    Device Name :

    NO TOUCH+FOREHEAD NTF 3000 THERMOMETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The No Touch + Forehead Thermometer (Model NTF3000US) is a non-sterile, reusable clinical thermometer intended for the intermittent determination of human body temperature in a touch and no touch on the centre of the forehead as the measurement site on people of all ages.

    Device Description

    The No Touch+Forehead Thermometer ( Model NTF3000US) is a hand-held, battery powered device designed to measure human body temperature. The NTF3000 is an Infrared thermometer that converts a user's forehead temperature using the infrared energy emitted in the area around the user's forehead to an oral equivalent temperature when placed in contact with the subject's forehead or up to two (2) inches away. The NTF 3000 thermometer uses a thermopile sensor with integrated thermistor for the target reading, a thermistor mounted in the head of the thermometer for ambient temperature readings, a parabolic mirror to help focus the infrared energy emitted from the forehead, and an infrared distance sensor for detection of contact or non-contact use and compensation of the temperature reading.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the KAZ USA, Inc. No Touch+ Forehead Thermometer (Model NTF3000US), based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The primary acceptance criteria for the No Touch+ Forehead Thermometer are established by its conformance to the ASTM E1965-03 standard for clinical accuracy and repeatability. The device needed to demonstrate non-inferiority or substantial equivalence to the predicate device against a gold standard.

    Acceptance Criteria (from ASTM E1965-03)Reported Device Performance (No Touch+ Forehead Thermometer)
    Bias: Clinical acceptability (less than predicate device when compared to reference)Within clinical acceptability
    Clinical Repeatability: Less than 0.3 °C (0.58 °F)Statistically acceptable (less than 0.3 °C or 0.58 °F)

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document states that a comparison study was performed across four age groups: 0-12 months, 12 months-