    Device Name :

    PAIRED HYDROGEL ELECTRODE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Paired Hydrogel Electrode is used for evoked electrophysiological monitoring in soft tissue surgeries were motor nerves are at risk.

    Device Description

    The Paired Hydrogel Electrodes are a modification of the M203KEN electrode currently distributed as the Singlet and Triplet Hydrogel Electrodes by Neurovision Medical Products. The Paired Hydrogel Electrode is created by attaching two Singlet M203KEN electrodes by their lead wires to the same DIN plug (shorting the electrodes). The Paired version is designed to provide a single pole of electrophysiological monitoring across a surface too wide to be covered by a single electrode.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Neurovision Medical Paired Hydrogel Electrode, focusing on acceptance criteria and supporting studies:

    This submission is a Special 510(k), indicating a modification to an already cleared device. The focus of the performance testing is typically to demonstrate that the modified device (Paired Hydrogel Electrode) is substantially equivalent to the predicate devices and that the modification does not introduce new questions of safety oreffectiveness, or alter the fundamental scientific technology.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaPerformed Test/Reported Performance
    Material BiocompatibilityFreedom from sensitizationTested free of Sensitization (ISO 10993: Part 10)
    Freedom from cytotoxicityTested free of Cytotoxicity (Agarose Overlay Method)
    Freedom from irritationTested free of Irritation (ISO 10993: Part 10)
    Physical PerformanceElectrode adhesionDemonstrated equivalence to predicate singlet
    ImpedanceDemonstrated equivalence to predicate singlet
    Electrophysiological PropertiesElectrophysiological propertiesDemonstrated equivalence to predicate singlet
    Shelf-LifeShelf-life durationDemonstrated a 3-year shelf life (ANSI/AAMI EC12:2000)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for performance tests (adhesion, impedance, electrophysiological properties).
      • For biocompatibility, it states "Animal testing demonstrated the Biocompatibility," implying animal subjects, but the number is not specified.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The testing seems to be internal or contracted by the manufacturer. The studies are prospective as they were conducted to support the 510(k) submission for the modified device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This type of device (hydrogel electrode) does not typically involve human-read interpretations or ground truth established by experts in the context of clinical imagery. The performance criteria are technical and measurable (e.g., adhesion strength, electrical impedance, biological reactions).

    4. Adjudication Method for the Test Set

    Not applicable, as the performance testing involves measurable technical parameters, not subjective assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is not the type of device or study for which an MRMC comparative effectiveness study would be performed. The device is a surface electrode for physiological monitoring, not an imaging or diagnostic AI tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. The device itself is hardware (an electrode) that facilitates electrophysiological monitoring, which then requires an EMG unit (compatible device) and human interpretation of the signals. There is no "algorithm only" performance for an electrode.

    7. Type of Ground Truth Used

    • Biocompatibility: In-vitro (cytotoxicity via Agarose Overlay) and in-vivo (irritation and sensitization via animal testing according to ISO 10993).
    • Physical and Electrophysiological Properties: Measured technical specifications of the device compared against the known performance of the predicate device (Singlet Hydrogel Electrode). The "ground truth" here is the established performance characteristics of the predicate device.
    • Shelf-Life: Time-dependent measurements demonstrating stability over the claimed 3-year period, adhering to ANSI/AAMI EC12:2000 standards.

    8. Sample Size for the Training Set

    Not applicable. This is a hardware device (electrode), not a software/AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this hardware device.