Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K053123
    Device Name
    PD CATH
    Manufacturer
    Date Cleared
    2006-03-29

    (142 days)

    Product Code
    Regulation Number
    876.5630
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PD CATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PD Cath is indicated for use in attaining long term peritoneal access for peritoneal dialysis via the peritoneum. The catheter is intended for implantation dwell time of greater than 30 days.

    Device Description

    The PD Cath is a radiopaque silicone, single lumen catheter used to insert and remove peritoneal fluid. The fixed Dacron cuff allows for tissue ingrowth for long-term placement. A flexible silicone overtube is permanently bonded to the lumen passages and the cuff to insulate the cuff from forces applied to the external or internal lumens.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for a medical device called "PD Cath." This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting extensive clinical studies to establish safety and efficacy from scratch.

    Therefore, the information typically requested in your prompt (such as acceptance criteria tables, sample sizes for test sets, expert ground truth, MRMC studies, standalone performance, and extensive training set details) is generally not applicable to a 510(k) submission for a device like the PD Cath, which is a relatively low-risk device without complex diagnostic algorithms or image analysis components.

    Here's a breakdown based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: The document directly states that "Clinical data was not deemed necessary since in vitro testing was sufficient to demonstrate safety and efficacy by way of comparison to legally marketed predicate device intended for peritoneal dialysis." The acceptance criteria are implicitly tied to demonstrating substantial equivalence in performance characteristics to the predicate devices, specifically in terms of flow rate and tensile strength. The exact numerical acceptance criteria are not specified in the public summary.
    • Reported Device Performance:
      Performance MetricReported Device Performance (PD Cath)Predicate Device (Dermaport™ Access Device - K894131, Missouri Catheter - K874650)
      Flow RateSubstantially equivalentData not provided, but inferred to be similar to legally marketed devices
      Tensile StrengthSubstantially equivalentData not provided, but inferred to be similar to legally marketed devices

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not applicable. No clinical test set data was used. The performance testing was in vitro.
    • Data Provenance: Not applicable. In vitro testing was conducted, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical test set requiring expert ground truth was performed for this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set requiring adjudication was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical catheter, not an AI-powered diagnostic tool. MRMC studies or AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is not an algorithmic device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable for the reasons mentioned above. The "ground truth" for this submission was the established performance characteristics of the legally marketed predicate devices, against which the in vitro performance of the PD Cath was compared.

    8. The sample size for the training set

    • Not applicable. No AI/machine learning training set was used.

    9. How the ground truth for the training set was established

    • Not applicable. No AI/machine learning training set was used.

    Summary of the Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria was an in vitro performance test demonstrating that the PD Cath's flow rate and tensile strength were "substantially equivalent" to legally marketed predicate devices (Dermaport™ Access Device -- K894131 and Missouri Catheter-K874650). The FDA determined that clinical data was not necessary for this substantial equivalence determination. The specific details of the in vitro testing methodology, raw data, and precise quantitative equivalence metrics are not provided in this public summary, as is typical for 510(k) summaries. The acceptance criteria were met by showing that the physical characteristics (flow rate, tensile strength) of the PD Cath were comparable to established, safe, and effective peritoneal dialysis catheters already on the market.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031351
    Manufacturer
    Date Cleared
    2003-09-17

    (141 days)

    Product Code
    Regulation Number
    876.5630
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    FLEX-NECK PD CATHETER, INFANT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    If the patient is a suitable candidate for peritoneal dialysis (PD) therapy and is too small for the current Flex-Neck™ catheters, the Flex-Neck™ Catheter, Infant, can be implanted either surgically, peritoneoscopically or percutaneously. The only contraindication to implantation of the Flex-Neck™ PD Catheter, Infant, is if the patient is not a candidate for peritoneal dialysis.

    Device Description

    This device consists of the following components:
    a. 1-Flex-Neck™ Catheter, Infant, made of long-term, implantable grade silicone tubing with radiopaque strip and 1 or 2 polyester cuffs. All components are made of the same materials as the Flex-Neck™ Catheter, Adult and Pediatric. The Infant catheter also has 4 rows of side holes punched into the tubing at the distal end to facilitate inflow and outflow of dialysate.
    b. 1-Right hand and 1-left hand tunnel marking stencils to mark and guide and thereby assist the implanting doctor to make a sound subcutaneous tunnel for the catheter.
    c. 1-Surgical grade marking pen
    d. 1-Catheter connector, plastic
    e. 1-Catheter cap, plastic
    f. 1-Tube (20 cc) of water-soluble lubricating gel

    AI/ML Overview

    This document describes a 510(k) submission for the Flex-Neck™ PD Catheter, Infant. The submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria for device performance. Therefore, many of the requested elements (like MRMC study, separate training/test sets, expert qualifications, etc.) are not applicable to the information provided.

