Search Results
Found 47 results
510(k) Data Aggregation
(189 days)
The Pulse Oximeter is intended for spot-checking in measuring and displaying functional arterial hemoglobin (SpO2) and pulse rate. It is intended for adult, adolescent, child users in hospitals, hospital facilities and home healthcare environments.
The Pulse Oximeter model BM20, can display the SpO2 and PR, plethysmogram wave and other indication parameters, such as pulse battery power status and the PI. The model BM20 Pulse Oximeter include the mainboard, display and lithium battery. It has small size, light weight, easy to carry. It adopts low power consumption design, and has the function of battery capacity display.
The pulse oximeter, model BM20, is designed for spot checking of the pulse oxygen saturation and pulse rate for adults, pediatric and adolescent patients in professional healthcare facility or home conditions when physician follow-up and operated by a physician. And it is not intended to be used under motion or low perfusion scenarios. This medical device can be reused. Not for continuously monitoring.
N/A
Ask a specific question about this device
(158 days)
The Pulse Oximeter is a non-invasive device indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult patients. It is intended for spot- check and/or continuous data collection, and not continuous monitoring. It can be used in sleep labs, long-term care, hospitals and home use.
The Pulse Oximeter is a lightweight, portable health finger ring oximeter for home or healthcare facilities. SpO2 measurement technology is based on the developed photoelectron method and circuit design, and Shenzhen Viatom Technology Co., Ltd. developed calculation software. The SpO2 sensor receives the optical signal from the red light and infrared light through the finger into the oximeter. Two emitting tubes (red light diodes and infrared diodes) are located on the inner right side of the sensor, and they can emit red light and infrared. The receiving end is located on the inner left side of the sensor, and it can receive the red light and infrared through the finger. The MCU receives the pulse signal, gets the frequency signal by counting, processes its digital signal, and finally gets the measured SpO2 value. The PR is calculated on average by the above peak intervals of the PR waveform.
The provided text is a 510(k) summary for a Pulse Oximeter (Models PO2, PO2A, PO2B) and does not contain information about an AI/ML-driven device. Therefore, it is not possible to describe acceptance criteria or a study related to an AI/ML device based on this document.
The document primarily focuses on demonstrating the substantial equivalence of the subject pulse oximeter to a predicate pulse oximeter (K191088 Checkme O2 Pulse Oximeter) by meeting established performance standards for pulse oximeters, such as ISO 80601-2-61.
Here's an analysis based on the information available in the document, demonstrating why it doesn't fit the AI/ML framework you've described:
General Device Performance (Pulse Oximeter):
-
A table of acceptance criteria and the reported device performance:
The document does not present a formal "acceptance criteria table" in the context of an AI/ML model. Instead, it provides a comparison table of the subject device's specifications and performance metrics against a predicate device.Characteristic Acceptance Criteria (Implicit from Predicate & Standards) Reported Device Performance (Subject Device) SpO2 Accuracy (70-100%) ±2% 1.77% ARMS Pulse Rate Accuracy ±2bpm or ±2% (whichever is greater) ±2bpm or ±2% (whichever is greater) SpO2 Measurement Accuracy (ARMS) ≤ 2% (from ISO 80601-2-61) 1.77% Work Mode Spot-check and continuous data collection (not continuous monitoring) Spot-check and continuous data collection (not continuous monitoring) Intended Application Site Finger Finger -
Sample sized used for the test set and the data provenance:
- Test Set Description: The "clinical validation testing of the SpO2 performance" was conducted on "healthy adult volunteers."
- Sample Size: The exact number of healthy adult volunteers is not specified in the document.
- Data Provenance: The document states "Measured values are from a controlled lab study in healthy volunteers." It does not specify the country of origin. The study appears to be prospective, as it involved actively testing subjects with the device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This question is not applicable to this device. For a pulse oximeter, "ground truth" for SpO2 and pulse rate is established using a co-oximeter and ECG (or similar reference standard) on induced hypoxia studies, as per ISO 80601-2-61. There are no "experts" in the AI/ML sense establishing ground truth labels for images or other complex data. The "ground truth" clinical values are collected by qualified personnel trained in conducting such studies. -
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., radiology reads) to resolve discrepancies and establish a consensus ground truth. For a pulse oximeter, the reference measurements from the co-oximeter are the ground truth, and human adjudication is not part of the process. -
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This question is not applicable. The device is not an AI-assisted diagnostic tool. No MRMC study was performed or is relevant for this type of device. -
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This question is not applicable in the context of an AI/ML algorithm. The performance of the pulse oximeter itself (which contains an algorithm to calculate SpO2 and PR from optical signals) was evaluated in a standalone manner against reference standards in the clinical study. The device's performance is inherently "standalone" in its measurement function, as it provides direct numerical outputs. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For SpO2 accuracy, the ground truth was arterial oxygen saturation measured by a co-oximeter (from blood samples) during induced hypoxia, as per ISO 80601-2-61. For pulse rate, the ground truth would typically be derived from an electrocardiogram (ECG) monitor. This is a physiological reference standard, not expert consensus or pathology. -
The sample size for the training set:
This question is not applicable. The document describes a traditional medical device (pulse oximeter) that uses a pre-defined algorithm based on physical principles (absorption of red and infrared light). It does not mention any "training set" for an AI/ML model. The "algorithm" for SpO2 calculation is based on principles of spectroscopy and physiological models, not machine learning from a dataset in the way a deep learning model would be trained. -
How the ground truth for the training set was established:
This question is not applicable for the same reasons as #8. There is no training set for an AI/ML model described.
