Search Results
Found 3 results
510(k) Data Aggregation
(130 days)
PALACOS R pro, PALACOS R+G pro, PALACOS MV+G pro
PALACOS® R pro is indicated for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
PALACOS® R+G pro is indicated for use in the second stage revision for total joint arthroplasty after the initial infection has been cleared.
PALACOS® MV+G pro is indicated for use in the second stage revision for total joint arthroplasty after the initial infection has been cleared.
PALACOS® R pro is a standard-setting, high-viscosity, radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). It contains the X-ray contrast medium zirconium dioxide. To improve visibility in the surgical field, it has been colored with chlorophyll-copper-complex (E141). The bone cement consists of two components and is prepared immediately before use by mixing the polymer powder (= powder) with the monomer liquid (= liquid). A ductile dough forms that sets within a few minutes.
PALACOS® R+G pro is a standard-setting, high-viscosity, radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). It contains the aminoglycoside antibiotic gentamicin to protect the cured bone cement and contiguous tissue against colonization by bacteria that are sensitive to gentamicin. It contains the X-ray contrast medium dioxide. To improve visibility in the surgical field, it has been colored with chlorophyll-copper-complex (E141). The bone cement consists of two components and is prepared immediately before use by mixing the polymer powder (= powder) with the monomer liquid (= liquid). A ductile dough forms that sets within a few minutes.
PALACOS® MV+G pro is a standard-setting, medium viscosity, radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). It contains the aminoglycoside antibiotic gentamicin to protect the cured bone cement and contiguous tissue against colonization by bacteria that are sensitive to gentamicin. It contains the X-ray contrast medium zirconium dioxide. To improve visibility in the surgical field, it has been colored with chlorophyll-copper-complex (E141). The bone cement consists of two components and is prepared immediately before use by mixing the polymer powder (= powder) with the monomer liquid). A ductile dough forms that sets within a few minutes.
This FDA 510(k) Premarket Notification is for PALACOS® R pro, PALACOS® R+G pro, and PALACOS® MV+G pro bone cements. It does not describe an AI/ML device, a clinical study with human readers, or a ground truth established by experts. Instead, it focuses on demonstrating substantial equivalence to previously cleared predicate devices through non-clinical performance testing.
Here's a breakdown based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document does not present a formal table of acceptance criteria with reported numerical performance values in the way you might expect for an AI device (e.g., sensitivity, specificity, AUC). Instead, it states that the devices meet existing industry standards and guidance documents. The "acceptance criteria" are implied by adherence to these standards, and the "reported performance" is that the devices meet them.
Acceptance Criteria Category | Reference Standard/Guidance | Reported Device Performance |
---|---|---|
Biocompatibility | FDA Guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (Sept 4, 2020) and ISO 10993-1 | - Focus on changes in mixing/application system and accessories. |
- Concluded that possible transfer of leachables is low.
- Cytotoxicity test performed to support this conclusion.
- No negative impact estimated from changes. |
| Mechanical & Functional | FDA Guidance document Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement - Guidance for Industry and FDA (July 17, 2002) (Annex I and Annex II) | - Tests conducted to demonstrate intended function, safety, and effectiveness. - Data provided to state substantial equivalence to predicate device. |
| Sterilization | ISO 11135 | - Validation performed using ethylene oxide gassing. - Cycle designed for sterile units with a sterility assurance level (SAL) of 10⁻⁶.
- Chosen sterilization process considered valid, showing equivalence to predicate device. |
| Pyrogenicity | ANSI/AAMI ST72 (LAL test) | - Subject device meets endotoxin limit specification of ≤ 20 EU/device. - Shows equivalence to predicate device. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This document details non-clinical laboratory testing, not a test set of patient data. Therefore, the concepts of "sample size used for the test set" and "data provenance (country of origin, retrospective/prospective)" as they relate to patient data or imagery do not apply. The testing was conducted in a laboratory setting, likely in Germany where the manufacturer is located, or by authorized testing facilities.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The ground truth for this type of device (bone cement) is not established by human experts in the context of medical image interpretation. The "truth" is determined by established physical, chemical, and biological testing standards and measurements.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. There is no human adjudication process described, as the testing is based on objective laboratory measurements against defined standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this submission is based on:
- Established objective standards: International Standards (e.g., ISO 10993-1, ISO 11135, ANSI/AAMI ST72) and FDA Guidance documents (e.g., "Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement").
- Laboratory measurements: Results from specific tests (cytotoxicity, mechanical property measurements, sterility assurance level determination, LAL test for endotoxins) designed to objectively evaluate the device's conformance to these standards and demonstrate substantial equivalence to predicate devices.
