Search Results
Found 58 results
510(k) Data Aggregation
(187 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The subject device is disposable medical PVC gloves that made from PVC compound, Clear (non-colored), powder free and nonsterile. The device meets the specifications in ASTM D5250-06(Reapproved 2015) Standard specification for poly (vinyl chloride) gloves for medical application.
This document is a 510(k) Premarket Notification summary for "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". It describes the device, its intended use, and the studies conducted to demonstrate substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
Test Methodology | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D5250-06 (Reapproved 2015) | Dimension (mm) IL S-2 AQL4.0 | ||
Length | $\geq$ 230 | 233-241 | |
Width (Small) | 80-90 | 81-89 | |
Width (Medium) | 90-100 | 92-99 | |
Width (Large) | 100-110 | 102-109 | |
Width (X-Large) | 110-120 | 112-119 | |
Thickness (Fingertip) | $\geq$ 0.05 | 0.09-0.11 | |
Thickness (Palm) | $\geq$ 0.08 | 0.10-0.11 | |
ASTM D5250-06 (Reapproved 2015) | Physical Properties IL S-2 AQL4.0 | ||
Tensile strength | $\geq$ 14MPa (Before & After aging) | 17-25 MPa (implicitly for before & after, as it covers the range) | |
Elongation (Before aging) | $\geq$ 500% | 560-610% | |
Elongation (After aging) | $\geq$ 400% | 460-570% | |
21 CFR 800.20, ASTM D5250-06 (Reapproved 2015), ASTM D5151-19 | Freedom from pinholes | Waterleakage test: Inspection Level I, AQL2.5, and Accept/Reject criteria of 10/11. (5 noncompliance is allowed.) | Pass |
ASTM D5250-06 (Reapproved 2015), ASTM D6124-06 (Reapproved 2017) | Powder Residual | $ |
Ask a specific question about this device
(60 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Poly Vinyl Chloride (PVC) films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
This document pertains to the 510(k) premarket notification for Powder Free Vinyl Patient Examination Gloves, Clear (non-colored). The information provided heavily focuses on proving substantial equivalence to a predicate device through non-clinical performance data, primarily compliance with established ASTM and ISO standards for medical gloves.
Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance
Characteristics | Acceptance Criteria | Reported Device Performance |
---|---|---|
Dimensions | ASTM D 5250-06 (Reapproved 2015) | Met ASTM D 5250-06 (Reapproved 2015) |
Length | ≥230mm | Subject Device: 231-241mm (Predicate: 231-242mm) - Stated as "Similar" |
Width (Small) | 80-90 mm | Subject Device: 81-89mm (Predicate: 82-89mm) - Stated as "Similar" |
Width (Medium) | 90-100mm | Subject Device: 93-99mm (Predicate: 93-99mm) - Stated as "Similar" |
Width (Large) | 100-110mm | Subject Device: 102-110mm (Predicate: 103-109mm) - Stated as "Similar" |
Width (X-Large) | 110-120 mm | Subject Device: 111-119mm (Predicate: 114-119mm) - Stated as "Similar" |
Thickness (Fingertip) | ≥0.05mm | Subject Device: 0.09-0.10mm (Predicate: 0.05-0.10mm) - Stated as "Similar" |
Thickness (Palm) | ≥0.08mm | Subject Device: 0.10-0.11mm (Predicate: 0.09-0.13mm) - Stated as "Similar" |
Physical Properties | ASTM D 5250-06 (Reapproved 2015) | Met ASTM D 5250-06 (Reapproved 2015) |
Tensile Strength | ≥11MPa (Before & After aging) | Subject Device: 15-22 MPa (Predicate: 16-20 MPa) - Stated as "Similar" |
Elongation Rate | ≥300% (Before & After aging) | Subject Device: 350-420% (Predicate: 380-410%) - Stated as "Similar" |
Freedom from Pinholes | 21 CFR 800.20 | |
ASTM D5250-06 (Reapproved 2015) | ||
ASTM D5151-19 | ||
Holes at Inspection Level I AQL2.5 | Subject Device: "Passed Standard Acceptance Criteria" | |
Holes at Inspection Level I AQL2.5 (Same as predicate) | ||
Powder Residual | ASTM D5250-06 (Reapproved 2015) | |
ASTM D6124-06 (Reapproved 2017) |
Ask a specific question about this device
(60 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Poly Vinyl Chloride (PVC) films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
This document is a 510(k) Pre-Market Notification for Powder Free Vinyl Patient Examination Gloves. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a novel device meets specific acceptance criteria in a clinical setting.
