Search Results
Found 1 results
510(k) Data Aggregation
(158 days)
NovaLine SP-C35 Transducer Protector (956007)
The NovaLine SP-C35 Transducer Protector is a single-use, disposable, prescription device intended for use as a protective device for pressure monitors and to help prevent contamination of the fluid pathway in hemodialysis delivery systems.
The NovaLine SP-C35 Transducer Protector consists of a plastic housing that contains a hydrophobic filter with a reference pore size of 0.2 µ m. The housing has male and female luer lock connectors that allow its attachment between a blood tubing set and the pressure monitor on a hemodialysis machine.
The NovaLine SP-C35 Transducer Protector is to be used as protective device for pressure monitors as well as to help maintain the sterility of the blood tubing fluid pathway. The 0.2 ↓ m hydrophobic filter helps prevent cross-contamination by viruses, bacterial and particulate matter while preventing the flow of fluids to the hemodialysis machine pressure monitor.
The devices are packaged sterile and labeled for single use only.
The proposed devices are provided in sterile condition, it is subject to e-beam radiation sterilization prior to release to achieve a Sterility Assurance Level (SAL) of 10-6.
Here's a breakdown of the acceptance criteria and study information for the NovaLine SP-C35 Transducer Protector based on the provided FDA 510(k) summary.
Note: This document describes a medical device, not an AI or software-based medical device. Therefore, many of the requested fields related to AI performance, ground truth, and expert evaluation are not applicable.
1. Table of Acceptance Criteria and Reported Device Performance
Device: NovaLine SP-C35 Transducer Protector
Acceptance Criteria Category | Specific Test / Characteristic | Acceptance Criteria (Target) | Reported Device Performance (Result) | Relevant Standard (if applicable) |
---|---|---|---|---|
Physical Performance | Appearance | Complies with specifications | Complies with specifications | --- |
Structural Integrity | Maintained | Maintained | --- | |
Dimension | Complies with standard | Complies with standard | ISO 80369-7:2021 | |
Sterility | SAL of 10^-6 | Achieved SAL of 10^-6 | --- | |
Positive Pressure Limitation | 500 mmHg | 500 mmHg (matches predicate) | --- | |
Negative Pressure Limitation | -500 mmHg | -500 mmHg (matches predicate) | --- | |
Connections Conformance | Conforms to standard | Conforms to standard | ISO 80369-7:2016 | |
Biocompatibility | Cytotoxicity | Non-toxic | Complies with ISO 10993-5 | ISO 10993-5 |
Sensitization | Non-sensitizing | Complies with ISO 10993-10 | ISO 10993-10 | |
Irritation/Intradermal Reactivity | Non-irritating | Complies with ISO 10993-23 | ISO 10993-23 | |
Systemic Toxicity/Acute Systemic Toxicity | Non-toxic | Complies with ISO 10993-11 | ISO 10993-11 | |
Pyrogens/Material-mediated Pyrogens | Non-pyrogenic | Complies with ISO 10993-11 | ISO 10993-11 | |
In Vitro Hemolysis | Non-hemolytic | Complies with ISO 10993-4 | ISO 10993-4 | |
Complement Activity | No adverse complement activity | Complies with ISO 10993-4 | ISO 10993-4 | |
Partial Thromboplastin Time | No adverse effect | Complies with ISO 10993-4 | ISO 10993-4 | |
Platelet Count | No adverse effect | Complies with ISO 10993-4 | ISO 10993-4 | |
Hematology | No adverse effect | Complies with ISO 10993-4 | ISO 10993-4 | |
Sterilization | Sterility Assurance Level | SAL of 10^-6 | Achieved SAL of 10^-6 | --- |
Packaging | Seal Strength | Complies with standard | Complies with standard | ASTM F88/F88-15 |
Dye Penetration | Complies with standard | Complies with standard | ASTM F1929-15 | |
Shelf Life | Packaging Performance (aged samples) | Complies with acceptance criteria | Complies with acceptance criteria | ISO 80369-7 (for connections) |
Performance Tests (aged samples) | Complies with acceptance criteria | Complies with acceptance criteria | --- | |
Microbiological | Bacterial Endotoxin | |||
Microbial Challenge | Sterile Barrier Integrity | Maintains sterile barrier | Complies with standards | ASTM F1671/F1671M-2022, YYT1497-2016, ASTM F838-2020 |
Distribution Simulation | Withstands distribution stresses | Complies with standard | Complies with standard | ASTM D4169-22 |
Comparative Performance | Overall performance vs. predicate | Equivalent | Equivalent to predicate device | --- |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each test within the finished product performance, packaging, or shelf-life evaluations. It mentions "proposed devices" were tested, implying a statistical sample was used according to the relevant standards, but the specific number is not provided.
Data Provenance: The tests were conducted by Bain Medical Equipment (Guangzhou) Co., Ltd. in China. The data is retrospective in the sense that the testing was performed on manufactured devices to support the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not Applicable. This is a medical device (transducer protector) and not an AI/software device. The "ground truth" here refers to the physical and biological properties of the device, established through standardized laboratory testing, not human expert interpretation of data.
4. Adjudication Method for the Test Set
Not Applicable. As this is not an AI/software device involving human interpretation, there is no adjudication method in the traditional sense. Device performance is determined by objective, measurable laboratory tests against pre-defined specifications and standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is relevant for imaging or diagnostic AI, comparing human readers with and without AI assistance. This device is a component of a hemodialysis system, and its performance is assessed through physical, chemical, and biological testing.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not Applicable. This device is hardware, not an algorithm or software.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by:
- Standardized Test Results: Objective measurements from various laboratory tests (e.g., dimensional measurements, sterile barrier integrity, biocompatibility assays).
- Compliance with Recognized Standards: Adherence to international and national standards such as ISO 80369-7, ISO 10993 series, ASTM F88, ASTM F1929, ASTM D4169, USP , ASTM F1671, etc.
- Predicate Device Equivalence: Direct comparison to the performance characteristics of a legally marketed predicate device (Nipro Transducer Protector TP-SURE, K072988) where applicable.
8. The Sample Size for the Training Set
Not Applicable. This is a physical medical device, not an AI/machine learning model, so there is no training set.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. See response to #8.
Ask a specific question about this device
Page 1 of 1