In summary, the provided document details the regulatory clearance for a traditional medical device (pulse oximeter) based on established performance standards, and therefore, the questions tailored for AI/ML device evaluations are generally not relevant.
Ask a specific question about this device
(97 days)
The Pulse Oximeter FS20P is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
The Pulse Oximeter FS20C and FS10C is a handheld non-invasive device intended for spotchecking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
The subject device Pulse Oximeter is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. Place one fingertip into the photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. The device is normally applied to infants, children, adolescents and adult in hospitals, hospital-type facilities and homecare.
The subject device is composed of following components to achieve the above detection process: power supply module, detector and emitter LED, signal collection and process module (MCU), OLED/LED display screen, user interface and button control circuit.
The document provided is a 510(k) Premarket Notification from the FDA for a Pulse Oximeter. It describes the device, its intended use, and a comparison to a predicate device, including non-clinical and clinical data.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Acceptance Criteria Table and Reported Device Performance
| Acceptance Criteria (Stated or Implied) | Reported Device Performance |
|---|---|
| SpO2 Accuracy (for FS20P model)Measurement Accuracy: 70%~100%: ±2% (Implied from "Accuracy: 70% | FS20P Pulse Oximeter:ARMS (Accuracy Root Mean Square) of 1.70% over the range of 70-100%. This value 1.70% is less than or equal to the implied acceptance criteria of ±2% as defined by the predicate device and the new device's listed specification for accuracy. |
| SpO2 Accuracy (for FS20C model)Measurement Accuracy: 70%~100%: ±2% (Implied from "Accuracy: 70% | FS20C Pulse Oximeter:ARMS (Accuracy Root Mean Square) of 1.71% over the range of 70-100%. This value 1.71% is less than or equal to the implied acceptance criteria of ±2% as defined by the predicate device and the new device's listed specification for accuracy. |
| BiocompatibilityMust conform to the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1. | The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". (Statement of compliance, not a numerical performance metric). |
| Bench Testing / Electrical SafetyConform to specified IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61). | The Pulse Oximeter has been tested according to the following standards:- IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012- IEC 60601-1-2:2014- IEC 60601-1-11 Edition 2.0 2015-01- ISO 80601-2-61: 2017(Statement of compliance, not a numerical performance metric, beyond the clinical accuracy which is detailed elsewhere). Electrical safety explicitly stated: "Conformed to IEC60601-1, IEC 60601-1-11". |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- FS20P: 10 healthy female adult volunteer subjects.
- FS20C: 13 healthy adult volunteer subjects.
- Data Provenance: The studies involved volunteer subjects, implying prospective data collection. The document does not specify the country of origin of the study data, but the submitting company is based in China, so it is highly likely the studies were conducted there. The subject skin tones (Fitzpatrick 1-6) suggest a diverse population, but this is a characteristic, not a location.
3. Number of Experts and Qualifications for Ground Truth
- The document describes a clinical study for SpO2 accuracy compared to arterial blood CO-Oximetry. This is a direct physiological measurement, implying that no human expert adjudication was used to establish the ground truth for SpO2. The CO-Oximeter itself is the "expert" or gold standard.
- Therefore, the concept of "number of experts" for ground truth establishment, as typically applied to image-based diagnostic AI, is not directly applicable here.
4. Adjudication Method for the Test Set
- None, as the ground truth was established by direct physiological measurement using a CO-Oximeter, which is considered a gold standard for SpO2. No human adjudication process (like 2+1 or 3+1) was necessary or mentioned for the SpO2 values.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging AI devices where human readers interpret medical images with and without AI assistance.
- For a pulse oximeter, the device directly measures physiological parameters (SpO2 and PR). The clinical study focused on validating the device's accuracy against a gold standard (arterial blood CO-Oximetry), not on improving human reader performance.
6. Standalone (Algorithm Only) Performance
- Yes, a standalone performance study was done. The clinical studies directly measured the device's (i.e., the pulse oximeter's algorithm/measurement system) SpO2 accuracy (ARMS) against a gold standard (arterial blood CO-Oximetry). The reported ARMS values (1.70% for FS20P and 1.71% for FS20C) are metrics of the device's standalone performance.
7. Type of Ground Truth Used
- The ground truth used was outcomes data / physiological measurement, specifically arterial blood CO-Oximetry. This method provides a direct, highly accurate measurement of arterial oxygen saturation, serving as the gold standard for SpO2.
8. Sample Size for the Training Set
- The document does not specify a separate training set or its sample size. For a traditional medical device like a pulse oximeter, particularly one based on well-established principles of optical measurement, there isn't typically an "AI training set" in the sense of machine learning. The device's calibration and design would be based on engineering principles and potentially internal validation data, but not typically a labeled "training set" like an AI performs. The studies described are for validation/testing the accuracy of the final device.
9. How Ground Truth for the Training Set Was Established
- This question is not applicable as there is no mention of an "AI training set" or a separate training set in the context of this device's submission. The described clinical studies are for performance validation (test set), not for training an algorithm.