Therefore, many of the requested sections are not applicable in this context. However, I can extract information related to the device's technical specifications and the non-clinical tests performed to ensure it meets established safety and performance standards for gloves.
Here's a breakdown of the relevant information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally defined by compliance with established ASTM and ISO standards for medical examination gloves. The reported device performance is presented as meeting these standards.
Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | Width (mm) | |
Width (Small) | 80-90 mm | 82-88 |
Width (Medium) | 90-100 mm | 94-98 |
Width (Large) | 100-110 mm | 103-109 |
Width (X-Large) | 110-120 mm | 114-117 |
Thickness (mm) | ||
Fingertip | ≥0.05 mm (ASTM D 5250-06) | 0.09-0.10 |
Palm | ≥0.08 mm (ASTM D 5250-06) | 0.10-0.11 |
Physical Properties | ASTM D 5250-06 (Reapproved 2015) | |
Tensile Strength (Before & After aging) | ≥11 MPa | 15-22 MPa |
Elongation Rate (Before & After aging) | ≥300% | 350-420% |
Freedom from Pinholes | Holes at Inspection Level I AQL2.5 (21 CFR 800.20, ASTM D5250-06, ASTM D5151-19) | Met Acceptance Criteria; Holes at Inspection Level I AQL2.5 |
Powder Residual |
Ask a specific question about this device
(175 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The proposed device is Powder Free Examination Gloves. The proposed device is Clear (non-colored). The proposed device is non-sterile.
Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on what's available and noting what's not:
1. Table of Acceptance Criteria and Reported Device Performance
Test/Characteristic | Acceptance Criteria | Reported Device Performance (K191292) |
---|---|---|
Dimensions - Length | ASTM D 5250-06 (Reapproved 2011): ≥ 230mm | 233-241mm |
Dimensions - Width | ASTM D 5250-06 (Reapproved 2011): | |
Small 80-90 mm | ||
Medium 90-100mm | ||
Large 100-110mm | ||
X large 110-120 mm | 84-88 mm | |
94-98 mm | ||
103-109 mm | ||
114-117 mm | ||
Dimensions - Thickness | ASTM D 5250-06 (Reapproved 2011): | |
Fingertip ≥0.05mm | ||
Palm ≥0.08mm | Fingertip: 0.08-0.11mm | |
Palm: 0.10-0.11mm | ||
Physical Properties (Tensile Strength) | ASTM D 5250-06 (Reapproved 2011): ≥11MPa (Before & After aging) | 15-17 MPa (for "Physical Properties") |
Physical Properties (Elongation Rate) | ASTM D 5250-06 (Reapproved 2011): ≥300% (Before & After aging) | 350-390% (for "Physical Properties") |
Freedom from Pinholes | ASTM D5151-06 (Reapproved 2011): Holes at Inspection Level I, AQL2.5, Act/Re: 10/11 | Water leakage test: 3 noncompliance is allowed (This is a different test/criterion reported in the comparison table, but within the "similar" category for pinholes). The detailed result for the ASTM D5151-06 specific criterion is not explicitly stated as "Pass" or "Fail" in the result column, but the overall "similar" comparison implies conformity. |
Powder Residual | ASTM D6124-06 (Reaffirmation 2011): |
Ask a specific question about this device
(247 days)
Jaysun Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
PVC films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
This document, K182575, is a 510(k) Premarket Notification for "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". As such, it does not contain the type of acceptance criteria and study data typically associated with a "device" in the context of an AI/ML medical device, which would involve performance metrics like accuracy, sensitivity, specificity, and a detailed human-in-the-loop (HITL) study.
This submission is for a Class I medical device (patient examination glove), which, by regulation, has a much lower bar for demonstrating substantial equivalence to a predicate device. The "performance data" here refers to physical and biological characteristics of the glove, not diagnostic or predictive performance.
