Precision1; Precision1 for Astigmatism; Dailies Total1; Dailies Total1 for Astigmatism; Dailies Total1
Precision1 (verofilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes and approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.
Precision1 for Astigmatism (verofilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes and 6.00 diopters (D) or less of astigmatism.
Precision1 lenses are to be prescribed for single use, daily disposable wear, as recommended by the eye care professional. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Dailies Total1 (delefilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
Dailies Total1 Multifocal (delefilcon A) soft contact lenses are indicated for the optical correction of presbyopia. with or without refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
Dailies Total1 for Astigmatism (delefilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with up to 6.00 diopters (D) of astigmatism.
Dailies Total 1 Multifocal Toric (delefilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to 6.00 diopters (D) of astigmatism.
Dailies Total1 lenses are to be used for single use, daily disposable wear (less than 24 hours while awake) only. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Precision1 (P1) (verofilcon A lens material) and Dailies Total1 (DT1) (delefilcon A) are soft contact lenses, intended for on-eye use in persons with healthy eyes that need vision correction as determined and fitted by an eye care professional. The lensed for single use, daily disposable wear (less than 24 hours while awake) only.
Precision1 (verofilcon A) and Dailies Total1 (delefilcon A) soft contact lenses are supplied sterile, immersed in buffered saline solution and packaged in individual foil-blister packs, which are terminally sterilized in a validated autoclave (moist heat, steam under pressure). The foil-blister pack system consists of a polypropylene (PP) blister shell sealed with a coated aluminum foil lidding. The blister packs are packaged into carton boxes available in different pack sizes.
Precision1 (verofilcon A) soft contact lenses are currently available in a spherical and a toric (for astigmatism) lens design.
Dailies Total 1 (delefilcon A) soft contact lenses are available in spherical, toric (for astigmatism) and multifocal toric lens designs.
I am sorry, but the provided text does not contain information about the acceptance criteria for a device, nor does it describe a study that proves a device meets such criteria.
The document is an FDA 510(k) clearance letter for Alcon's contact lenses (Precision1 and Dailies Total1 families). It primarily discusses:
- The FDA's determination of substantial equivalence to predicate devices.
- The general controls and regulations applicable to the devices.
- The indications for use for the various contact lens products.
- A summary of the device description and intended use.
- A statement that the submission is for a manufacturing change (modification of extraction and coating processes) and that no clinical testing was required to establish safety and effectiveness for this specific change. Instead, non-clinical bench testing demonstrated that the lenses meet established finished product specifications and are substantially equivalent to the predicate devices.
Therefore, I cannot fulfill your request as there is no information in the provided text regarding:
- A table of acceptance criteria and reported device performance.
- Sample sizes for a test set or data provenance.
- Number and qualifications of experts for ground truth establishment.
- Adjudication methods.
- MRMC comparative effectiveness studies.
- Standalone algorithm performance.
- Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
- Sample size for a training set.
- How ground truth for a training set was established.
These elements are typically found in documents describing clinical trials, performance studies for AI/ML-driven devices, or detailed design verification and validation reports, none of which are present in this FDA 510(k) clearance letter.
Ask a specific question about this device
(24 days)
Precision1; Precision1 for Astigmatism; TOTAL30; TOTAL30 for Astigmatism; TOTAL30 Multifocal
Precision1 (verofilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves and approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.
Precision1 for Astigmatism (verofilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes and 6.00 diopters (D) or less of astigmatism.
Precision1 lenses are to be prescribed for single use, daily disposable wear, as recommended by the eye care professional. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
TOTAL30 (lehfilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
TOTAL30 for Astigmatism (lehfilcon A) toric soft contact lenses are indicated for the optical correction ofrefractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes with up to 6.00 diopters (D) of astigmatism.
TOTAL30 Multifocal (lehfilcon A) soft contact lenses are indicated for the optical correction of presbyopia with or without refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
TOTAL30 lenses are to be prescribed for daily wear, with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, or disposal, as recommended by the eye care professional. Lenses should be discarded and replaced with a new pair each month, or more often, if recommended by the eye care professional.
Precision1 (verofilcon A) soft contact lenses are currently available in a spherical and a toric lens design.
Precision1 (verofilcon A) soft contact lenses are supplied sterile, immersed in buffered saline solution and packaged in individual foilblister packs, which are terminally sterlized in a validated autoclave (moist heat, steam under pressure). The foil-blister pack system consists of a polypropylene (PP) blister shell sealed with a coated aluminum foil lidding. The blister packaged into carton boxes available in different pack sizes.