Ask a specific question about this device
(287 days)
The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
The oximeter consists of probe, electronic circuits, and display and plastic enclosures. And one side of probe is designed to locate light emitting diodes and a light detector (called a photo-detector). Red and Infrared lights are shone through the tissues from one side of the probe to the other. Then parts of the light emitted absorbed by blood and tissues. The light absorbed by the blood varies with the oxygen saturation of haemoglobin. After that, the photo-detector detects the light volume transmitted through the tissues which depends on blood pulse, Hereafter, the microprocessor calculates a value for the oxygen saturation (SpO2). The subject device is a reusable device, and need to reprocess as suggested in the user manual after each use. And the device is intended to be used on the finger, and powered by 2*1.5V AAA battery.
The provided document is a 510(k) summary for the Shenzhen Yimi Life Technology Co., Ltd. Pulse Oximeter (various models). It details the device's characteristics and its substantial equivalence to a predicate device. Here's an analysis of the acceptance criteria and the study information as requested:
1. Table of Acceptance Criteria and Reported Device Performance
The document explicitly refers to acceptance criteria for SpO2 accuracy and Pulse Rate accuracy based on the ISO 80601-2-61 standard.
| Acceptance Criteria (from ISO 80601-2-61) | Reported Device Performance |
|---|---|
| SpO2 Accuracy | |
| 70-100%: ±2% | 70-100%: ±2% |
| 0-69%: Undefined | 0-69%: Undefined |
| Pulse Rate Accuracy | |
| Not explicitly stated in the document, but the device "meets the requirements defined in ISO 80601-2-61, Clause 201.12.1.104" | ±3 bpm (This is the predicate's declared accuracy, and the subject device's declared accuracy is ±2 bpm, which is better and thus meets or exceeds the predicate's and implied standard.) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample size for the clinical test set. It only states that "Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-61:2011". Annex EE of ISO 80601-2-61 generally recommends a minimum of 10 healthy subjects for SpO2 accuracy testing, though more may be required depending on the specific design of the study and the range of SpO2 values assessed.
- Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given it's a 510(k) submission for a Chinese manufacturer, the clinical testing was likely conducted in China or a similar jurisdiction following international standards. The nature of SpO2 accuracy testing typically involves prospective studies where subjects are intentionally desaturated under controlled conditions.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts or their qualifications for establishing ground truth in the clinical test set. For pulse oximetry, the ground truth for SpO2 accuracy is typically established by co-oximetry measurements from arterial blood samples, which does not directly involve "experts" in the sense of clinical reviewers for image analysis. The "ground truth" (reference standard) in this context is the quantitative co-oximetry reading.
4. Adjudication Method for the Test Set
Not applicable in the context of pulse oximetry SpO2 accuracy testing. Ground truth is established by objective co-oximetry measurements, not through expert consensus or adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results, and the AI's impact on their performance is evaluated. Pulse oximeters provide a direct numerical output (SpO2 and pulse rate), so MRMC studies are not applicable.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, a standalone performance study was done. The entire premise of the SpO2 accuracy testing described (clinical data per ISO 80601-2-61 Annex EE) is to evaluate the device's ability to accurately measure oxygen saturation and pulse rate independently, without human interpretation influencing the measurement. The device's output is compared directly against the reference standard (co-oximetry).
7. Type of Ground Truth Used
The ground truth used for SpO2 accuracy testing is co-oximetry measurements from arterial blood samples. This is the gold standard for determining actual arterial oxygen saturation. For pulse rate, the ground truth would typically be from an ECG monitor or a reference pulse rate measurement device.
8. Sample Size for the Training Set
The document does not provide information about a "training set" or its sample size. Pulse oximeters traditionally rely on established physiological principles and algorithms (e.g., Beer-Lambert law and pulsatile blood flow), rather than machine learning models that require extensive training data in the same way an AI image analysis algorithm would. Therefore, a distinct "training set" as understood in AI/ML development is not typically applicable for this type of device.
9. How the Ground Truth for the Training Set was Established
As no training set is described, this question is not applicable. The device's underlying principles are physics-based, and its performance is validated against physiological standards with direct measurements.
Ask a specific question about this device
(192 days)
WS20A is a pulse oximeter indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for patients(≥10 years old, ≥30kg). It is designed for finger circumference more than 33mm. It is intended for spot-check, continuous data collection, recording and transmitting, not continuous monitoring. It can be used in sleep labs, long-term care, hospitals and home.
Pulse Oximeter WS20A is an internally powered pulse oximeter. The main functions of the devices include hemoglobin oxygen saturation (SpO2), pulse rate (PR) measurements and Pulse amplitude index (PAI), data storage and transmission. Place one fingertip into the sensor and the oxygen saturation (SpO2), pulse rate (PR) measurements and pulse amplitude index (PAI) will appear on the display. The device is intended to be applied to adult and pediatric patients in sleep labs, long-term care, hospital and home care environment. The subject device is composed of the following components to achieve the above detection process: power supply module, detector and emitter, signal collection and process module (MCU), TFT display screen and Bluetooth module.
The provided text describes the 510(k) summary for the Hunan Accurate Bio-Medical Technology Co., Ltd.'s Pulse Oximeter (WS20A). While it details the device, its intended use, and comparison to a predicate device, it does not contain a specific table of acceptance criteria with reported device performance in the format of typical medical device performance claims (e.g., sensitivity, specificity, accuracy for an AI/diagnostic device).
Instead, the performance criteria for this pulse oximeter are related to its accuracy in measuring SpO2 and Pulse Rate against a reference standard (CO-Oximetry). The acceptance criteria for pulse oximeters are typically defined by recognized standards such as ISO 80601-2-61, which specifies the accuracy requirements (e.g., ARMS value).