Therefore, I cannot provide the requested information in the format of AI/ML device studies because the provided document does not pertain to such a device. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to manufacturing and material standards, not clinical diagnostic accuracy.
However, I can extract the information relevant to this specific device type and its "studies":
Acceptance Criteria and Device Performance (for a physical medical device: Examination Glove)
Acceptance Criteria Category | Specific Acceptance Criteria (Standard & Value) | Reported Device Performance (Subject Device K182575) |
---|---|---|
Dimensions | ASTM D5250-06 (Reapproved 2011) | Meets ASTM D5250-06 (Reapproved 2011) |
Length | ≥230mm (Predicate) | 231-241mm (Subject Device) - Note: Predicate shown as 232-243mm, Subject shows 231-241mm and a comparison "Same" with Predicate. |
Width (Small) | 80-90mm (Predicate) | 81-89mm (Subject Device) - Note: Predicate shows 82-88mm, Subject shows 81-89mm and a comparison "Same" with Predicate. |
Width (Medium) | 90-100mm (Predicate) | 93-99mm (Subject Device) - Note: Predicate shows 93-98mm, Subject shows 93-99mm and a comparison "Same" with Predicate. |
Width (Large) | 100-110mm (Predicate) | 102-110mm (Subject Device) - Note: Predicate shows 102-109mm, Subject shows 102-110mm and a comparison "Same" with Predicate. |
Width (X-Large) | 110-120mm (Predicate) | 111-119mm (Subject Device) - Note: Predicate shows 113-118mm, Subject shows 111-119mm and a comparison "Same" with Predicate. |
Thickness (Finger) | ≥ 0.05mm (Predicate) | 0.05-0.10mm (Subject Device) - Note: Predicate shows 0.08-0.11mm, Subject shows 0.05-0.10mm and a comparison "Same" with Predicate. |
Thickness (Palm) | ≥ 0.08mm (Predicate) | 0.09-0.13mm (Subject Device) - Note: Predicate shows 0.10-0.11mm, Subject shows 0.09-0.13mm and a comparison "Same" with Predicate. |
Physical Properties | ASTM D5250-06 (Reapproved 2011) | Meets ASTM D5250-06 (Reapproved 2011) |
Elongation (Before/After aging) | ≥ 300% (Predicate) | 340-410% (Subject Device) - Note: Predicate shows 350-420%, Subject shows 340-410% and a comparison "Same" with Predicate. |
Tensile Strength (Before/After aging) | ≥ 14MPa (Predicate) | 15-25 MPa (Subject Device) - Note: Predicate shows 15-20 MPa, Subject shows 15-25 MPa and a comparison "Same" with Predicate. |
Freedom from Pinholes | ASTM D5151-06 (Reapproved 2015), Holes at Inspection Level I AQL 2.5 (Predicate: LAOL 2.5) | Meets ASTM D5151-06 (Reapproved 2011), Holes at Inspection Level I AQL 2.5 |
Residual Powder | ASTM D5250-06 (Reapproved 2011) and D6124-06 (Reaffirmation 2011): |
Ask a specific question about this device
(79 days)
Kindguard Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
PVC films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
The provided text is an FDA 510(k) summary for "Kindguard Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a medical device's performance against specific clinical efficacy or diagnostic accuracy criteria as would be required for a novel AI/imaging device.
Therefore, many of the requested categories (e.g., sample size for test/training set, number of experts, adjudication methods, MRMC studies, effect size, type of ground truth) are not applicable to this type of medical device submission. This submission relies on established performance standards for examination gloves.
Here's how the available information from the document applies to your request:
1. Table of Acceptance Criteria and Reported Device Performance:
The document compares the subject device's performance against established ASTM standards and FDA requirements, as well as against the predicate device.
