Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K011688
    Manufacturer
    Date Cleared
    2002-03-20

    (293 days)

    Product Code
    Regulation Number
    884.1425
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PelvicFlexer is intended to assist women in performing Kegel Exercises which may help control stress urinary incontinence.
    Treatment of stress and /or Urge incontinence in women

    Device Description

    The PelvicFlexer is a comprehensive, behaviorally based program designed for independent use by incontinence people at home. The program includes a PelvicFlexer device, instructional manual and direct support via phone and internet. The PelvicFlexer is hand-held and inserted into the vagina. The strength of the muscle contraction is reflected by the stainless steel springs.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for the PelvicFlexer Exercise Device

    This submission describes the PelvicFlexer Exercise Device and establishes its substantial equivalence to the predicate device, Pelvic Muscle Therapy (K# 002830). The assessment is based on non-clinical tests rather than clinical studies to demonstrate safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in quantitative metrics within the provided document for device performance. Instead, the "acceptance criteria" appear to revolve around demonstrating biocompatibility, safety, and functional equivalence to the predicate device through non-clinical testing.

    FeatureAcceptance Criteria (Implied)Reported Device Performance (PelvicFlexer)
    BiocompatibilityMeet ISO 10993 standards for medical devices.Extensive testing performed by NAMSA against ISO 10993 standards: Cytotoxicity (ISO Elution Method), Muscle Implantation Test (7 day), Acute Systemic Toxicity, Delayed Contact Sensitization (Maximization Method), Hemolysis Test (USP - direct and extraction methods). Results indicated material is biocompatible, nontoxic, and well tolerated by mucosal membranes.
    Material SafetyMaterials used must be safe for patient contact.Addressed by biocompatibility testing (ISO 10993).
    Functional Safety (Spring Load)Assure safety and effectiveness of the spring mechanism.Safety Testing (spring load testing) performed. (Specific performance metrics not detailed, but states assurance of safety and effectiveness).
    Intended UseTo assist women in performing Kegel Exercises which may help control stress urinary incontinence.Identical intended use as the predicate device.
    Target PopulationWomen with mild incontinence.Identical to the predicate device.
    Reusable/Single Patient UseMust be reusable and for single patient use.Reusable, single patient device.

    2. Sample Size Used for the Test Set and Data Provenance

    The submission explicitly states, "Discussion of Clinical Tests Performed: Not applicable."
    Therefore, there was no human "test set" in the context of clinical performance evaluation. The "test set" for the non-clinical tests would refer to the samples of materials and device components subjected to the biocompatibility and safety testing. The document does not specify the exact number of samples used for these non-clinical tests.

    The data provenance for the non-clinical tests would be the testing laboratories (e.g., NAMSA) where the biocompatibility and safety tests were conducted. The country of origin is not specified for these tests. The nature of these tests is laboratory-based (bench testing), not retrospective or prospective clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable, as no clinical test set was used to evaluate device performance against a medical ground truth (e.g., disease diagnosis, treatment efficacy). The "ground truth" for the non-clinical tests was established by adherence to ISO 10993 standards and general engineering safety principles.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring expert adjudication was performed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done, as no clinical studies involving human readers or AI assistance were conducted.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable, as this device is a physical medical device, not an algorithm or AI software. Therefore, no standalone algorithm performance study was relevant or performed.

    7. Type of Ground Truth Used

    The "ground truth" for demonstrating the safety and biochemical compatibility of the device's materials was based on internationally recognized standards and guidelines for medical device biocompatibility, specifically ISO 10993. For the functional safety (spring load testing), the "ground truth" was internal engineering specifications and generally accepted safety parameters for mechanical components, aiming to assure safety and effectiveness.

    8. Sample Size for the Training Set

    Not applicable, as this is a physical medical device and not an AI/ML algorithm requiring a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1