Based on the provided text, here's an attempt to extract the relevant information and present it as requested.
Overview of Device Performance and Study
The Pulse Oximeter (WS20A) underwent clinical studies to verify its accuracy in measuring functional oxygen saturation of arterial hemoglobin (SpO2) against arterial blood CO-Oximetry, in accordance with ISO 80601-2-61:2017 and FDA guidance for Pulse Oximeters - Premarket Notification Submissions.
1. Table of Acceptance Criteria and Reported Device Performance
For pulse oximeters, the key performance metric for SpO2 accuracy is typically the Accuracy Root Mean Square (ARMS) value when compared to a reference method (CO-Oximetry). The acceptance criterion is generally that the ARMS value should be within a specified limit, commonly ± 3% for 70-100% SpO2 range as per ISO 80601-2-61.
| Performance Metric | Acceptance Criteria (ARMS for SpO2, 70-100% range) | Reported Device Performance (ARMS for SpO2, 70-100% range) |
|---|---|---|
| SpO2 Accuracy | $\leq \pm 3$% (derived from ISO 80601-2-61) | Adult Patients: - All subjects: 1.81% - Female subjects: 1.80% - Male subjects: 1.83% - Light-skinned: 1.78% - Dark-skinned: 1.91% Pediatric Patients: - 10 female subjects: 1.80% - Light-skinned: 1.81% - Dark-skinned: 1.74% |
| PR Accuracy (bpm) | $\pm$ 3bpm (as per predicate device spec) | Not explicitly reported from clinical study for WS20A. The 510(k) submission document for WS20A lists its PR accuracy as $\pm$ 3 bpm, which aligns with its own stated specifications and is compared to a similar specification for the predicate. |
Note: The acceptance criteria for ARMS are not explicitly stated as a number in the provided text but are inferred from common regulatory expectations for pulse oximeters based on the ISO standard cited.
2. Sample Size Used for the Test Set and Data Provenance
- Adult Patients: 13 healthy adult subjects. 315 data sets (SpO2 vs SaO2) were obtained.
- Pediatric Patients: 10 healthy female subjects. 243 data sets (SpO2 vs SaO2) were obtained. (Note: The description states "healthy female subjects, aged 22-30" for 'pediatric patients' which seems to be a typo given the 'pediatric' labeling and the age range overlaps with 'adult'. Assuming this refers to subjects representing a pediatric patient population as defined by the study design, or a mislabeling of the subject group based on physical characteristics matching the pediatric profile for device testing.)
- Data Provenance: The document does not specify the country of origin of the data. The studies were described as "clinical studies" and appear to be prospective in nature, as they involved actively obtaining data sets from healthy subjects under stationary (non-motion) conditions through controlled desaturation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The ground truth for pulse oximetry accuracy studies is established through arterial blood CO-Oximetry, which is a direct invasive measurement of arterial oxygen saturation (SaO2). This is a gold standard laboratory method, not typically established by physician experts in the same way as image interpretation. The text does not mention the use of experts for ground truth establishment; rather, it relies on the direct physiological measurement.
4. Adjudication Method for the Test Set
Not applicable. Ground truth for oxygen saturation in these studies is derived from direct physiological measurement (CO-Oximetry) rather than expert consensus requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a pulse oximeter for direct physiological measurement, not an AI-assisted diagnostic imaging device requiring human reader interpretation for a comparative effectiveness study.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the performance presented (ARMS values) represents the standalone accuracy of the pulse oximeter device algorithm in measuring SpO2 against the CO-Oximetry reference. While a human uses the device, the accuracy is derived from the device's measurement capabilities.
7. The Type of Ground Truth Used
The ground truth used was arterial blood CO-Oximetry (SaO2), which is a gold standard for measuring oxygen saturation in the blood. This is a direct physiological measurement, not expert consensus or pathology.
8. The Sample Size for the Training Set
The document describes clinical studies conducted to verify the accuracy of the proposed device. It does not explicitly mention a "training set" for an algorithm in the sense of machine learning. The data described (from 13 adult and 10 "pediatric" subjects) appears to be the test/validation set used to demonstrate the device's accuracy. For traditional medical devices like pulse oximeters, the core measurement principles are physics-based, not reliant on machine learning models that require distinct training sets. Any internal calibration or algorithm development would typically occur during the R&D phase, prior to these described validation studies, and the size of data used for such internal development is not generally disclosed in 510(k) summaries unless it's a novel AI/ML device.
9. How the Ground Truth for the Training Set was Established
Not applicable, as a distinct "training set" in the context of an AI/ML device is not described for this traditional pulse oximeter. If there were internal algorithm development or calibration, the ground truth would likely have been established using similar reference methods (CO-Oximetry) in controlled lab settings.
Ask a specific question about this device
(283 days)
The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in home and healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
The Pulse Oximeter is a battery powered device in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 m, which is red light; the other is 905 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The device mainly composed of PCB board, On/Off button, mode button, OLED&LED screen, battery compartment, Bluetooth® module and plastic shell. There are five models AOJ-70A, AOJ-70C, AOJ-70E. Only AOJ-70E. Only AOJ-70D have wireless connection function via Bluetooth®. The device is a spot-check pulse oximeter and does not intended for life-supporting or life-sustaining.