Characteristics | Standard / Acceptance Criteria | Predicate Device Reported Performance (K091663) | Subject Device Reported Performance (K173017) | Result of Comparison |
---|---|---|---|---|
Dimension | ASTM D5250-06 (Reapproved 2011) | |||
Length | ≥ 230mm | 231-241mm (Range) | 232-241mm (Range) | Same |
Width (Small) | 80-90 mm | 81-89 mm (Range) | 82-88 mm (Range) | Same |
Width (Medium) | 90-100mm | 93-99 mm (Range) | 93-98 mm (Range) | Same |
Width (Large) | 100-110mm | 102-110 mm (Range) | 102-108 mm (Range) | Same |
Width (X Large) | 110-120 mm | 111-119 mm (Range) | 113-118 mm (Range) | Same |
Thickness (Fingertip) | ≥ 0.05mm | 0.05-0.10mm (Range) | 0.08-0.11mm (Range) | Same |
Thickness (Palm) | ≥ 0.08mm | 0.09-0.13mm (Range) | 0.10-0.12mm (Range) | Same |
Physical Properties | ASTM D5250-06 (Reapproved 2011) | |||
Elongation (Before & After aging) | ≥ 300% | 340-410% (Range) | 350-420% (Range) | Same |
Tensile Strength (Before & After aging) | ≥ 11MPa | 15-25 MPa (Range, specified as "> 14MPa" for predicate) | 15-20 MPa (Range) | Same |
Freedom from Pinholes | 21 CFR 800.20 Passed Standard Acceptance Criteria; ASTM D5250-06; ASTM D5151-06 | Meets ASTM D5151-06 (Reapproved 2011), ASTM D5250-06 (Reapproved 2011) | Meets ASTM D5151-06 (Reapproved 2011), ASTM D5250-06 (Reapproved 2011) | Same (Holes at Inspection Level IAQL2.5) |
Powder Residual | **ASTM D5250-06 (Reapproved 2011), ASTM D6124-06 (Reaffirmation 2011) Meets |
Ask a specific question about this device
(132 days)
DAIYANG Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
DAIYANG Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
The provided document is a 510(k) summary for the "DAIYANG Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". It describes the device, its intended use, and a comparison to a predicate device to establish substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance:
Characteristics | Standard / Acceptance Criteria | Reported Device Performance (DAIYANG Powder Free Vinyl Patient Examination Gloves) | Predicate Device Performance (K091663) | Result of Comparison (DAIYANG vs. Predicate) |
---|---|---|---|---|
Dimension: | ASTM D 5250-06(Reapproved 2011) | Meets ASTM D5250-06 | Similar | |
Length | ≥230mm | 231-241mm | 232-240mm | Similar |
Width (Small) | 80-90 mm | 81-89 | 81-87 | Similar |
Width (Medium) | 90-100mm | 93-99 | 93-98 | Similar |
Width (Large) | 100-110mm | 102-110 | 102-109 | Similar |
Width (X Large) | 110-120 mm | 111-1119 (likely typo, should be 111-119) | 113-118 | Similar |
Thickness (Fingertip) | ≥0.05mm | 0.05-0.10 | 0.08-0.11 | Similar |
Thickness (Palm) | ≥0.08mm | 0.09-0.13 | 0.10-0.12 | Similar |
Physical Properties: | ASTM D 5250-06(Reapproved 2011) | Before aging/after aging: Elongation ≥300%, Tensile Strength ≥ 14MPa | Similar | |
Tensile strength | ≥11MPa (Before & After aging) | 15-25 MPa | 15-20 MPa | Similar |
Elongation rate | ≥300% (Before & After aging) | 340-410% | 350-420% | Similar |
Freedom from pinholes | 21 CFR 800.20 Passed Standard Acceptance Criteria; ASTM D5250-06; ASTM D5151-06 | Holes at Inspection Level I AQL2.5 | Holes at Inspection Level I AQL2.5 | Similar |
Powder Residual | ASTM standard D 5250-06 (Reapproved 2011) and D6124-06 (Reaffirmation 2011). Meets |
Ask a specific question about this device
(158 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Ask a specific question about this device
(148 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
PVC films form a barrier to body fluids and bloodborne Pathogens. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.