    However, based on the available text, here's an analysis of what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance
    Mechanical IntegrityFunctional testingPerformed to demonstrate mechanical integrity.
    RetentionFunctional testingPerformed to demonstrate retention.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Functional testing has been performed," but does not specify the sample size used for these tests.
    The data provenance is not explicitly stated, but given this is a 510(k) submission for a medical device in the US, the testing would generally be conducted by the manufacturer, MEDIGROUP, Inc. It is implicitly internal company testing rather than external clinical data. It is retrospective in the sense that the testing was completed prior to the submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. The submission relies on functional testing for mechanical integrity and retention, not on expert-adjudicated ground truth like in image analysis or diagnostic studies.

    4. Adjudication Method for the Test Set

    Not applicable. See point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This type of study is not mentioned as it's not relevant for demonstrating substantial equivalence of a physical medical device based on mechanical properties. The focus is on the device's physical and functional attributes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical medical catheter, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the functional testing would be the physical measurement and observation of the device's mechanical integrity and retention properties under defined test conditions, evaluated against engineering specifications. This is not "expert consensus" or "pathology" in the typical sense, but rather objective physical testing results.

    8. The Sample Size for the Training Set

    Not applicable. This submission describes a medical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970159
    Manufacturer
    Date Cleared
    1997-09-05

    (232 days)

    Product Code
    Regulation Number
    876.5630
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    FLEX-NECK PD CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    If the patient is a suitable candidate for peritoneal dialysis (PD) therapy, the Flex-Neck™ PD Catheter can be implanted either surgically or peritoneoscopically. The only contraindication to implantation of the Flex-Neck™ PD Catheter is if the patient is not a candidate for peritoneal dialysis. Numerous prior surgeries or suspected or documented intraperitoneal adhesions may be relative contraindications to PD. However, since the Y-TEC® System of peritoneoscopic implantation enables inspection of the peritoneum to confirm the presence of adhesions and to avoid them, all patients who are suitable for PD can receive this catheter.

    Device Description

    The current Tenckhoff peritoneal catheter is generally successful in providing peritoneal access for fluid infusion and drainage. The Flex-Neck™ PD Catheter is extruded of medical long-term implantable silicone basically identical to the existing Tenckhoff peritoneal catheters. All these peritoneal catheters include one or two Dacron® cuffs, the "deep" cuff in the abdominal musculature and the "superficial" cuff near the skin exit site of the catheter. Ingrowth of fibrous tissue over a few weeks period results in a thick fibrous plug preventing passage of bacteria around the Dacron® cuff. The coiled or straight intraperitoneal portion of the catheters contains multiple holes: 0.75mm in the Flex-Neck™ and 1.0mm in the Tenckhoff. Generally, the smaller the hole, the less likely omentum is to become trapped within the holes. During infusion of PD fluid, most of the flow is through the tip of the catheter. During outflow of the fluid, the soft surfaces of the peritoneum over bowel loops and omentum may be drawn to the tip, however, outflow continues through the multiple side holes.

    AI/ML Overview

    The provided text, K970159, describes a 510(k) premarket notification for the Flex-Neck™ PD Catheter. This document asserts the substantial equivalence of the new device to legally marketed predicate devices rather than proving the device meets specific acceptance criteria through a study.

    Therefore, the information required to populate the fields regarding acceptance criteria and a study demonstrating performance is not present in the provided text. The document is generally focused on comparing the new device's characteristics to existing devices to establish substantial equivalence for regulatory approval.

    Specifically:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document focuses on comparing physical characteristics (material, dimensions, number of cuffs, hole diameter) to predicate devices, not performance against acceptance criteria.
    2. Sample size used for the test set and the data provenance: Not mentioned, as no study demonstrating performance against acceptance criteria is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no study with a test set requiring ground truth establishment is described.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device (catheter), not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device (catheter), not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as no study with a test set requiring ground truth is described.
    8. The sample size for the training set: Not applicable. This is a medical device, and the concept of a "training set" is not relevant here.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1