510(k) Data Aggregation

    K Number
    K241353
    Date Cleared
    2024-11-27

    (198 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PowerLoc™ Max Power Injectable Infusion Set; SafeStep™ Huber Needle Set

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PowerLoc™ Max Power Injectable Infusion Set:

    The PowerLoc™ Max Power Injectable Infusion Set is a device intended for insertion into the septum of a subcutaneously implanted port for the infusion of fluids and drugs, as well as blood sampling through surgically implanted vasular ports. When used with ports that are indicated for power injection of contrast media into the central venous system, the PowerLoc™ Max Power Injectable Infusion Set is also indicated for power injection of contrast media. These devices may be used in any patient population with an implanted vascular port.

    For power injection of contrast media, the maximum recommended infusion rate at 11.8 cPs is 5 mL/s for 19 gauge and 20 gauge needles, and 2 mL/s for 22 gauge needles.

    SafeStep™ Huber Needle Set:

    The SafeStep™ Huber Needle Set is a device intended for insertion into the septum of a subcutaneously implanted port for the infusion of fluids and drugs, as well as blood sampling through surgically implanted vascular ports.

    These devices may be used in any patient population with an implanted vascular port.

    Note: The SafeStep™ Huber Needle Set is NOT indicated for power injection of contrast media or any other high pressure injection.

    Device Description

    PowerLoc™ Max Power Injectable Infusion set is a non-coring Huber needle and infusion set with a manually activated needlestick prevention safety mechanism. These devices access surgically implanted subcutaneous vascular ports by penetrating the port septum to provide a closed fluid pathway for the infusion of fluids and drugs, as well as blood sampling. In addition, it may be used for power injection of contrast media through an implanted vascular port that is also indicated for power injection up to 325 psi (2241 kPa). It is supplied sterile and non-pyrogenic, for single use only.

    The PowerLoc™ Max Power Injectable Infusion Set is offered with and without a Y-site.

    SafeStep™ Huber Needle Set:

    The SafeStep™ Huber Needle Set is a device intended for insertion into the septum of a subcutaneously implanted port for the infusion of fluids and drugs, as well as blood sampling through surgically implanted vascular ports.

    These devices may be used in any patient population with an implanted vascular port.

    Note: The SafeStep™ Huber Needle Set is NOT indicated for power injection of contrast media or any other high pressure injection.

    AI/ML Overview

    This document describes the premarket notification for the "PowerLoc™ Max Power Injectable Infusion Set" and "SafeStep™ Huber Needle Set". The study presented focuses on demonstrating substantial equivalence to a predicate device (K171735) through performance testing, rather than a standalone clinical study to establish new acceptance criteria. Therefore, much of the requested information regarding clinical studies, expert-established ground truth, and training data is not applicable in this context.

    Here's a breakdown of the available and applicable information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each test are implied by the "All testing passed the predetermined acceptance criteria" statement. Specific quantitative acceptance criteria or detailed performance results are not provided in this summary. Instead, the document lists the type of test conducted and the standard it references, indicating compliance.

    Test CategoryReference Standard / Test DescriptionDevice Performance (as reported)
    BiocompatibilityISO 10993-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. Tests include Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity, Subchronic Systemic Toxicity, Hemocompatibility.Passed, free from biological hazard, material acceptability confirmed by health-based risk assessment.
    Assembly BurstISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – Part 1: General requirements. Test to confirm assembly burst pressure exceeds peak pressure at maximum flow conditions.Passed (exceeded peak pressure).
    Assembly TensileISO 8536-4:2019 – Infusion equipment for medical use – Part 4: Infusion sets for single use, gravity feed. Test to demonstrate peak tensile force of each assembly exceeds minimum peak tensile force.Passed (exceeded minimum peak tensile force).
    Assembly Air LeakISO 8536-10:2015 – Infusion equipment for medical use – Part 10: Accessories for fluid lines for single use with pressure infusion equipment. Test to confirm no leak when occluded.Passed (no leak).
    Assembly High Pressure Water Leak/Burst (2X)ISO 8536-10:2015 – Infusion equipment for medical use – Part 10: Accessories for fluid lines for single use with pressure infusion equipment. Test to confirm PowerLoc™ Max does not leak or burst at maximum indicated flow rate.Passed (no leak or burst at max indicated flow rate).
    Assembly Occlusion/Flow RateISO 7864:2016: Sterile hypodermic needles for single use – Requirements and test methods. Test to confirm minimum flow rates.Passed (confirmed minimum flow rates).
    Clamp FunctionInternal Standard. Test to confirm fluids will not pass through tubing when the clamp is engaged.Passed (no fluid passage when clamp engaged).
    Luer Adapter TestingISO 80369-7: Small-bore connectors for liquids and gases in healthcare applications – Part 7: Connectors for intravascular or hypodermic applications. Tests include Leak, Leak Decay, Stress Cracking, Resistance to Separation from Axial Load, Unscrewing, Overriding.Passed (met all requirements).
    Particulate TestingUSP : Sizing and Counting Particulate Matter. Test to ensure particulate matter on the catheter post-manufacture is not exceeded for prescribed particle sizes.Passed (particulate matter levels not exceeded).
    Needleless Y-Site TestingGuidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)] (2008). Tests include Microbial Ingress, Valve Activation, Air Introduction Due to Device Connect/Disconnect.Passed (met all requirements for SafeStep™ Huber Needle Set configurations with Y-site).
    Residual Toxicity (Ethylene Oxide Residuals)ISO 10993-7:2008: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals. Test to ensure EO residuals are below requirements for any patient with an implanted port.Passed (EO residuals below required levels).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes used for each individual performance test. It broadly states "All testing passed the predetermined acceptance criteria."
    • Data Provenance: The studies were conducted "by or for Bard Access Systems (BD)". The data is presumably from laboratory testing, not patient data (retrospective or prospective), given the nature of the device and the tests performed. The country of origin for the data is not specified, but the applicant's address is in Salt Lake City, Utah, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The device's substantial equivalence was established through engineering and performance testing against recognized standards, not through clinical trials requiring expert-established ground truth on patient data.