- Chemical and material analysis: Confirmation of identical (or acceptably different) chemical composition and materials compared to predicate devices.
8. The sample size for the training set
Not applicable. This is not a machine learning device and therefore does not have a "training set" in that context.
9. How the ground truth for the training set was established
Not applicable. As above, there is no "training set" for this type of device.
Ask a specific question about this device
(87 days)
PALACOS R, PALACOS R pro, PALACOS R+G, PALACOS R+G pro, PALACOS MV+G
PALACOS® R is indicated for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
PALACOS® R pro is indicated for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
PALACOS® R+G is indicated for use in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.
PALACOS® R+G pro is indicated for use in the second stage revision for total joint arthroplasty after the initial infection has been cleared.
PALACOS® MV+G is indicated for use in the second stage revision for total joint arthroplasty after the initial infection has been cleared.
A bundled Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance for modifications to PALACOS® R, PALACOS® R pro, PALACOS® R+G, PALACOS® R+G pro, and PALACOS® MV+G previously cleared in K030902, K150119, K031673, K142157 and K050855.
Modifications include changes of instructions for use (IFU) and labels as well as the addition of MRI safety information.
This submission encompasses multiple devices that have similar intended use and indications for use as well as rely on similar data.
PALACOS® bone cements without Gentamicin:
PALACOS® R and PALACOS® R pro are polymethylmethacrylate (PMMA) bone cements.
PALACOS® R: is a standard-setting, high-viscosity, PMMA-based bone cement for orthopaedic surgery.
PALACOS® R pro: is a PMMA based PALACOS® R bone cement, packed in a closed mixing and application system ready for processing (ready-to-mix).
The bone cements consist of two components, a monomer liquid and a polymer powder. The liquid component contains the monomer, accelerator, and a stabilizer. The powder contains the polymer, X-Ray-opacifier, and initiator. They are intended for single-use and are provided sterile (ethylene oxide and sterile filtration).
PALACOS® bone cements with Gentamicin:
PALACOS® R+G, PALACOS® R+G pro and PALACOS® MV+G are PMMA bone cements, containing the antibiotic Gentamicin.
PALACOS® R+G (previously cleared under the name PALACOS® G); is a standard-setting. high-viscosity, PMMA-based bone cement for orthopaedic surgery.
PALACOS® R+G pro is a PMMA based PALACOS® R+G bone cement, packed in a closed mixing and application system ready for processing (ready-to-mix) and
PALACOS® MV+G (previously cleared under the name PALAMED G) is a standard-setting, medium-viscosity, radiopaque, poly(methyl methacrylate)-based bone cement for orthopaedic surgery.
The bone cements consist of two components, a monomer liquid and a polymer powder. The liquid component contains the monomer, accelerator, and a stabilizer. The powder contains the polymer, X-Ray-opacifier, initiator and the antibiotic Gentamicin. They are intended for single-use and are provided sterile (ethylene oxide and sterile filtration).
The provided document is a 510(k) premarket notification for PMMA bone cements (PALACOS® R, PALACOS® R pro, PALACOS® R+G, PALACOS® R+G pro, and PALACOS® MV+G). This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than establishing new acceptance criteria or conducting a clinical study to prove device performance against specific metrics.
Therefore, the document does not contain the information requested regarding acceptance criteria, a study proving the device meets those criteria, sample sizes for test/training sets, data provenance, number or qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance studies, or how ground truth was established for training sets.
Instead, the document states:
"A risk-based assessment was performed as per DIN EN ISO 14971 to evaluate the impact of the modifications to the labelling... The modified IFUs do not alter the previously validated packaging or sterilization data as well as the existing results of non-clinical performance testing and biocompatibility in accordance with the FDA Class II Special Controls Guidance Document 'PMMA Bone Cement'. The risk-based assessment concludes that the IFU changes do not significantly affect the device's risk profile because no new risks or significantly modified existing risks are identifically based rationale has been prepared to designate the bone cements as "MR Safe"."
And, for each device: "No clinical testing... has been conducted."
The substantial equivalence determination is based on a comparison of technological characteristics with predicate devices, confirming that the subject devices are modified versions with no significant changes to their safety or effectiveness.