The provided text is a 510(k) summary for a Pulse Oximeter (AOJ-70A-E). It describes the device, its intended use, and comparative testing to show substantial equivalence to a predicate device. However, it does not provide a table of acceptance criteria with reported device performance or information regarding a typical "AI/algorithm" study (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, types of ground truth, etc.). This document primarily focuses on regulatory compliance for a medical device (pulse oximeter), not an AI/ML-based diagnostic or assistive technology.
Therefore, I cannot fulfill the request as the necessary information (acceptance criteria table, detailed study design for an AI/ML model, expert data, ground truth establishment for AI/ML) is not present in the provided text. The text only mentions "software verification and validation testing" and "clinical data is referenced to K221039" for SpO2 accuracy, which refers to a standard medical device clinical trial for a pulse oximeter, not an AI/ML performance study.
The closest information found is the clinical study for SpO2 accuracy performance of the pulse oximeter itself, not an AI/ML component. Here's what is available regarding that:
Study on SpO2 Accuracy Performance (Not an AI/ML study):
- Study Name/Reference: Referenced to K221039.
- Location: Sir Run Run Shaw Hospital, Zhejiang University School of Medicine.
- Methodology: Conducted in accordance with ISO 14155-1, ISO 80601-2-61:2017, and FDA Guidance Document for Pulse Oximeters. Evaluated SpO2 accuracy performance during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels compared to arterial blood CO-Oximetry (which served as ground truth).
- Sample Size: 11 healthy adult volunteer subjects (ages 21-47yr, with skin tones varying from Fitzpatrick 2-6) were included.
- Ground Truth: Arterial blood CO-Oximetry.
- Reported Performance: The SpO2 accuracy performance results showed that the subject device had an ARMS (Accuracy Root Mean Square) < 2 during steady-state conditions over the range of 70-100%.
Missing Information (for an AI/ML study):
- A table of acceptance criteria with reported device performance (for an AI/ML algorithm).
- Sample sizes for AI/ML test sets.
- Data provenance for AI/ML datasets.
- Number of experts used to establish ground truth for AI/ML test sets and their qualifications.
- Adjudication method for AI/ML test sets.
- Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done for AI assistance.
- Effect size of human reader improvement with AI vs. without AI assistance.
- Standalone AI/ML algorithm performance.
- Sample size for training set (for AI/ML).
- How ground truth for the training set was established (for AI/ML).
The document is a regulatory submission for a traditional medical device (pulse oximeter), and the testing described is standard for such devices, not for AI/ML-based medical software.
Ask a specific question about this device
(274 days)
The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent child and infant patients in hospitals, hospital-type facilities and homecare. The device can be used by the people whose finger thickness is between 8mm and 22mm(0.3 inches to 0.9 inches).
The proposed device, Pulse Oximeter, is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. The MD300C228 is adopted LCD screen to display to display SpO2, Pulse Rate value (PR), Perfusion Index (PI), pulse bar, brightness level, battery indicator, signal indicator and waveform, it has 7 display modes, the brightness level can be adjusted 1-5 level. The measured data can be transmitted to APP through Bluetooth. The device is normally applied to adult, adolescent child and infant patients in hospitals, hospital-type facilities and homecare.
The subject device is composed of following components to achieve the above detection process: power supply module, detector and emitter LED, signal collection and process module (MCU), LCD display screen, user interface and button control circuit.
Principle of the oximeter is as follows: The pulse oximeter works by applying a sensor to a fingertip. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660mm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The enclosure of the subject device is made of ABS and the fingertip cushion is made of Silicone Gel.
The subject device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile, and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
The device does not contain drug or biological products.
The provided text describes the acceptance criteria and a clinical study conducted for the Pulse Oximeter MD300C228. This device is a pulse oximeter, not an AI/ML device, so certain requested details such as number of experts for ground truth, adjudication methods, MRMC studies, and training set information are not applicable in this context. However, I will extract and present the relevant information provided for this medical device.
Acceptance Criteria and Device Performance:
The primary performance metric mentioned is SpO2 accuracy.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| SpO2 Accuracy | Arms of 1.8 during steady state conditions over the range of 70-100% SaO2. |
| SpO2 Display Range | 0%~100% |
| SpO2 Measurement Range | 70%~100% (Accuracy defined for 70-100%, no definition for 0-69%) |
| SpO2 Resolution | 1% |
| PR Display Range | 30 bpm~250 bpm |
| PR Measurement Range | 30 bpm~250 bpm |
| PR Accuracy | 30 bpm |
| PR Resolution | 1 bpm |
| PI Display Range | 0.1%~20% |
| PI Measure Range | 0.3~20.0% |
| PI Resolution | 0.1% |
| Working Time | Continuously operated as long as 20 hours |
| Operating Temperature | 0°C~40°C |
Study Details:
-
Sample Size and Data Provenance:
- Sample Size: 11 healthy adult volunteer subjects (ages 20-42yr, with light to dark pigmentation, including male and female).
- Data Provenance: The document does not explicitly state the country of origin for the clinical study, but the manufacturer is based in Beijing, China. The study appears to be prospective, as it involved inducing hypoxia in volunteers to collect data.
-
Number of Experts and Qualifications for Ground Truth:
- This criterion is not applicable as the "ground truth" for SpO2 accuracy was established through direct measurement using a reference co-oximeter, not through expert consensus or interpretation of images.
