Search Results
Found 7 results
510(k) Data Aggregation
(23 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES, WHITE
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent containation between healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powered Vinyl Patient Examination Gloves, 80 L YZ, and meets all 21 C requirements of ASTM standard D-5250-00E4.
This document describes the 510(k) summary for Super Safe Plastic Products Co., Ltd.'s Powdered Vinyl Patient Examination Gloves, White. The summary focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
ASTM D-5250-00E4 Standard (Physical & Dimensions) | All testing meets requirements for physical and dimensions testing conducted on gloves. |
FDA 1000 ml Water Fill Test (Pinhole) | Sampling conducted at AQL 2.5, inspection level I, meeting requirements. |
Primary Skin Irritation | Testing conducted with results showing no primary skin irritant reactions. |
Skin Sensitization (Allergic Contact Dermatitis) | Testing conducted with results showing no sensitization reactions. |
Residual Powder (ASTM D-6124-06 for Starch) | Testing conducted to ensure gloves meet "powdered" claims (contain no more than 10mg/dm2). |
2. Sample Size Used for the Test Set and Data Provenance
The document provides the following information:
- ASTM D-5250-00E4 (Physical & Dimensions): "Inspection level S-2, AQL 4.0." This refers to sampling plans defined within the ASTM standard. The specific sample size is not explicitly stated in the provided text but would be determined by these parameters.
- FDA 1000 ml Water Fill Test (Pinhole): "Samplings of AQL 2.5, inspection level I." Similar to the above, this indicates the sampling plan, but the exact numerical sample size is not presented.
- Primary Skin Irritation and Skin Sensitization: The document doesn't specify sample sizes for these tests.
- Residual Powder Test: The document doesn't specify sample size for this test.
Data Provenance: Not explicitly stated, but based on the manufacturer's location (Shijiazhuang, Hebei, China), the testing was likely conducted in China. The document does not specify if the data is retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The tests performed are objective, laboratory-based tests (e.g., physical measurements, water fill tests, chemical tests, biocompatibility tests). They do not involve expert interpretation or ground truth establishment in the way, for example, a diagnostic imaging study would.
4. Adjudication Method for the Test Set
Not applicable, as the tests are objective and do not require expert adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device for personal protection (gloves), not an AI-powered diagnostic or assistive tool. MRMC studies are not relevant to its evaluation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for the tests performed is based on:
- Established ASTM Standards: For physical properties, dimensions, and residual powder content. These standards define objective criteria and test methods.
- FDA Requirements: For the pinhole test (1000 ml Water Fill Test).
- Biocompatibility Testing Protocols: For primary skin irritation and skin sensitization, which involve established biological assessment methods.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/machine learning model and does not involve a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable.
Summary of the Study that Proves the Device Meets Acceptance Criteria:
The study proving the device meets acceptance criteria is a series of non-clinical tests conducted on the Super Safe Plastic Products Co., Ltd. Powdered Vinyl Patient Examination Gloves, White. These tests were designed to demonstrate compliance with recognized industry standards and regulatory requirements for patient examination gloves.
The key aspects of this study include:
- Compliance with ASTM D-5250-00E4: The gloves were tested for physical and dimensional characteristics, with all results meeting the requirements of this ASTM standard. The sampling plan for these tests was based on "Inspection level S-2, AQL 4.0."
- Pinhole Integrity (FDA 1000 ml Water Fill Test): The gloves underwent the FDA 1000 ml Water Fill Test, meeting the requirements with a sampling plan of "AQL 2.5, inspection level I."
- Biocompatibility Testing:
- Primary Skin Irritation: Tests were conducted, showing no primary skin irritant reactions.
- Skin Sensitization (Allergic Contact Dermatitis): Tests were conducted, showing no sensitization reactions.
- Residual Powder Content: A Residual Powder Test based on ASTM D-6124-06 was performed to ensure the gloves meet the "powdered" claim, meaning they contain no more than 10mg/dm2 of starch.