Ask a specific question about this device
(148 days)
PALACOS R pro
PALACOS® R pro is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
PALACOS® R pro is an acrylic bone cement for use in orthopedic surgery. It is formed from powder and liquid by exothermic polymerization. It secures the fixation of the grafted artificial joint improving the transfer of forces at the interface implant - bone. The bone cement powder and liquid of PALACOS® R pro are pre-packed in a vacuum mixing and application system. This reduces the user steps and processing time during mixing of the bone cement. It also decreases the exposure to monomer fumes. PALACOS® R pro is available in one size: 75 g and is for single use. The PALACOS® R pro device includes: The mixing and application device pre-packed with the bone cement powder One ampoule of monomer liquid pre-packed in a monomer cartridge Accessories: a nozzle, a femur pressurizer, a vacuum sealed vacuum tube and in a separate box, an adaptor ring for the use with bone cement gun.
The provided document is a 510(k) summary for a medical device called PALACOS® R pro, a PMMA bone cement. It is a submission to the FDA for market clearance, asserting substantial equivalence to a predicate device. The information details non-clinical testing performed to demonstrate this equivalence, rather than a clinical study with acceptance criteria for device performance in human patients.
Therefore, many of the requested categories for acceptance criteria and study details are not directly applicable within the context of this document, as it focuses on demonstrating bio-mechanical and material equivalence to an already approved device.
Here's an attempt to address your request based on the provided text, noting where information is not available or not applicable:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria here are derived from the standards and tests applied to the predicate device and the reference device. The "reported device performance" indicates that PALACOS® R pro (or its relevant precursors) met these criteria, supporting substantial equivalence.
Acceptance Criteria (Based on Predicate/Reference Device Standards) | Reported Device Performance (PALACOS® R pro / Reference Device) |
---|---|
Stability of liquid component (per ISO 5833) | Characterized (met standards of predicate/reference) |
Maximum temperature (per ISO 5833) | Characterized (met standards of predicate/reference) |
Setting time (per ISO 5833) | Characterized (met standards of predicate/reference) |
Intrusion (per ISO 5833) | Characterized (met standards of predicate/reference) |
Compressive strength (per ISO 5833) | Characterized (met standards of predicate/reference) |
Bending modulus (per ISO 5833) | Characterized (met standards of predicate/reference) |
Bending strength (per ISO 5833) | Characterized (met standards of predicate/reference) |
Impact strength (per Dynstat test method) | Measured (for PALACOS® R+G pro, considered worst-case) |
Bending strength (per Dynstat test method) | Measured (for PALACOS® R+G pro, considered worst-case) |
EtO sterilization validation (per ISO 11135) | Validated |
Biocompatibility: Cytotoxicity (per ISO 10993) | Performed (results valid for PALACOS® R pro) |
Biocompatibility: Irritation (per ISO 10993) | Performed (results valid for PALACOS® R pro) |
Biocompatibility: Sensitization (per ISO 10993) | Performed (results valid for PALACOS® R pro) |
Biocompatibility: Acute systemic toxicity (per ISO 10993) | Performed (results valid for PALACOS® R pro) |
Biocompatibility: Genotoxicity (per ISO 10993) | Performed (results valid for PALACOS® R pro) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide specific sample sizes for the individual non-clinical tests (e.g., number of specimens for compressive strength testing). It states that the tests were performed "per ISO 5833" and "per ISO 10993," which would implicitly define testing protocols and sample sizes within those standards.
- Test set sample size: Not explicitly stated for each test, but implied by adherence to ISO standards.
- Data Provenance: The manufacturer, Heraeus Medical GmbH, is located in Wehrheim, Germany. The tests appear to be conducted as part of the product development and regulatory submission process, which is inherently prospective for demonstrating compliance. The data refers to non-clinical laboratory testing.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This concept is not applicable here. The "ground truth" for these tests are the established parameters and thresholds defined by international standards (ISO 5833, ISO 11135, ISO 10993) for bone cement properties and sterilization/biocompatibility. These are objective measurements, not subjective expert assessments.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for reconciling differing expert opinions in clinical ground truth establishment. Here, the "truth" is determined by direct physical and chemical measurements against defined standard limits.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, this was not done. MRMC studies are associated with evaluating diagnostic imaging devices where human readers interpret results, often to assess AI's impact on reader performance. This document concerns a bone cement, which is a therapeutic device, not a diagnostic one.
6. If a Standalone (algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a bone cement, a material used in surgery, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the non-clinical tests is based on established international standards and measurement methodologies for material properties, sterilization efficacy, and biocompatibility. For instance, compressive strength has a defined physical measurement and an acceptable range per ISO 5833.
8. The Sample Size for the Training Set
Not applicable. This device is a material product, not an AI/machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable for the same reason as point 8.