-
Adjudication Method:
- Not applicable, as the ground truth was based on direct co-oximeter measurements, not subjective evaluations requiring adjudication.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- Not applicable. This is a study to validate the accuracy of a standalone pulse oximeter device, not an AI-assisted diagnostic tool for human readers.
-
Standalone Performance:
- Yes, a standalone performance study was conducted. The clinical study aimed to evaluate the SpO2 accuracy performance of the pulse oximeter during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood co-oximeter. The reported Arms of 1.8% is a measure of the device's standalone accuracy.
-
Type of Ground Truth Used:
- Physiological Measurement/Reference Device: The ground truth for SpO2 accuracy was established by analyzing arterial blood samples with a reference co-oximeter, providing functional SaO2.
-
Sample Size for Training Set:
- Not applicable. This is a hardware pulse oximeter device, not an AI/ML algorithm that requires a training set. The device's performance is based on its optical and signal processing design.
-
How Ground Truth for Training Set was Established:
- Not applicable as there is no training set for this type of device.
Ask a specific question about this device
(337 days)
The Pulse Oximeter is intended for spot-checking oxygen saturation and pulse rate, and the device is a reusable device and intended to be used with the finger of adults in professional healthcare facility or home conditions when physician follow-up and operated by a physician. And it is not intended to be used under motion.
The pulse oximeter, Model OHT60 and OXH78, consists of probe, electronic circuits, and display and plastic enclosures, is designed for spot checking of the pulse oxygen saturation and pulse rate for adults in a professional healthcare facility or home conditions when physician follow-up and operated by a physician. This medical device can be reused. Not for continuously monitoring.
SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The measurement principle of Pulse oximeter is based on the absorption spectrum characteristics of hemoglobin and oxyhemoglobin in red and infrared regions, and the empirical formula of data is established by using ' Lambert Beer ' law. The working principle of the instrument adopts photoelectric blood oxygen detection technology combined with volume pulse recording technology, The specific process is as follows: Firstly, the infrared light emitting at 660 nm and the near-infrared light emitting at 905 are used to irradiate the finger and the measurement data are obtained by the infrared receiver tube. Secondly, the obtained measurement data were calculated and processed by electronic circuits and microprocessors to obtain the blood oxygen saturation and pulse rate, and the calculation results were displayed on the screen.
The subject device is a reusable device, and need to reprocess as suggested in the user manual after each use. And the device is intended to be used on the finger, and powered by 2*1.5V AAA alkaline batteries.
The Model OHT60 and OXH78 have the same critical components and its materials and means of power supply, PCB layout and Software are little different.
The provided text describes the 510(k) summary for a Pulse Oximeter (Models OHT60, OXH78). This document primarily details the device's technical specifications, non-clinical testing, and a clinical study to demonstrate its accuracy against a predicate device. It does not describe a study involving an AI component or a multi-reader multi-case (MRMC) comparative effectiveness study, as the device is a standalone hardware product (pulse oximeter).
Therefore, I will focus on the acceptance criteria and the study that proves the device meets those criteria based on the information available for this specific medical device.
Device Name: Pulse Oximeter (Model OHT60, OXH78)
Intended Use: Spot-checking oxygen saturation and pulse rate in adults, reusable device, for use with the finger in professional healthcare facilities or home conditions with physician follow-up. Not intended for use under motion.
Acceptance Criteria and Reported Device Performance
The primary performance criterion for this pulse oximeter is its accuracy in measuring SpO2. The relevant standard cited is ISO 80601-2-61:2017. This standard typically specifies the acceptable accuracy (Root Mean Square, ARMS) for pulse oximeters over a defined SpO2 range.
Table 1: Acceptance Criteria and Reported Device Performance (SpO2 Accuracy)
| Criterion (from ISO 80601-2-61:2017 & FDA Guidance) | Acceptance Level | Reported Device Performance | Comments |
|---|---|---|---|
| SpO2 Accuracy (ARMS) for 70-100% SpO2 range | ≤ 3.0 % (Typical) | 1.98 % | Meets the generally accepted criterion for pulse oximeter accuracy within the specified range. |
Note: While the predicate device had an SpO2 accuracy of ±2% and the subject device has ±3%, the document states "The SpO2 measurement range and accuracy is a little different, but they are within the limits of the specified range of the standard ISO 80601-2-61 and FDA guidance for Pulse Oximeters. The different has no effect on substantive equivalence." This implies that the ±3% for the subject device is within the acceptable range defined by the standard and FDA guidance.
Study Details to Prove Device Meets Acceptance Criteria
The study described is a clinical accuracy study for the pulse oximeter.
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 12 healthy adult volunteer subjects.
- Data Provenance: Clinical study conducted at Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, China (implied from the submitter's country and study location). The study was prospective as it involved inducing fractional inspiration O2 (FiO2) control tests and obtaining arterial blood samples.
- Subject Demographics: Ages 21-42 years, 46-75 kg, 150-180 cm. Included 4 males and 8 females, with varying skin pigmentations (4 subjects with Fitzpatrick V, VI, and 8 subjects with Fitzpatrick I ~ Fitzpatrick IV).
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- The ground truth for SpO2 (SaO2) was established by an arterial blood gas analyzer (CO-OXIMETER), which is a laboratory instrument. Experts were involved in conducting the clinical study, drawing arterial blood, and operating the analyzer, but the ground truth itself is an instrumental measurement, not expert consensus reading of images or similar. The document does not specify the number or specific qualifications of medical professionals involved in operating the CO-OXIMETER or conducting the clinical measurements beyond being part of the clinical study team at the hospital.
