Found 1 results

510(k) Data Aggregation

    K Number
    K945627
    Manufacturer
    Date Cleared
    1996-03-04

    (474 days)

    Product Code
    Regulation Number
    870.3610
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    PIKOS LP 01, PIKOS LP E01

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pikos LP 01/LP B01 models, like the Pikos 01/E01 models, are indicated for use in the following conditions:

    • · Sinus node arrest or bradycardia with or without AV conduction disorder.
    • · Intermittent or complete AV conduction block.
    • · Brady/tachy syndrome or other manifestation of sick sinus syndrome which results in symptomatic bradycardia.
    • · Atrial fibrillation and ventricular bradyarrhythmia.
    Device Description

    The Pikos LP 01/LP B01 is a multi-programmable single chamber pulse generator which is designed and recommended for use with atrial or ventricular leads. The Pikos LP 01/LP E01 models offer a limited number of programmable options compared to those of the Pikos 01/E01 models.

    AI/ML Overview

    This 510(k) submission (K945627) for the Pikos LP 01/LP B01 Implantable Pacemaker Pulse Generators focuses on demonstrating substantial equivalence to a predicate device (K914109/A, Pikos 01 and Pikos H01) rather than establishing new acceptance criteria through a clinical study. Therefore, most of the requested information regarding a study proving acceptance criteria will not be present.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly define a table of new acceptance criteria and corresponding performance metrics from a distinct study for this device. Instead, it relies on demonstrating that the new device performs equivalently to a predicate device (Pikos 01/E01 models) based on pre-established performance of that predicate.

    The qualification testing mentioned is primarily for design verification and validation of components and modifications, not for proving clinical performance against new acceptance criteria.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance
    Functional Equivalence:The Pikos LP 01/LP B01 is a multi-programmable single chamber pulse generator designed for use with atrial or ventricular leads, similar to the Pikos 01/E01 models.
    Intended Use Equivalence:"The Pikos LP 01/LP B01 models, like the Pikos 01/E01 models, are indicated for use in the following conditions: Sinus node arrest or bradycardia with or without AV conduction disorder. Intermittent or complete AV conduction block. Brady/tachy syndrome or other manifestation of sick sinus syndrome which results in symptomatic bradycardia. Atrial fibrillation and ventricular bradyarrhythmia." The intended uses are identical to the predicate device.
    Design/Component Qualification (Modifications):- Battery Qualification: "identical to that submitted in the Pikos 01/501 premarket notification (K914109/A)." (i.e., meets prior acceptance criteria).
    • Self-Sealing Header: "qualification reports describing the testing conducted for validation... included vibration, shock, temperature and transport tests meeting the applicable IS-1 requirements; header leakage resistance tests. Sterilization... was validated."
    • Hybrid Circuit Modifications: "qualified with complete electrical and mechanical testing." (Implies meeting internal design specifications and performance standards). |
      | Safety and Effectiveness: | "The product labeling includes instructions for use adequate to assure safe and effective operation of the device." (Relies on the established safety and effectiveness of the predicate device and verification of modifications). |

    2. Sample Size Used for the Test Set and the Data Provenance

    No specific "test set" in the context of clinical data for proving new acceptance criteria is described. The "qualification testing" refers to engineering and bench testing.

    • Sample Size: Not applicable in the context of a clinical test set from human subjects for this submission. The "sample" likely refers to components or finished devices subjected to various engineering tests (e.g., a certain number of self-sealing headers tested for leakage). These numbers are not provided in the summary.
    • Data Provenance: Not applicable in the context of clinical data. The qualification testing data would be generated in-house by BIOTRONIK GmbH & Co.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This submission is for a medical device that relies on engineering qualification and demonstrated equivalence, not a diagnostic algorithm requiring expert-established ground truth from a test set of human data.

    4. Adjudication Method for the Test Set

    Not applicable for the reasons stated above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study or AI components are mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable for the reasons stated above.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the qualification testing of components and modifications (e.g., self-sealing header, hybrid circuits), the "ground truth" would be engineering specifications, international standards (like IS-1 mentioned for the header), and internal design requirements.

    For the substantial equivalence claim, the "ground truth" is effectively the established safety and effectiveness profile and performance characteristics of the predicate device (Pikos 01/E01).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.

    In summary: This 510(k) submission is for an implantable pacemaker pulse generator that relies on demonstrating substantial equivalence to an already approved predicate device. The "qualification testing" described is for verifying the design and manufacturing of the new device and any modifications, ensuring it meets engineering standards and is comparable to the predicate. It does not involve a clinical study with human subjects designed to establish new performance acceptance criteria for a novel device.