is that because Radio BioAquaCare™ is "technologically identical" to BioAquaCare™, the performance observed in BioAquaCare™'s in vitro and animal studies (which would have used various experimental endpoints as their "ground truth") is directly applicable. The document does not detail the specific ground truths of those underlying studies (e.g., pathology, measured outcomes).

    8. The sample size for the training set

    Not applicable. This is a medical device (wound dressing), not an AI/ML model, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no "training set" for this device.