Search Results
Found 1 results
510(k) Data Aggregation
(28 days)
Navigator HD Ureteral Access Sheath Set 11/13 F x 28cm (M0062502210); Navigator HD Ureteral Access
The Navigator HD Ureteral Access Sheath is indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract.
Navigator™ HD Ureteral Access Sheath Set consists of a semi-rigid outer sheath and a semi-rigid inner dilator with interlocking hub. The set permits utilization of a guidewire to assist in device placement.
Sheath: The Navigator™ HD Sheath has a sheath hub which is over-molded onto the sheath shaft has three layers: Outer Pebax layer, Reinforced stainless steel coil enhancing torqueability and maneuverability, Inner PTFE liner. The sheath outer layer contains Barium Sulphate as a radiopacifier. In addition to the stainlesssteel coil, a radiopaque marker band is sandwiched between the Pebax and PTFE layers. The radiopaque marker band is located at the sheath tip and, along with the radiopacifier added to the sheath outer layer, enhances fluoroscopic visualization of sheath placement during the procedure. A hydrophilic coating is applied to the entire length of the shaft to reduce friction and facilitate placement of the device.
Dilator: The Navigator™ HD dilator has a hub with a standard luer lock at the proximal end. The tab on the dilator hub allows it to lock into the sheath hub. The dilator shaft is made of extruded tube Pellethane to allow for dimensional stability, low moisture absorption, and chemical resistance. The dilator tip is made of a softer grade of Pellethane to improve tip flexibility to assist in atraumatic placement. Both the dilator shaft and tip contain Barium Sulfate as a radiopacifier, enhancing visualization under fluoroscopy during the procedure. A hydrophilic coating is applied to the entire length of the dilator to reduce friction and facilitate placement of the device, either when assembled with the sheath or when used separately as a dilatation device.
This is a 510(k) premarket notification for a medical device, not an AI/ML device. Therefore, the questions regarding acceptance criteria and studies that prove the device meets these criteria in the context of AI/ML are not applicable here.
The document discusses the substantial equivalence of the Navigator™ HD Ureteral Access Sheath Set to a previously cleared predicate device.
Here's the relevant information that can be extracted, addressing the spirit of the request as much as possible for a non-AI/ML device:
1. A table of acceptance criteria and the reported device performance:
The document doesn't present "acceptance criteria" in the typical AI/ML sense (e.g., specific sensitivity/specificity thresholds). Instead, it relies on demonstrating substantial equivalence to an existing predicate device. The performance is assessed by comparing technological characteristics and performing design verification and usability testing related to a minor change.
Characteristic / "Acceptance Criteria" | Proposed Device Performance (Navigator™ HD Ureteral Access Sheath Set K250517) |
---|---|
Intended Use / Indication for Use | Identical to Predicate (K140323): Facilitate passage of endoscopes, urological instruments, and injection of fluids into the urinary tract in endoscopic procedures. |
Reusability | Identical to Predicate: Single Use |
Supplied Condition | Identical to Predicate: Sterile |
Sterilization Method | Identical to Predicate: Ethylene Oxide (EO) |
Packaging | Identical to Predicate: Tyvek/Poly pouch |
Hydrophilic Coating | Similar but different (Lubricent UV540 vs. Lubricent UV460 in Predicate). Performance verified via Design Verification and Summative Usability. |
Dilator & Sheath Mechanical Specifications | Unchanged from Predicate via implied equivalence. Confirmed by Design Verification. |
Sizes Offered | Identical to Predicate: 11/13 F, 12/14 F, 13/15 F |
Length (cm) | Identical to Predicate: 28, 36 & 46 |
Biocompatibility | Biological evaluation assessment concluded no biocompatibility risks, identical contact classification to predicate. |
Shelf-life impact of change | Additional Design Verification testing executed to verify no impact. |
2. Sample size used for the test set and the data provenance:
This is a physical medical device. The "test set" would refer to samples of the device undergoing engineering performance testing. The document does not specify sample sizes for the "Design Verification" or "Summative Usability" testing, nor does it mention data provenance (e.g., country of origin, retrospective/prospective clinical data). These would typically be detailed in internal test reports referenced by the submission but not included in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as this is not an AI/ML device requiring expert ground truth for classification or diagnosis. The "ground truth" for the device's performance relies on engineering specifications and established safety/efficacy profiles of similar devices.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable for a physical medical device. Adjudication methods are relevant for subjective interpretations of data, typically in clinical studies or AI/ML ground truth establishment.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI/ML device, and no human reader study with or without AI assistance was performed or needed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the device's acceptable performance is based on:
- Engineering specifications and standards: Ensuring the device meets its design requirements.
- Performance testing results: Demonstrating functional equivalence and safety (e.g., mechanical performance, biocompatibility).
- Historical performance of predicate device: The K140323 device, which has a proven track record.
8. The sample size for the training set:
Not applicable. This is a physical medical device, not an AI/ML model that requires a training set.
9. How the ground truth for the training set was established:
Not applicable, as there is no training set for a physical medical device.
Ask a specific question about this device
Page 1 of 1