Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Integra Ti6 Internal Fixation System,Subtalar MBA System,MetaSurg Subtalar Implant,NewDeal BOLD Screw,NewDeal
    HALLU Lock Plate System,NewDeal HALLU Plates,QWIX Positioning Screw,SPIN Snap-Off Screw,NewDeal TIBIAXYS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use/indications for use of the predicate devices identified remain the same as previously cleared in their respective 510(k)s. The respective 510(k)s for the devices can be referenced in the predicate device section above.

    Device Description

    The purpose of this submission is the addition of MR Conditional information to the labeling for the predicate devices. The addition of MR labeling to the subject devices does not impact indications, materials, design features or dimensions, packaging or sterilization.

    AI/ML Overview

    This is a 510(k) summary for a submission that adds MR Conditional information to the labeling of several existing medical devices. The submission does not introduce new devices or changes to the fundamental design, materials, or indications for use of the listed devices. Therefore, the "acceptance criteria" and "device performance" in this context relate to the MR compatibility of the devices according to established standards.

    Here's the breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Test)Reported Device Performance
    Magnetically Induced Displacement Force (ASTM F2052)The devices were tested and found to be MR conditional. (Specific force values or thresholds are not provided in this summary, but the conclusion states conditional MRI safety was established).
    Magnetically Induced Torque (ASTM F2213)The devices were tested and found to be MR conditional. (Specific torque values or thresholds are not provided in this summary, but the conclusion states conditional MRI safety was established).
    RF-induced Heating (ASTM F2182)The devices were tested and found to be MR conditional. (Specific temperature increases or thresholds are not provided in this summary, but the conclusion states conditional MRI safety was established).
    Image Artifact (ASTM F2119)The devices were tested and found to be MR conditional. (Specific artifact sizes or impacts are not provided in this summary, but the conclusion states conditional MRI safety was established).
    Overall MR Compatibility (per ASTM F2503 and FDA Guidance)The completed MR compatibility testing establishes the conditional safety and compatibility of the passive implant devices in the MR environment, and supports the addition of MR Conditional labeling.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in the provided text. For MR compatibility testing, the "sample size" typically refers to the number of device models/configurations tested. It's implied that "the devices" (referring to the listed product lines and their components) were tested.
    • Data Provenance: The nature of this testing (MR compatibility) suggests it was conducted in a controlled environment as a prospective evaluation of the devices. The country of origin of the data is not specified but would likely have been where the testing laboratory is located, presumably in a country with recognized testing standards (e.g., USA or Europe).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This type of submission (MR compatibility of implants) does not typically involve human expert "ground truth" derived from clinical images. The "ground truth" here is objective measurements against engineering standards for MR safety and compatibility. The "experts" would be the engineers and physicists conducting the tests and interpreting the results according to ASTM standards and FDA guidance. Their qualifications would be expertise in MR safety testing and relevant engineering fields.

    4. Adjudication Method for the Test Set

    • Not applicable. This is not a study requiring adjudication of human-interpreted data. The results are based on objective physical measurements and adherence to specified test protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This submission is for device labeling updates based on physical properties (MR compatibility), not for evaluating the clinical effectiveness of a diagnostic or therapeutic algorithm with human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No, a standalone algorithm performance study was not done. This submission is about the physical properties of medical implants in an MRI environment, not about an AI algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" used is defined by internationally recognized engineering standards for MR compatibility: ASTM F2052 (displacement force), ASTM F2213 (torque), ASTM F2182 (RF-induced heating), and ASTM F2119 (image artifact). The FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" also serves as a framework for the "ground truth" criteria.

    8. The Sample Size for the Training Set

    • Not applicable. This submission is about MR compatibility testing of existing devices, not about developing or training an AI algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083154
    Manufacturer
    Date Cleared
    2009-07-10

    (259 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    NEWDEAL HALLU LOCK PLATE SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HALLU® Lock Plates are intended for fixation of fractures, osteotomies or arthrodesis of the first metatarso-phalangeal joint. Including cases of:

    • Hallux rigidus
    • Severe hallux valgus (IM angle >20° and HV angle >40°)
    • Deformity from rheumatoid arthritis
    • Failed previous surgical procedure
    • Traumatic arthritis
    • Neuromuscular instability.
      The HALLU® Lock plates must be fixed with the SURFIX® fixed angle locking system and with the SURFTX®-Alpha variable angle locking system of 2.7mm or 3.0mm diameter (screws and lock-screws).
      Addition of a Newdeal® QWIX® screw crossing the joint is strongly recommended for optimal arthrodesis consolidation
    Device Description

    The Newdeal HALLU® Lock Plate System is a low profile Titanium plate dedicated to first metatarso-phalangeal arthrodesis. The HALLU Lock C plates are pre-bent with 10° valgus and 10° dorsiflexion. The HALLU Lock S plates are available with 10° valgus and 10° or 5° dorsiflexion. Pre-operative bending is possible to completely fit to the anatomical shape of the joint. Their fixation is provided by SURFIX and SURFIX ALPHA locking screws.

    AI/ML Overview

    The provided text describes a medical device, the "Newdeal HALLU® Lock Plate System," which is a bone fixation appliance. It is a 510(k) summary, indicating that it aims to demonstrate substantial equivalence to an existing device rather than presenting entirely new clinical trial data for novel acceptance criteria.

    Therefore, the structure of the response will reflect the information available in a 510(k) summary for a substantially equivalent device, focusing on mechanical testing rather than extensive clinical studies.

    Here's the information as requested, based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Equivalence to Predicate Device (HALLU® Plate System, K021626)Mechanical tests have been carried out. Results have shown that the mechanical properties of the modified HALLU® Lock Plate System are equivalent to the properties of the unmodified device, HALLU® Plates, K021626.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes mechanical testing, not a clinical study involving human subjects or data related to patient outcomes. Therefore, information regarding "sample size used for the test set" in the context of clinical data, "data provenance (e.g. country of origin of the data, retrospective or prospective)," "number of experts," "adjudication method," "MRMC study," "standalone performance," "type of ground truth (e.g., pathology, outcomes data)," "training set sample size," and "how ground truth for training set was established" are not applicable to this type of submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable (as per explanation in point 2). The ground truth for mechanical testing is typically based on established biomechanical standards and measurements, not expert clinical consensus.

    4. Adjudication Method for the Test Set

    Not applicable (as per explanation in point 2).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a bone fixation plate, and the submission focuses on mechanical equivalence. An MRMC study is relevant for diagnostic imaging AI, which is not the subject here.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Not applicable. This device is a physical implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    For the mechanical testing performed, the ground truth would be established through objective physical measurements and engineering analyses under controlled laboratory conditions, demonstrating that the new device's mechanical properties (e.g., strength, stiffness, fatigue resistance) fall within acceptable predefined limits or are equivalent to the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set was Established

    Not applicable (as per explanation in point 8).


    Ask a Question

    Ask a specific question about this device

    Page 1 of 1