The conclusions of these tests collectively demonstrated that the device "conform[s] fully to ASTM D-5250-00E4 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims." The results provided the basis for the FDA's determination of substantial equivalence to the predicate device.
Ask a specific question about this device
(112 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that mects all of the requirements of ASTM standard D 5250-0004.
-
Table of Acceptance Criteria and Reported Device Performance:
Characteristics Acceptance Criteria (Standard) Reported Device Performance Dimension ASTM standard D 5250-00c4 Meets Physical Properties ASTM standard D 5250-00c4 Meets Freedom from pinholes 21 CFR 800.20 Meets Powder Amount ASTM standard D 5250-00c4 Meets Biocompatibility (extract.) Not specified (implied
Ask a specific question about this device
(32 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES WHITE (NON-COLORED)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard I) 525()-00'd
The provided text is for a 510(k) premarket notification for "Powdered Vinyl Patient Examination Gloves, White (Non-colored)." It describes the device, its intended use, and indicates its substantial equivalence to a predicate device. The information focuses on regulatory compliance based on established standards rather than a study proving the device meets acceptance criteria through clinical or comparative effectiveness trials.
Here's a breakdown based on your request, highlighting the information that is not present in the provided document, as it's not typically required for this type of device and submission:
Acceptance Criteria and Reported Device Performance
Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-00e4 | Meets |
Physical Properties | ASTM standard D 5250-00e4 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Amount | ASTM standard D 5250-00e4 | Meets |
Biocompatibility: Primary Skin Irritation (Rabbits) | Not a Primary Skin Irritation (Implicitly: Passes standard for no irritation) | Passes ( |
Ask a specific question about this device
(47 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00-4
Here's an analysis of the provided text regarding acceptance criteria and study information for the "Powdered Vinyl Patient Examination Gloves, White (Non-colored)":
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-00c4 | Meets |
Physical Properties | ASTM standard D 5250-00c4 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Amount | ASTM standard D 5250-00c4 |
Ask a specific question about this device
(30 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)
powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00-4.
This document is a 510(k) Premarket Notification summary for medical gloves, not a study of an AI device. As such, it does not contain the information requested in your prompt regarding AI device acceptance criteria and study details.
However, I can extract the relevant information about the device performance and acceptance criteria for the medical gloves as presented in this document.
1. A table of acceptance criteria and the reported device performance:
Characteristics | Standard / Acceptance Criteria | Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-00e4 | Meets |
Physical Properties | ASTM standard D 5250-00e4 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Amount | ASTM standard D 5250-00e4 |
Ask a specific question about this device
(14 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-0004
The provided document describes the safety and effectiveness information for "Powdered Vinyl Patient Examination Gloves, White (Non-colored)" submitted by ZIBO ZHUANGYUAN PLASTIC & RUBBER CO.,LTD. The study presented is a non-clinical evaluation comparing the device's technical characteristics against established standards.
Here's an analysis of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance:
Characteristics | Standard | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Dimension | ASTM standard D 5250-00e4 | Meets ASTM standard D 5250-00e4 requirements | Meets |
Physical Properties | ASTM standard D 5250-00e4 | Meets ASTM standard D 5250-00e4 requirements | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets 21 CFR 800.20 requirements (implying a specific AQL for waterleak test) | Meets |
Powder Amount | ASTM standard D 5250-00e4 |
Ask a specific question about this device
(22 days)
POWDERED VINYL PATIENT EXAMINATION GLOVES, WHITE (NON-COLORED)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00c4
The provided text describes the regulatory submission for Powdered Vinyl Patient Examination Gloves. It details the device's characteristics and its equivalence to a predicate device, primarily through adherence to established standards.
Here's an analysis of the acceptance criteria and the study information, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-00e4 | Meets |
Physical Properties | ASTM standard D 5250-00e4 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Amount | ASTM standard D 5250-00e4 | Meets |
Biocompatibility | Primary Skin Irritation in rabbits | Passes, Not a Primary Skin Irritation ( |
Ask a specific question about this device
Page 1 of 1