510(k) Data Aggregation

    K Number
    K140310
    Manufacturer
    Date Cleared
    2015-09-03

    (573 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Peristeen Anal Irrigation System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peristeen™ Anal Irrigation System is intended to instill water into the colon through a rectal catheter - which incorporates an inflatable balloon - inserted into the rectum to promote evacuation of the contents of the lower colon. The Peristeen Anal Irrigation System is indicated for use by children (2 years -

    Device Description

    The Peristeen™ Anal Irrigation (PAI) System is a Class II device intended for intermittent use that facilitates emptying of the colon/bowel in patients with neurogenic bowel dysfunction. The PAI system consists of a single-use irrigation catheter that incorporates an inflatable balloon to keep the catheter in place during the procedure and retain the water that flows into the colon. The rectal catheter is non-sterile, intended for single-use, and packaged and labeled accordingly. The other components may be used multiple times; usage guidelines are detailed in the labeling. All System components are also provided separately in various accessory packages.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Peristeen™ Anal Irrigation System. It states that no changes have been made to the device design, materials, components, technological characteristics or manufacturing processes of the subject device compared to the predicate. The only change being effected is the release of an updated Instructions For Use (IFU).

    Therefore, the document explicitly states: "As such, testing for Performance, Shelf Life and Biocompatibility was not deemed necessary to support substantial equivalence. No clinical testing was performed, referenced, or relied on in the 510(k) for a determination of substantial equivalence."

    Because no new performance testing was conducted for this submission, the document does not contain the information required to answer the questions about acceptance criteria and a study proving the device meets those criteria.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    PERISTEEN ANAL IRRIGATION SYSTEM, PERISTEEN ANAL IRRIGATION ACCESSORY UNIT, PERISTEEN ANAL IRRIGATION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peristeen™ Anal Irrigation System is intended to instill water into the colon through a rectal catheter-which incorporates an inflatable balloon-inserted into the rectum to promote evacuation of the contents of the lower colon. The Peristeen™ Anal Irrigation System is indicated for use by children (2 years -

    Device Description

    The Peristeen™ Anal Irrigation system is a Class II device, consisting of a single-use irrigation catheter with balloon for retention; a control unit with a manual switch that allows for addition of pressure to the water bag, and inflation and deflation of the balloon on the catheter; a bag with a lid to hold water, leg straps that may be used to fasten the control unit and tubing to the thigh, and tubes with connectors. The system is provided with a nylon storage case. The rectal catheter is single-use, but the other components may be used multiple times.

    AI/ML Overview

    The provided FDA 510(k) summary for the Peristeen™ Anal Irrigation System (K112860) does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the context of an AI/ML device.

    This document describes a medical device (a rectal catheter system) that underwent biocompatibility and mechanical testing. It does not mention any AI/ML components or algorithms, nor does it discuss performance metrics like sensitivity, specificity, or AUC, which are typically found in studies validating AI/ML devices.

    Therefore, I cannot provide the requested information. The document explicitly states:

    • SUMMARY OF CLINICAL TESTS SUBMITTED (AS APPLICABLE): Not applicable

    This means no clinical trials were conducted or submitted as part of this 510(k) to evaluate the device's performance in human subjects, let alone an AI/ML component. The "nonclinical tests" listed are bench tests related to the physical properties and biocompatibility of the catheter.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    PERISTEEN ANAL IRRIGATION SYSTEM, IRRIGATION ACCESSORY UNIT, IRRIGATION RECTAL CATHETER, SYSTEM STRAP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peristeen™ Anal Irrigation (PAI) System is intended to instill water into the colon through a rectal catheter-which incorporates an inflatable balloon-inserted into the rectum to promote revacuation of the contents of the lower colon. The PAI System is indicated for use by Children (2 years -

    Device Description

    The Peristeen™ Anal Irrigation system consists of a single-use irrigation catheter with a balloon for retention; a control unit with a manual switch to add pressure to the water bag, inflate and deflate the balloon on the catheter; a bag with a lid to hold water or isotonic saline solution, leg straps that may be used to fasten the control unit and tubing to the thigh, and tubes with connectors. The system may be purchased with a carrying case (toilet bag). The rectal catheter is single-use, but the other components may be used multiple times. Accessory kits are available for the components. The PAI system does not contain natural rubber latex components.

    AI/ML Overview

    The provided text is a 510(k) summary for the Peristeen™ Anal Irrigation System, which is a medical device. It does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the typical sense of a clinical performance study with statistical metrics.

    The document states that the Peristeen™ Anal Irrigation System is substantially equivalent to a previously cleared device (K083770). The equivalence is supported by a comparison of design, materials, intended use, and acceptable results from functional performance and biocompatibility testing. This indicates that the "acceptance criteria" here are likely related to engineering specifications, safety standards, and material compatibility, rather than clinical performance metrics like sensitivity, specificity, or reader agreement.

    Since the document focuses on substantial equivalence for a physical medical device, not an AI/algorithm-based diagnostic tool, most of the requested information (like sample size for test sets, expert ground truth, MRMC studies, standalone performance, training set details) is not applicable or not provided in this specific type of regulatory filing.

    However, I can extract the relevant information that is present:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategoryReported Device Performance (Summary)
    Functional PerformanceAcceptable results (implied by substantial equivalence claim)
    BiocompatibilityAcceptable results (implied by substantial equivalence claim)
    Design ComparisonSubstantially equivalent to predicate device
    Materials ComparisonSubstantially equivalent to predicate device
    Intended Use ComparisonExpanded indications, but substantially equivalent to predicate device

    2. Sample size used for the test set and the data provenance:

    • Not Applicable/Not Provided. The submission relies on functional performance and biocompatibility testing and substantial equivalence to a predicate device, not clinical performance data from a patient test set in the context of diagnostic accuracy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable/Not Provided. This information is relevant for diagnostic accuracy studies, which are not detailed in this 510(k) summary for a physical irrigation system.

    4. Adjudication method for the test set:

    • Not Applicable/Not Provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-based diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI/algorithm-based diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For functional performance and biocompatibility testing, the ground truth would typically be established by established engineering standards, material science testing protocols, and biological safety evaluations.
    • For substantial equivalence, the ground truth is the performance, design, and indications for use of the predicate device.

    8. The sample size for the training set:

    • Not Applicable/Not Provided. This information is relevant for AI models, not for a physical medical device like the Peristeen™ Anal Irrigation System.

    9. How the ground truth for the training set was established:

    • Not Applicable/Not Provided.

    Summary of the "study" that proves the device meets acceptance criteria:

    The "study" referenced in the document is a comparison to a legally marketed predicate device (K083770) combined with functional performance and biocompatibility testing.

    • Comparison to Predicate: The report states: "Substantial equivalence of the Peristeen™ Anal Irrigation System is supported by a comparison of the design, materials, and intended use compared to the predicate..." This comparison essentially serves as the primary "study" for establishing regulatory acceptance based on the 510(k) pathway. If the new device is sufficiently similar to an already cleared device in terms of safety and effectiveness, it can be cleared.
    • Functional Performance and Biocompatibility Testing: The document mentions "acceptable results from functional performance and biocompatibility testing." While the details of these tests are not provided in this summary, they would involve engineering tests to ensure the device operates as intended (e.g., proper water flow, balloon inflation/deflation, pressure control) and material tests to ensure the components are safe for human contact (e.g., ISO 10993 series of standards). The "acceptance criteria" for these would be specific engineering and biocompatibility thresholds defined prior to testing.