    4. Adjudication Method for the Test Set

    This information is not applicable, as no clinical study with human observers or adjudication committees was performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This is not applicable. This is a medical device (infusion set and Huber needle set), not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests would be the established requirements and limits defined by the referenced ISO standards, USP standards, and internal protocols for a device of this type. For example, for "Assembly Burst," the ground truth is that the device must withstand a certain pressure without bursting, as defined by ISO 10555-1.

    8. The Sample Size for the Training Set

    This is not applicable. There is no AI component or training set involved in the assessment of this physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set. The "ground truth" for the device's acceptable performance is derived from established international and national standards for medical devices and internal quality control protocols.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153440
    Manufacturer
    Date Cleared
    2016-08-18

    (265 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PowerLoc MAX Power Injectable Infusion Set, SafeStep Huber Needle Set

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerLoc MAX® Power-Injectable Infusion Set is an intravascular administration set with a non-coring right angle needle and manually activated needle stick prevention safety mechanism which reduces the risk of accidental needlestick injuries by shielding the needle is used to access surgically implanted vascular ports.

    The PowerLoc MAX® Power-Injectable Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports.

    When used with ports that are indicated for power injection of contrast media into the central venous system, the PowerLoc MAX® Power-Injectable Infusion Set is also indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate at 11.8 cPs is 5 ml/s for 19 gauge and 20 gauge needles, and 2 ml/s for 22 gauge needles.

    The SafeStep® Huber Needle Set is a device intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids into the port. The SafeStep® Huber Needle safety feature is manually activated during needle removal, and is designed to aid in the prevention of accidental needlesticks.

    Device Description

    The PowerLoc MAX® Power-Injectable Infusion Set is a standard non-coring intravascular administration set with a non-coring Huber type right angle needle and a manually activated needle-stick prevention safety mechanism which reduces the risk of accidental needlestick injuries by shielding the needle. The device is used to access surgically implanted vascular ports and is indicated for use in the administration of fluids and drugs, as well as blood sampling. The PowerLoc MAX® Power-Injectable Infusion Set is also indicated for power injection of contrast media into the central venous system only through an implanted port that is also indicated for power injection. The maximum recommended infusion rate at 11.8 cPs is 5 ml/s for 19 gauge and 20 gauge needles and 2 ml/s for 22 gauge needles. The PowerLoc MAX® Power-Injectable Infusion Set is offered with and without a Y-site.

    The SafeStep® Huber Needle Set is a standard right angle Huber needle and administration set with a needlestick prevention feature, designed for use with a vascular access infusion system. It is manufactured with conventional medical grade, biocompatible materials. The SafeStep® Huber Needle Set operates as a standard Huber needle with the addition of a safety feature to aid in the prevention of needlestick injuries to the health practitioner. The SafeStep® Huber Needle Set is offered with and without a Y-site.

    The stabilization device is intended for use as an accessory to the subject PowerLoc MAX® and SafeStep® infusion sets, and is supplied pre-loaded with the subject devices. It is placed centrally over the implanted port so that the base surrounds the implanted port under the skin. The infusion set needle handle is pressed down until the needle has entered the port septum, then the stabilization device is lifted off of the needle. The stabilization device is then discarded per hospital protocol.

    AI/ML Overview

    The provided document is a 510(k) summary for medical devices (PowerLoc MAX® Power-Injectable Infusion Set and SafeStep® Huber Needle Set). It details the regulatory clearance for these devices. However, the document does not contain information about acceptance criteria and studies proving the device meets these criteria in the context of an AI/algorithm-driven device.

    The "Performance Tests" section details a series of tests conducted to ensure the device's physical and functional performance, materials, and safety in relation to its intended use as an intravascular administration set. These tests include:

    • Slide Force testing (to ensure easy needle slide within the stabilization device).
    • Stabilization Device Removal Force testing (to ensure easy removal of the stabilization device).
    • Acceptability of the Stabilization Device (clinical acceptability of placement).
    • Acceptability of GuardIVa® with Stabilization Device (validation of antimicrobial dressing use).
    • Various ISO standards compliance (e.g., sharps injury protection, infusion equipment, biocompatibility, sterilization, packaging, luer fittings, needle tubing, intravascular catheters, and hypodermic needles).
    • FDA Guidance compliance (e.g., sharps injury prevention features, MR compatibility).
    • Risk management (FMEA) in accordance with BS EN ISO 14971:2012.

    The document states: "The subject devices met all predetermined acceptance criteria derived from the above listed references and demonstrated substantially equivalent performance as compared to the cited predicate devices."

    Therefore, it is not possible to extract the requested information (sample size for test/training set, data provenance, number/qualification of experts, adjudication method, MRMC study, standalone performance, type of ground truth) because the supplied document describes the regulatory approval of a physical medical device (infusion set/needle) and not an AI/algorithm-driven device.

    The questions posed relate to the validation of AI/ML-based medical devices, which involve different types of studies (e.g., diagnostic accuracy, reader studies) and data considerations than those detailed for this hardware device.