The provided document describes the acceptance criteria and the study that proves Daxwell, LLC's Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet those criteria, primarily through a comparison to a predicate device and adherence to established ASTM and ISO standards.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Standard / Acceptance Criteria | Reported Device Performance |
---|---|---|
Dimension | ASTM D 5250-06(Reapproved 2011) | |
Length | ≥230mm | 232-240mm |
Width (Small) | 80-90mm | 81-90mm |
Width (Medium) | 90-100mm | 93-98mm |
Width (Large) | 100-110mm | 102-109mm |
Width (X-Large) | 110-120mm | 110-118mm |
Thickness (Fingertip) | ≥0.05mm | 0.08-0.11mm |
Thickness (Palm) | ≥0.08mm | 0.09-0.12mm |
Physical Properties | ASTM D 5250-06(Reapproved 2011) | |
Tensile strength (Before & After aging) | ≥11MPa | 15-20MPa |
Elongated rate (Before & After aging) | ≥300% | 360-420% |
Freedom from Pinholes | 21 CFR 800.20, ASTM D5250-06 (Reapproved 2011), ASTM D5151-06 (Reapproved 2011), Inspection Level I AQL2.5 | Passed Standard (Waterleak test), Meets ASTM D5151-06 (Reapproved 2011) |
Powder Residual | ASTM D 5250-06 (Reapproved 2011) and D6124-06 (Reaffirmation 2011) | Meets |
Ask a specific question about this device
(158 days)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder-Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).
-- How the device functions:
PVC films form a barrier to body fluids and bloodborne Pathogens
-- Scientific concepts that form the basis for the device
The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.
-- Physical and performance characteristics such as design, materials and physical properties:
PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.
The document describes the acceptance criteria and performance of the "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)" manufactured by Better Plastic Co., Ltd. against regulatory standards and a predicate device.
Table of Acceptance Criteria and Reported Device Performance
Feature & Description | Acceptance Criteria (Predicate / Standard) | Reported Device Performance (Subject Device) | Result of Comparison |
---|---|---|---|
Product Name | Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) (K120968) | Powder-Free Vinyl Patient Examination Gloves, Clear (Non-colored) | Same |
Product Code | LYZ (K120968) | LYZ | Same |
Size | Small/ Medium/ Large/X large (K120968) | Small/ Medium/ Large/X large | Substantially Equivalent |
Intended Use | (K120968) Disposable device for medical purposes, worn on examiner's hand/finger to prevent contamination between patient and examiner. | Disposable device for medical purposes, worn on examiner's hand/finger to prevent contamination between patient and examiner. | Substantially Equivalent |
Device Description & Specs | Meets ASTM D5250-06 (Reapproved 2011) | Meets ASTM D5250 -06 (Reapproved 2011) | Substantially Equivalent |
Dimensions - Length | Meets ASTM D5250-06 (Reapproved 2011) ≥230mm min. | Meets ASTM D5250-06 (Reapproved 2011) 230mm min for all sizes | Substantially Equivalent |
Dimensions - Width | Meets ASTM D5250-06 (Reapproved 2011) Small 80-90 mm, Medium 90-100mm, Large 100-110mm, X large 110-120 mm | Meets ASTM D5250-06 (Reapproved 2011) Small 85-87 mm, Medium 95-98 mm, Large 104-107mm, X large 112-116 mm | Substantially Equivalent |
Dimensions - Thickness | Meets ASTM D5250-06 (Reapproved 2011) Finger 0.05mm min., Palm 0.08mm min. | Meets ASTM D5250-06 (Reapproved 2011) Finger 0.05mm min., Palm 0.08mm min. | Substantially Equivalent |
Physical Properties | Meets ASTM D5250-06 (Reapproved 2011) Before aging/after aging: Elongation ≥300%, Tensile Strength ≥11MPa | Meets ASTM D5250-06 (Reapproved 2011) Before aging/after aging: Elongation ≥300%, Tensile Strength ≥ 11MPa | Substantially Equivalent |
Freedom from Pinholes | Meets 21 CFR 800.20, ASTM D5250-06 (Reapproved 2011), ASTM D 5151-06 (Reapproved 2011) | Meets ASTM D5151-06 (Reapproved 2011) Holes Inspection Level I AQL 2.5 | Substantially Equivalent |
Residual Powder | Meets ASTM D6124-06 (Reaffirmation 2011) | ASTM D6124-06 (Reaffirmation 2011) Results generated values below 2mg of residual powder | Substantially Equivalent |
Materials | PVC (K120968) | PVC | Substantially Equivalent |
Dusting or Donning Powder | PU (K120968) | PU / Surface Coating Agent | Substantially Equivalent |
Performance Data | Meets ASTM D5151-06 (Reapproved 2011), ASTM D5250-06 (Reapproved 2011), ASTM D6124-06 (Reaffirmation 2011) | Meets ASTM D5151-06 (Reapproved 2011), ASTM D5250-06 (Reapproved 2011), ASTM D6124-06 (Reaffirmation 2011) | Substantially Equivalent |