-
Adjudication Method for the Test Set:
- Not applicable as the ground truth (SaO2) was obtained directly from an arterial blood gas analyzer, not through human interpretation requiring adjudication.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, an MRMC comparative effectiveness study was not done. This device is a direct measurement device (pulse oximeter), not an AI-assisted diagnostic imaging tool that would typically involve human readers.
-
If a Standalone (algorithm only without human-in-the-loop performance) was done:
- Yes, in essence, the clinical study assessed the "standalone" performance of the pulse oximeter device itself by comparing its SpO2 readings directly against the functional oxygen saturation (SaO2) from the arterial blood gas analyzer. The device provides a direct measurement, not an output that requires further human interpretation for its primary function.
-
The Type of Ground Truth Used:
- Outcomes Data/Physiological Measurement: The ground truth used was functional oxygen saturation (SaO2) measured by an arterial blood gas analyzer (CO-OXIMETER) from stable arterial blood samples. This is considered a gold standard for blood oxygen measurement.
-
The Sample Size for the Training Set:
- The document does not specify a separate "training set" in the context of machine learning. The device is a hardware product relying on established physiological principles (spectrophotometry, photoplethysmography) and empirical formulas, not a "learnable" AI algorithm in the contemporary sense that would require a distinct training set for model development. The clinical study data (12 subjects) described is the test set used for validation of the device's accuracy.
-
How the Ground Truth for the Training Set Was Established:
- Not applicable, as a distinct training set for an AI model is not described or relevant for this type of device. The underlying principles and empirical formulas are based on general scientific and medical knowledge, rather than specific data training.
Ask a specific question about this device
(204 days)
The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent child and infant patients in hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
The subject device Pulse Oximeter is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. Place one fingertip into the photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. The device has 2 display modes. The device is normally applied to adult, adolescent child and infant patients in hospitals, hospital-type facilities and homecare. The subject device is composed of following components to achieve the above detection process: power supply module, detector and emitter LED, signal collection and process module (MCU), LED display screen, user interface and button control circuit. Principle of the oximeter is as follows: The pulse oximeter works by applying a sensor to a fingertip. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The enclosure of the subject device is made of ABS and the fingertip cushion is made of Silicone Gel. The subject device is not for life-supporting or life-sustaining, not for implant. The device is not sterile, and the transducers are reusable and do not need sterilization and re-sterilization. The device is for prescription. The device does not contain drug or biological products.
Here's a breakdown of the acceptance criteria and the study details for the Pulse Oximeter, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria (from predicate/standards) | Reported Device Performance (Subject Device MD300C19) |
|---|---|---|
| SpO2 Accuracy | 70%~100%, ±2% | ARMS of 1.5% (over 70-100% SpO2 range) |
| SpO2 Display Range | 0~100% | 0~100% |
| SpO2 Measurement Range | 70%~100% | 70%~100% |
| SpO2 Resolution | 1% | 1% |
| PR Display Range | 30bpm~250bpm (Predicate) | 0bpm~255bpm |
| PR Measurement Range | 30bpm~250bpm | 30bpm~250bpm |
| PR Accuracy | 30bpm | 30bpm |
| PR Resolution | 1bpm | 1bpm |
| Operating Temperature | 5°C~40°C (Predicate) | 0°C~40°C |
| Biological Evaluation | ISO 10993-1, -5, -10 | Compliant (Biocompatibility tests done) |
| Electrical Safety | IEC60601-1, IEC60601-1-11 | Compliant |
| EMC | IEC60601-1-2 | Compliant |
| Software | FDA Guidance for Software in Medical Devices | Compliant |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 11 healthy adult volunteer subjects.
- Data Provenance: The study was conducted in a controlled clinical setting. The specific country of origin is not explicitly stated, but the manufacturer is based in China. The study was prospective in nature, as it involved actively recruiting subjects and conducting measurements to evaluate the device's performance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts used or their qualifications for establishing the ground truth. It states that the ground truth was obtained via "arterial blood CO-Oximetry," which is a laboratory method, not an expert assessment.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method involving experts for the test set. The ground truth was established through a direct objective measurement (arterial blood CO-Oximetry).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study focuses on the standalone performance of the Pulse Oximeter against an objective ground truth, not on evaluating human reader performance with or without AI assistance.
6. Standalone (Algorithm Only) Performance
Yes, the study primarily assessed the standalone performance of the Pulse Oximeter (algorithm only, as it's a direct measurement device). The reported ARMS value of 1.5% is a measure of the device's accuracy in determining SpO2 levels without human interpretation beyond reading the display.
7. Type of Ground Truth Used
The ground truth used was arterial blood CO-Oximetry. This is considered a gold standard for measuring oxygen saturation in arterial blood.
8. Sample Size for the Training Set
The document does not provide any information about a training set or its sample size. Pulse oximeters, particularly those based on established principles like spectrophotometry, often rely on physics-based algorithms and extensive calibration rather than machine learning models requiring large training datasets in the same way AI algorithms do.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned, there is no information on how its ground truth was established.
Ask a specific question about this device
(822 days)
The pulse oximeter is a non-invasive device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios. The device is reusable.
