Search Results
Found 1 results
510(k) Data Aggregation
(86 days)
PICCOLO TRIGLYCERIDES TEST SYSTEM
The Piccolo Triglycerides Test System used with the Piccolo Point-of-Care Chemistry Analyzer is intended for the in vitro quantitative determination of triglycerides in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism, or various endocrine disorders.
The Piccolo® Lipid Panel Reagent Disc (which contains the Piccolo® Triglycerides Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analvzer. Alternately, the disc mav also be used with serum.
This document describes the regulatory submission for the Abaxis Piccolo® Triglycerides Test System. The device is a laboratory test system designed to measure triglyceride levels in blood samples.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and the Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined set of thresholds. Instead, it presents performance characteristics of the new device and compares them to those of a legally marketed predicate device (Bayer Triglycerides-RA Assay). The underlying acceptance criterion for regulatory approval in this context is "substantial equivalence" to the predicate device, meaning the new device performs as safely and effectively. The tests conducted (linearity, precision, sample type comparison, and method comparison) are standard ways to demonstrate this equivalence for in vitro diagnostic devices.
Below is a table summarizing the reported device performance and comparing it to the characteristics of the predicate device, which implicitly defines the 'acceptance' for a substantially equivalent device.
Performance Characteristic | Acceptance Criteria (Implied by Predicate Device) | Reported Device Performance (Piccolo® Triglycerides Test System) |
---|---|---|
Intended Use | Quantitative analysis of triglycerides | Quantitative analysis of triglycerides |
Methodology | Enzymatic endpoint reaction | Enzymatic endpoint reaction |
Sample Type | Heparinized plasma and serum | Heparinized whole blood, heparinized plasma, and serum |
Sensitivity | 0.99 mA per mg/dL or 0.088 A per mmol/L | 3.82 mA per mg/dL or 0.338 A per mmol/L; 20 mg/dL |
Reagents | Dry reagents (reconstituted by user prior to use) and liquid reagents | Dry test-specific reagent beads and liquid diluent; reconstitution performed by analyzer |
Temperature of Reaction | 37°C | 37°C |
Calibration | Calibrated periodically using calibrators supplied by vendor | Bar code with factory calibrated lot specific data |
Assay Range | 0 - 500 mg/dL | 20 - 500 mg/dL |
Testing Environment | Professional use | Professional use |
Sample Size | 2 µL | Approx 100 µL |
Linearity (Slope) | (Implied to be close to 1) | 1.000 |
Linearity (Intercept) | (Implied to be close to 0) | 3.0 |
Linearity (Correlation Coefficient 'r') | (Implied to be close to 1) | 1.000 |
Precision (Serum 1, Mean) | (Not specified as a target, but precision is expected) | 206.8 mg/dL (Within-Run & Total) |
Precision (Serum 1, SD) | (Not specified as a target, but precision is expected) | 4.7 (Within-Run), 5.5 (Total) |
Precision (Serum 1, %CV) | (Not specified as a target, but precision is expected) | 2.3 (Within-Run), 2.6 (Total) |
Precision (Serum 2, Mean) | (Not specified as a target, but precision is expected) | 163.7 mg/dL (Within-Run & Total) |
Precision (Serum 2, SD) | (Not specified as a target, but precision is expected) | 1.8 (Within-Run), 2.4 (Total) |
Precision (Serum 2, %CV) | (Not specified as a target, but precision is expected) | 1.1 (Within-Run), 1.5 (Total) |
Method Comparison (Correlation Coefficient 'r') | (Implied to be close to 1 for substantial equivalence) | 0.999 |
Method Comparison (Slope) | (Implied to be close to 1 for substantial equivalence) | 0.983 (Linear), 0.984 (Deming) |
Method Comparison (Intercept) | (Implied to be close to 0 for substantial equivalence) | 8.2 (Linear), 8.1 (Deming) |
2. Sample Size Used for the Test Set and the Data Provenance:
- Linearity Study: The sample size for the linearity study is not explicitly stated in terms of number of patient samples, but the results for slope, intercept, and correlation coefficient are provided.
- Precision Study:
- Sample Size: n = 160 for both within-run and total precision for each serum level tested (Serum 1 and Serum 2). This appears to be 160 replicates per serum level, not 160 distinct patient samples.
- Data Provenance: Not specified, but generally, such studies use commercially available control sera or pooled patient samples.
- Sample Type Comparison Study:
- Sample Size: Samples from 20 different patients. Each sample type (serum, heparinized plasma, heparinized whole blood) was tested in quadruplicate.
- Data Provenance: Not specified, presumed to be clinical samples.
- Method Comparison Study:
- Sample Size: n = 172 data points. The document clarifies: "number of patient samples = n/2", meaning 86 unique patient samples were run in duplicate.
- Data Provenance: Not specified, but these are patient samples tested against a predicate device, presumably from a clinical setting. Retrospective or prospective is not specified, but given the nature of the study, it's likely a mix or prospective collection for comparison purposes. No country of origin is mentioned.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not provided in the document. For in vitro diagnostic devices like this, the "ground truth" for the test set (patient samples) is typically established by the reference method (the predicate device, in the case of method comparison) or by highly accurate laboratory methods, rather than by human expert consensus or adjudication.
4. Adjudication Method for the Test Set:
This information is not applicable and therefore not provided. Adjudication methods (like 2+1, 3+1) are common in image-based diagnostic studies where human readers interpret results which are then reconciled. For quantitative laboratory tests, the result is a numerical value, and "adjudication" in the traditional sense is not performed on the test results themselves. Method comparison studies statistically compare the new device's numerical results to those of the predicate device.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable and therefore not provided. MRMC studies are specific to diagnostic tools that involve human interpretation, often in imaging. The Abaxis Piccolo® Triglycerides Test System is an automated in vitro diagnostic device for quantitative chemical analysis, not an AI-assisted diagnostic tool that involves human reader interpretation for primary diagnosis.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Yes, the studies presented (Linearity, Precision, Sample Type Comparison, Method Comparison) represent the standalone performance of the Abaxis Piccolo® Triglycerides Test System. The device is designed to operate as an automated system producing a quantitative result without direct human interpretation influencing the measurement itself (beyond sample collection and loading). The studies assess the device's inherent analytical performance.
7. The Type of Ground Truth Used:
For the method comparison study, the predicate device's measurement (Bayer Triglycerides-RA Assay) served as the de facto "ground truth" for comparison to establish substantial equivalence. For linearity, precision, and sample type comparison, the "ground truth" is typically established by the inherent analytical capabilities of the device itself and its ability to consistently and accurately measure known concentrations or produce reproducible results.
8. The Sample Size for the Training Set:
This information is not applicable and therefore not provided. The Abaxis Piccolo® Triglycerides Test System is a chemical analyzer, not a machine learning or AI-based system that requires a "training set" in the computational sense. Its reagents and methods are developed through biochemical and analytical chemistry principles, not by training an algorithm on a dataset.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable due to the nature of the device (see point 8).
Ask a specific question about this device
Page 1 of 1