Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K140970
    Manufacturer
    Date Cleared
    2014-05-13

    (27 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ONE-PIECE PEEK FUSION IMPLANT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The One-Piece PEEK Fusion Implant is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe and mallet toe.

    Device Description

    The One-Piece PEEK Fusion Implant is manufactured from polyetheretherketone (PEEK). It is threaded on one end and ridged on the other to engage either side of an osteotomy or reconstruction site. The One-Piece PEEK Fusion Implant is threaded into the proximal phalanx and then it is press fit into the middle phalanx, allowing for reduction and fixation of the bone fragments.

    AI/ML Overview

    This report describes the acceptance criteria and the study proving the device meets those criteria, based on the provided 510(k) summary for the One-Piece PEEK Fusion Implant.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Tensile StrengthPre-determined acceptance criteriaMet the pre-determined acceptance criteria
    Bending StrengthPre-determined acceptance criteriaMet the pre-determined acceptance criteria
    Torsional StrengthPre-determined acceptance criteriaMet the pre-determined acceptance criteria
    Fatigue TestingPre-determined acceptance criteriaMet the pre-determined acceptance criteria
    Post-Fatigue Tensile StrengthPre-determined acceptance criteriaMet the pre-determined acceptance criteria

    Note: The specific numerical values for the acceptance criteria and corresponding device performance are not provided in the 510(k) summary. It only states that the device "met the pre-determined acceptance criteria." The summary also mentions that these are "the same tests performed on the predicate device to establish its substantial equivalence in submission K133515."

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size used for the test set or the data provenance (country of origin, retrospective/prospective). The testing described is nonclinical (mechanical/material properties), not clinical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The described testing is nonclinical mechanical testing, not a study requiring expert-established ground truth on patient data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was nonclinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was mentioned. This device is an implant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is an implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the nonclinical testing, the "ground truth" would be the engineering or materials science standards and specifications defining acceptable performance for each test (Tensile Strength, Bending Strength, etc.).

    8. The sample size for the training set

    Not applicable. This device is an implant, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133515
    Manufacturer
    Date Cleared
    2014-03-28

    (133 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PEEK FUSION IMPLANT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PEEK Fusion Implant is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe and mallet toe.

    Device Description

    The PEEK Fusion Implant is manufactured from polyetheretone (PEEK) and is composed of two threaded components, each designed to engage either side of an osteotomy or reconstruction site. Once in position, the components mate with one another allowing for reduction and fixation of the bone fragments.

    AI/ML Overview

    The PEEK Fusion Implant underwent non-clinical performance testing to ensure its substantial equivalence to predicate devices and its intended performance.

    Acceptance Criteria and Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Tensile Strength TestingThe device should demonstrate comparable tensile strength to the predicate device.Equivalent time zero tensile strength compared to the predicate.
    Bending Strength TestingThe device should demonstrate comparable or greater bending strength compared to the predicate device.Greater bending strength compared to the predicate.
    Torsion Strength TestingThe device should demonstrate comparable or greater torsional strength compared to the predicate device.Greater torsional strength compared to the predicate.
    Shear Strength TestingThe device should demonstrate comparable or greater shear strength compared to the predicate device.Greater shear strength compared to the predicate.
    Screw Insertion TestingNot explicitly stated as a pass/fail criterion in the provided text, but implied as part of demonstrating functional performance.Implied successful insertion with adequate torque, as driver torque failure loads exceeded required insertion torque.
    Driver Torque to Failure TestingThe driver/implant interface torque failure loads should exceed the torque required to insert the implants.Driver/implant interface torque failure loads exceeded the torque required to insert the implants.
    Association & Disassociation TestingNo disassociation of the implants should be observed during fatigue testing.No disassociation of the implants was observed during fatigue testing.
    Four-point Bending Fatigue TestingThe device should demonstrate comparable fatigue performance to the predicate device.Equivalent fatigue performance compared to the predicate.
    Tensile Load after Fatigue TestingThe device should demonstrate comparable or greater tensile strength after undergoing fatigue testing when compared to the predicate.Greater post-fatigue tensile strength compared to the predicate.

    Study Details:

    1. Sample Size for Test Set and Data Provenance:

      • The document states that "All testing was performed on test units representative of finished devices." However, the specific sample size for each mechanical test (e.g., number of implants tested for tensile strength) is not provided.
      • The data provenance is from laboratory testing (non-clinical) conducted by the manufacturer, MTP Solutions LLC. There is no information regarding country of origin, but it is implied to be within the scope of the manufacturer's operations. The testing is described as pre-clinical and therefore inherently retrospective in the context of device design and validation.
    2. Number of Experts and Qualifications for Ground Truth:

      • This is a non-clinical mechanical testing study, not a clinical study involving expert assessment of patient data. Therefore, the concept of "experts used to establish ground truth" related to clinical interpretation or diagnosis does not apply. The "ground truth" here is the objective measurement of mechanical properties of the device and its predicates.
    3. Adjudication Method:

      • Not applicable as this is a non-clinical mechanical testing study, not a study requiring adjudication of clinical observations or diagnoses. The results are based on objective measurement of physical properties.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. This document describes non-clinical mechanical testing of an implantable device, not an AI-assisted diagnostic or interpretive system that would typically use MRMC studies.
    5. Standalone (Algorithm Only) Performance:

      • Not applicable. This document describes a physical medical device (an implant), not a software algorithm or AI. Therefore, standalone algorithm performance is not relevant.
    6. Type of Ground Truth Used:

      • The ground truth for the performance claims derived directly from objective physical measurements and mechanical testing data (e.g., force, torque, displacement) as specified in the various tests (tensile, bending, torsion, fatigue, etc.). This data establishes the inherent properties of the PEEK Fusion Implant and its predicate devices.
    7. Sample Size for Training Set:

      • Not applicable. This is a non-clinical mechanical performance study of a physical device. There is no "training set" in the context of machine learning or AI models. The "training" in this context would refer to the engineering design and manufacturing processes, which are not detailed as a "training set" here.
    8. How Ground Truth for Training Set Was Established:

      • Not applicable. As stated above, there is no "training set" in the context of an AI/ML model for this device. The physical properties being tested are intrinsic to the manufactured device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1