510(k) Data Aggregation

    K Number
    K981431
    Date Cleared
    1999-03-03

    (317 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Panasonic wrist blood pressure meter, model EW280, is intended to measure systolic and diastolic blood pressure using a pressurized cuff worn around the wrist.

    Device Description

    The wrist blood pressure meter is a battery-charged non-invasive digital electronic blood pressure meter manufactured by Matsushita Electric Works, Ltd., (MEW) Osaka, Japan. The wrist blood pressure meter is intended to measure systolic and diastolic blood pressure using a pressurized cuff worn around the wrist.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Panasonic Wrist Blood Pressure Meter, Model EW280, focusing on acceptance criteria and the study conducted.

    Acceptance Criteria and Device Performance for Panasonic Wrist Blood Pressure Meter, Model EW280

    The provided 510(k) summary focuses on demonstrating substantial equivalence to predicate devices, rather than establishing de novo performance acceptance criteria for a novel device. The performance data presented primarily addresses verification that modifications to the new model (EW280) did not negatively impact the safety and effectiveness compared to the cleared predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/SpecificationAcceptance Criteria (from Predicate Device EW273)Reported Device Performance (EW280)
    Main body (w) x (h) x (d)77 x 66 x 23 (mm)67 x 67 x 26 (mm)