Lehfilcon A soft contact lenses are available in a spherical design (for correction of vision in persons with myopia or hyperopia), toric design (for correction of myopia, with astigmatism) and multifocal design (for correction of presbyopia, with or without myopia or hyperopia) in a range of powers and parameters.
Lehfilcon A lenses immersed in package saline and provided sterile in sealed blister packages are fully functioning (i.e., ready to use) out of pack (no need for accessories). Lenses in sealed blister packs (primary packaging) are provided in an outer carton (secondary packaging). For daily wear/reuse, products such as commercially available soft contact lens care cleaning and disinfecting solutions, rewetting drops, saline solutions and contact lens cases may be used. Accessories are sold and provided separately from the device.
The provided FDA 510(k) clearance letter and summary for Alcon's contact lenses (Precision1, TOTAL30, etc.) do not contain any information about an AI/ML-driven medical device, a study involving AI assistance, or acceptance criteria related to a device's performance based on AI algorithms.
The document discusses soft contact lenses and their indications for use, and the 510(k) submission relates to:
- The "Trade/Device Name" and "Regulation Number/Name" for contact lenses.
- The legal basis for marketing these contact lenses based on substantial equivalence to predicate devices.
- Quality System (QS) regulations for manufacturing medical devices.
- Indications for Use for various contact lens types.
- A 510(k) summary indicating "no proposed device modifications related to this change only affects quality control measures." This means the submission is likely for a minor manufacturing or quality control change to existing contact lens products, not a new or modified AI device.
Therefore, I cannot extract the information required to answer your prompt about acceptance criteria and a study proving an AI device's performance, as this information is not present in the provided text. The prompt's questions pertain to AI/ML device validation, which is unrelated to the content of this FDA document for contact lenses.
Ask a specific question about this device
(29 days)
Precision1, Precision1 for Astigmatism
Precision1 (verofilcon A) Spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes and approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.
Precision1 for Astigmatism (verofilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes and 6.00 diopters (D) or less of astigmatism.
The lenses are to be prescribed for single use, as recommended by the eye care professional. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Precision1 (verofilcon A) soft contact lenses are currently available in a spherical and a toric lens design.
Precision1 (verofilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes and approximately 1.50 diopters of astigmatism that does not interfere with visual acuity.
Precision1 for Astigmatism (verofilcon A) toric soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes and 6.00 diopters (D) or less of astigmatism.
Precision 1 (verofilcon A) soft contact lenses are supplied sterile, immersed in buffered saline solution and packaged in individual foilblister packs, which are terminally sterilized in a validated autoclave (moist heat, steam under pressure). The foil-blister pack system consists of a polypropylene (PP) blister shell sealed with a coated aluminum foil lidding. The blister packaged into carton boxes available in different pack sizes.
This document describes the 510(k) premarket notification for the Precision1 and Precision1 for Astigmatism soft contact lenses. The submission is for a modification to an already legally marketed predicate device (Precision1 for Astigmatism, K182902). The modification involves adding foil from an alternate supplier for primary packaging and extending the product expiration dating from 72 to 84 months.
The document states that the scope of the device modification did not require clinical testing to establish safety and effectiveness. Therefore, the information typically associated with studies proving device performance against acceptance criteria (such as sample sizes, ground truth establishment, expert adjudication, or MRMC studies) is not present in this document. The device stability and process validation were completed to verify equivalence to the predicate device.
Here's the information that could be extracted from the document regarding acceptance criteria and the study, as much as applicable:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Equivalence to Predicate Device: Successful stability testing (to ensure product integrity over extended shelf life) | All acceptance criteria being met. |
Equivalence to Predicate Device: Process validation (to ensure manufacturing consistency and sterilization efficacy) | All acceptance criteria being met. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document as part of a clinical study. The document refers to "stability testing" and "process validation," which are typically laboratory-based tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the document states that clinical testing was not required for this submission. The "ground truth" for the non-clinical tests would be defined by pre-established scientific and engineering standards for stability and process validation, not by human expert consensus or clinical outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as there was no clinical test set requiring expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There was no MRMC comparative effectiveness study done. This submission is for a material and shelf-life extension change to an existing contact lens, not an AI-powered diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable as this device is a contact lens, not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical "stability testing" and "process validation," the ground truth would be established by scientific and engineering specifications and standards for contact lens materials, sterility, and performance over time. This is not medical expert consensus, pathology, or outcomes data.
8. The sample size for the training set
This information is not applicable as there was no training set for an algorithm or AI.
9. How the ground truth for the training set was established
This information is not applicable as there was no training set for an algorithm or AI.