The oximeter consists of probe, electronic circuits, and display and plastic enclosures. And one side of probe is designed to locate light emitting diodes and a light detector (called a photo-detector). Red and Infrared lights are shone through the tissues from one side of the probe to the other. Then parts of the light emitted absorbed by blood and tissues. The light absorbed by the blood varies with the oxygen saturation of haemoglobin. After that, the photo-detector detects the light volume transmitted through the tissues which depends on blood pulse, Hereafter, the microprocessor calculates a value for the oxygen saturation (SpO2). The subjected device is a reusable device, and need to reprocess as suggested in the user manual after each use. And the device is intended to be used on the finger, and powered by 2*1.5V AAA battery.OP-101 display the measuring results on 1.5' LED screen, and the backlight of the three models are red, white and green respectively. And the screen of OP-102 and OP-103 are 0.96' OLED and 1.3' OLED. Additionally, battery indicator and pulse waveform can be displayed on OP-102 and OP-103. The subjected device is indicated for continuous or spot check monitoring of functional arterial oxygen saturation (SpO2) and pulse rate of adult patients in hospitals and clinics.
This document describes the validation of a Pulse Oximeter device (models OP-101 and OP-102).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily derived from the ISO 80601-2-61:2017 standard for pulse oximeters, particularly for SpO2 accuracy.
| Metric | Acceptance Criteria (ISO 80601-2-61:2017) | Reported Device Performance (OP-101, OP-102) | Predicate Device Performance (C101H1) |
|---|---|---|---|
| SpO2 Accuracy (70-100% range) | (Not explicitly stated as a single value criterion in the provided text, but implied by adherence to ISO 80601-2-61 guideline which involves comparing device SpO2 readings to arterial oxygen saturation (SaO2) measurements obtained from a co-oximeter.) The standard requires the root mean square deviation (ARMS) between SpO2 and SaO2 to be within a specified range, and typically, individual readings within a certain accuracy range (e.g., ±2% or ±3%). | 70-100%: ±2% | 70-100%: ±3% |
| SpO2 Accuracy (0-69% range) | Unspecified per the standard's current scope for typical physiological measurements. | 0-69%: unspecified | 0-69%: unspecified |
| Pulse Rate Accuracy | Not explicitly detailed but implied that the device meets the "pre-specified criteria" of the clinical trial and related standards. | ±3 bpm or ±1%, whichever is greater | ±1 bpm or ±1%, whichever is greater |
| Biocompatibility | Compliance with ISO 10993-1, ISO 10993-5, ISO 10993-10 | Passes Cytotoxicity, Skin Sensitization, Skin Irritation tests | Compliance with ISO 10993-1, ISO 10993-5, ISO 10993-10 |
| Electrical Safety | Compliance with IEC 60601-1 and IEC 60601-1-11 | Complies with IEC 60601-1: 2005+CORR. 1 (2006)+CORR. 2 (2007)+AM1 (2012) and IEC 60601-1-11:2015 | Complies with IEC 60601-1 and IEC 60601-1-11 |
| Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 | Complies with IEC 60601-1-2: 2014 | Complies with IEC 60601-1-2 |
| Performance Effectiveness (General) | Compliance with ISO 80601-2-61:2017 | Complies with ISO 80601-2-61: 2017 | Complies with ISO 80601-2-61 |
| Software Verification and Validation | Adherence to FDA Guidance for content of premarket submissions for software contained in medical devices (moderate level of concern) | Software works as expected, stable performance | (Not explicitly stated for predicate in comparison table, but assumed compliance for clearance) |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set (Clinical Data): 12 adult volunteers were used to validate the accuracy of the finger pulse oximeter (Model: OP-101).
- Data Provenance: The clinical study was conducted outside of the United States. It was a prospective, randomized clinical study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number or qualifications of experts directly involved in establishing the ground truth measurements. However, the ground truth was established by measuring SaO2 with a blood gas analyzer, which is a standard method that implicitly relies on trained medical professionals to operate and interpret.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for the test set in the sense of expert review of discrepancies. The accuracy validation appears to be a direct comparison between the device's SpO2 readings and the SaO2 measurements from the blood gas analyzer.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study was done or reported in this document. This type of study is more common for diagnostic imaging devices where human interpretation is a key component. For a pulse oximeter, the primary comparison is between the device's output and a gold standard physiological measurement.
6. Standalone Performance Study
Yes, a standalone performance study was done. The clinical study described involved evaluating the algorithm's performance (SpO2 and pulse rate measurements) against a recognized gold standard (blood gas analyzer) without human input / assistance in the measurement itself (beyond initial placement of the device and blood sample collection).
7. Type of Ground Truth Used
The ground truth for the clinical accuracy validation was arterial oxygen saturation (SaO2) measured by a blood gas analyzer. This is considered a gold standard for oxygen saturation in medical contexts.
8. Sample Size for the Training Set
The document does not explicitly mention a separate "training set" or its sample size. For medical devices like pulse oximeters, the development typically involves calibration data, and potentially internal validation data, but the focus of regulatory submissions is often on the performance demonstration of the final device (test set). Given the nature of a hardware-based measurement device, "training set" in the machine learning sense is less applicable here unless specific algorithms for signal processing or artifact rejection were developed with a dedicated training phase, which is not detailed.
9. How the Ground Truth for the Training Set Was Established
As no specific "training set" is described for this device in the provided document, the method for establishing its ground truth is not detailed.
Ask a specific question about this device
Page 1 of 5