Single Patient Use | Single Patient Use (K120968) | Single Patient Use | Substantially Equivalent |
Biocompatibility | Non-irritant and non-sensitizer (studies meets ISO 10993-10:2002/Amd.1:2006) | Not an irritant, not a sensitizer (studies meets ISO 10993-10 Third Edition 2010-08-01) | Substantially Equivalent |
Labeling | -Powder Free, -devices color: Clear(Non-colored), -Patient Examination Glove, -Non sterile, -Single Use Only, - Manufactured For:, - Lot (K120968) | -Powder Free, -devices color: Clear(Non-colored), -Patient Examination Glove, -Non sterile, -Single Use Only, - Manufactured For:, - Lot | Substantially Equivalent |
Study Details:
The document describes a submission for a 510(k) premarket notification, which is a process to demonstrate that the new device is substantially equivalent to a legally marketed predicate device. The study is a comparison of the subject device's performance against established ASTM standards and the predicate device's characteristics rather than a traditional clinical study with patients.
-
Sample sizes used for the test set and the data provenance:
- The document does not specify a distinct "test set" in the context of a clinical trial or algorithm evaluation. Instead, the "testing" refers to the manufacturing quality control and performance evaluation of the gloves based on established ASTM standards.
- The data provenance is from non-clinical testing performed by the manufacturer, Better Plastic Co., Ltd. The country of origin for the manufacturing and testing is China, as indicated by the submitter's address. The data is prospective in the sense that it's generated for the submission of this specific device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable to this type of device and submission. "Ground truth" in the context of diagnostic AI or image analysis typically refers to definitively confirmed diagnoses by expert clinicians, pathology, or long-term outcomes. For patient examination gloves, the "truth" is whether they meet specified physical, chemical, and biological performance standards (e.g., tensile strength, freedom from pinholes, biocompatibility). These are evaluated through standardized laboratory tests, not by human expert consensus on a diagnostic outcome.
-
Adjudication method for the test set:
- Not applicable. The "adjudication" for this product involves laboratory testing against defined standards. There is no human observer consensus or adjudication required beyond verifying that the test results meet the specified criteria.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a medical device (patient examination gloves), not an AI diagnostic imaging tool or a system that involves human "readers" or AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This device is not an algorithm or AI system. Its performance is evaluated through physical and chemical testing.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for this device's performance is defined by adherence to established international and national standards for patient examination gloves, specifically:
- ASTM D5250-06 (Reaffirmation 2011): Standard Specification for Poly(Vinyl Chloride) Gloves for Medical Application.
- ASTM D5151-06 (Reapproved 2011): Standard Test Method for Detection of Holes in Medical Gloves.
- ASTM D6124-06 (Reaffirmation 2011): Standard Test Method for Residual Powder on Medical Gloves.
- 21 CFR 800.20: General requirements for medical devices.
- ISO 10993-10 Third Edition 2010-08-01: Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization.
- The testing also focuses on demonstrating substantial equivalence to a legally marketed predicate device (K120968) which presumably met these same standards.
- The "ground truth" for this device's performance is defined by adherence to established international and national standards for patient examination gloves, specifically:
-
The sample size for the training set:
- Not applicable. There is no "training set" in the context of this device. The manufacturing process is controlled, and samples are taken for quality assurance testing against the standards. This is not an AI/ML product that learns from a training dataset.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set for this type of device. The performance standards are established by consensus national and international bodies (e.g., ASTM, ISO, FDA regulations) for medical devices of this type.
Ask a specific question about this device
Page 1 of 6