510(k) Data Aggregation

    K Number
    K142265
    Date Cleared
    2014-11-13

    (90 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ONE Snare Endovascular Microsnare System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ONE Snare Endovascular Microspare System is intended for use in adult and pediations for the retrieval and manipulation of atraumatic foreign bodies located in the coronary and peripheral cardiovascular system and the extracranial anatomy.

    Device Description

    ONE Snare Endovascular Microsnare System contains: (1) Snare, (1) Snare Catheter, (1) Insertion Tool and (1) Torque Device. The snare is constructed of nitinol cable and a gold plated tungsten loop. The pre-formed snare loop can be introduced through catheters without risk of snare deformation because of the snare's super-elastic construction. The snare catheter is constructed of polyether block amide (Pebax®) and contains a platinum/iridium radiopaque marker band.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the ONE Snare Endovascular Microsnare System. However, it does not explicitly define acceptance criteria or specific study results in the format requested. The document primarily lists performance attributes that were tested to demonstrate substantial equivalence to predicate devices, rather than providing quantitative acceptance criteria and corresponding detailed device performance data.

    Based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific numerical or quantitative acceptance criteria and corresponding performance metrics for the ONE Snare Endovascular Microsnare System. Instead, it lists a comprehensive set of tests performed and states that these tests were "successfully completed" or "designed to demonstrate that the device meets critical design specifications as well as clinical performance attributes for its intended use."

    The identified performance attributes tested include (but are not limited to):

    CategoryTested Performance Attributes
    Catheter & Insertion ToolSurface Finish, Radiopacity, Catheter tip atraumatic, Marker band placement, Effective Length, ID & OD, Catheter stiffness, Markerband retention, Tip integrity, Torque Strength, Kink resistance, Buckling force, Freedom from Liquid leakage, Freedom form Air leakage, Peak tensile force of catheter at hub, Force at break insertion tool at hub, Hub with Female luer testing, Corrosion Resistance
    ONE Snare Endovascular Microsnare SystemSnare Loop Resistance, Guide Catheter resistance, Snare shape integrity, Insertion Tool Fitment, Snare insertion, Insertion Tool tear away force, Snare jacket after insertion tool removal inspection, Snare loop protruding
    SnareSize designation, Snare Head Assembly, Radio detectability, Surface Finish, Tensile strength, Torqueability, Torque Strength, Flexing Test, Fracture Test, Tip Flexibility, Corrosion Resistance
    BiocompatibilityCytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Genotoxicity, Hemolysis, Thrombogenicity, Complement Activation, Chemical Tests
    Packaging PerformanceBefore and after exposure to accelerated aging and simulated shipping and handling conditions (bubble emission, dye penetration, seal peel tensile strength, burst strength, visual inspection)

    The document implies that the device "meets critical design specifications as well as clinical performance attributes for its intended use" by successfully completing these tests, but the specific numerical "acceptance criteria" and "reported device performance" are not detailed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for any of the tests conducted. It also does not provide information on data provenance (e.g., country of origin, retrospective or prospective) as the studies appear to be bench and laboratory-based performance tests rather than clinical studies with patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This information is not applicable as the studies described are in vitro or bench tests, not clinical studies requiring expert interpretation of medical images or patient outcomes for ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as #3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No MRMC comparative effectiveness study is mentioned in the provided text. The studies focus on the device's physical and functional performance and biocompatibility.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable as the device is a physical medical instrument (an endovascular microsnare system), not an algorithm or AI software. There is no "algorithm only" performance to evaluate.

    7. The Type of Ground Truth Used

    For the performance tests described (e.g., tensile strength, corrosion resistance, biocompatibility), the "ground truth" would be objective measurements and established standards (e.g., ISO, ASTM, FDA guidance documents). The tests were designed to demonstrate adherence to critical engineering specifications and regulatory requirements.

    8. The Sample Size for the Training Set

    This is not applicable as the device is a physical medical device, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as #8.

    Summary of Study Type:

    The studies described are a series of bench testing, laboratory testing, and biocompatibility assessments to demonstrate the physical, chemical, and functional performance of the ONE Snare Endovascular Microsnare System. The purpose of these tests was to establish substantial equivalence to predicate devices under the 510(k) pathway, not to evaluate clinical efficacy or diagnostic accuracy using patient data or expert adjudicated ground truth. The acceptance of the device is based on its ability to perform robustly under various conditions and meet recognized standards (e.g., ISO, ASTM) and design specifications.