Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K060998
    Device Name
    QMS TOBRAMYCIN
    Manufacturer
    Date Cleared
    2006-07-21

    (101 days)

    Product Code
    Regulation Number
    862.3900
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Tobramycin assay is intended for the quantitative determination of tobramycin in human serum or plasma on automated clinical chemistry analyzers.

    The results obtained are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to help ensure appropriate therapy.

    Device Description

    The QMS® Tobramycin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle. The assay consists of reagents R1: anti- tobramycin monoclonal antibody and R2: tobramycin -coated microparticles. A six-level set of QMS® Tobramycin Calibrators (A through F) is used to calibrate the assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Seradyn QMS® Tobramycin assay, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    TestAcceptance CriteriaReported Device Performance
    Accuracy% Recovery: 100 ± 10%Mean Percent Recovery: 94.14%
    Linearity% Recovery: 100 ± 10%Correlation coefficient (R²): 0.9996
    PrecisionTotal CV: < 10%Low Control: 7.57%
    Mid Control: 4.23%
    High Control: 4.26%
    Interferences (Endogenous Substances)% Recovery: 100 ± 10%Albumin: 97.66%
    Bilirubin: 92.27%
    Cholesterol: 105.84%
    Gamma Globulins (IgG): 97.44%
    Hemoglobin (20 mg/dL): 97.16%
    Hemoglobin (500 mg/dL): 108.65%
    Uric Acid: 90.31%
    Rheumatoid Factor: 103.29%
    Triglyceride: 90.98%
    Interferences (HAMA)% Recovery: 100 ± 10%HAMA Type-1: 92.58%
    HAMA Type-2: 93.82%
    Sensitivity (LOQ)Lowest concentration reliably detected meeting accuracy requirements0.4 µg/mL
    Assay RangeBased on Accuracy, Linearity, Sensitivity0.4 to 10 µg/mL
    Method ComparisonExcellent correlation between predicate and new deviceN = 67, Slope = 0.979, y-intercept = -0.086, R = 0.992, R² = 0.984
    On-Board Stability (Calibration Curve)Sufficient stability for intended claim14 days
    On-Board Stability (Reagent)Sufficient stability for intended claim45 days

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Accuracy: The exact number of spiked samples is not explicitly stated beyond "across the range of the assay," but each sample was analyzed in triplicate. The samples used were human serum negative for the drug. Data provenance is not specified (e.g., country of origin), but it is a retrospective laboratory study using controlled samples.
      • Linearity: Similar to accuracy, a tobramycin in human serum pool was diluted, and samples were analyzed in triplicate. Data provenance is not specified.
      • Sensitivity (LOQ): Not specified.
      • Method Comparison: N = 67 patient samples. Data provenance is not specified, but it's a retrospective comparison using patient samples.
      • Precision: 80 measurements per control level (Low, Mid, High) over a period of time (e.g., 20 days with 4 replicates per day, which is a common NCCLS EP5-A2 setup, though not explicitly stated here). Data provenance not specified.
      • Interferences: 3 replicates per interfering substance. Data provenance not specified.
      • Cross-reactivity: Each cross-reactant drug was tested at a specific concentration (e.g., 30 µg/mL for 5-Fluorocytosine, 200 µg/mL for Amikacin).
      • Anticoagulants: Not explicitly stated but implies testing with different types of plasma samples and serum.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This is a quantitative immunoassay for a drug, and the "ground truth" for the test set samples is typically established by precisely preparing known concentrations of tobramycin (for accuracy, linearity, sensitivity) or by using a a well-established comparative method (for method comparison). There are no human expert readers involved in establishing the ground truth for these types of in vitro diagnostic tests.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. As described above, this is laboratory testing of a quantitative assay, not an image interpretation or diagnostic performance study requiring expert adjudication. Recovered concentrations are compared to theoretical concentrations or to a predicate device's results.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI-assisted diagnostic device, nor does it involve human readers interpreting results in the way an MRMC study evaluates. It's an automated clinical chemistry analyzer.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance studies described (Accuracy, Linearity, Sensitivity, Precision, Method Comparison, Interferences, Stability) are all standalone performance evaluations of the assay system itself. This device is entirely automated; there is no "human-in-the-loop" once the sample is loaded and the assay initiated. The results are quantitative measurements produced directly by the instrument.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Accuracy and Linearity: The ground truth was established by known, theoretical concentrations of tobramycin prepared by spiking or diluting human serum.
      • For Method Comparison: The "ground truth" was essentially the results obtained from the legally marketed predicate device (Abbott TDx/TDxFLx Tobramycin assay). The study aimed to show correlation and agreement with an already established method.
      • For Precision: The ground truth was the mean concentration of the control materials run multiple times.
      • For Interferences/Cross-reactivity: The ground truth involved known concentrations of tobramycin with the addition of known concentrations of interfering substances or cross-reactants.
    7. The sample size for the training set:

      • Not applicable. This is a traditional immunoassay, not a machine learning or AI-based device that requires a "training set" in the computational sense. The reagents and assay parameters are developed and optimized through standard biochemical and analytical chemistry processes, not through training on a dataset.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the context of this device. Development of the assay's reagents and methodologies would involve empirical testing and optimization to achieve desired analytical performance characteristics.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1